F. Chee, T. Fernando, A. Savkin, P. V. van Heerden
{"title":"The use of MiniMed CGMS in real-time glucose monitoring","authors":"F. Chee, T. Fernando, A. Savkin, P. V. van Heerden","doi":"10.1109/ANZIIS.2001.974068","DOIUrl":null,"url":null,"abstract":"A study was conducted to examine the use of the MiniMed(R) Continuous Glucose Monitoring System (CGMS) to measure blood sugar level (BSL) in real-time. Eight patients participated in the study after informed consent was obtained. When more than a 3 mmol/L deviation occurred between CGMS reading and BSL as determined by conventional blood glucose meter using blood sample drawn from an arterial line, CGMS was re-calibrated, and \"regression calibration\" was applied to better estimate the real-time BSL. The mean difference between the CGMS and Meter BSL readings was found to be -0.2 /spl plusmn/5.3 mmol/L (Mean/spl plusmn/1.96/spl times/Standard Deviation). Clarke's Error grid analysis showed that 64.6% of the CGMS BSL readings were clinically accurate (i.e. <20% deviation from Meter BSL) while 30.9% were clinically acceptable. A calibration technique suitable for use in estimating BSL in real-time is needed to make CGMS useful in clinical situations.","PeriodicalId":383878,"journal":{"name":"The Seventh Australian and New Zealand Intelligent Information Systems Conference, 2001","volume":"3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Seventh Australian and New Zealand Intelligent Information Systems Conference, 2001","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ANZIIS.2001.974068","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5
Abstract
A study was conducted to examine the use of the MiniMed(R) Continuous Glucose Monitoring System (CGMS) to measure blood sugar level (BSL) in real-time. Eight patients participated in the study after informed consent was obtained. When more than a 3 mmol/L deviation occurred between CGMS reading and BSL as determined by conventional blood glucose meter using blood sample drawn from an arterial line, CGMS was re-calibrated, and "regression calibration" was applied to better estimate the real-time BSL. The mean difference between the CGMS and Meter BSL readings was found to be -0.2 /spl plusmn/5.3 mmol/L (Mean/spl plusmn/1.96/spl times/Standard Deviation). Clarke's Error grid analysis showed that 64.6% of the CGMS BSL readings were clinically accurate (i.e. <20% deviation from Meter BSL) while 30.9% were clinically acceptable. A calibration technique suitable for use in estimating BSL in real-time is needed to make CGMS useful in clinical situations.