Installation, commissioning, and performance evaluation of Bhabhatron-II TAW – An Indian-made telecobalt unit

Shashi Bhushan Sharma, G. Sarma, Mithu Barthakur, Pranjal Goswami, Bhaveshwar Yadav, Shrayanti Goswami
{"title":"Installation, commissioning, and performance evaluation of Bhabhatron-II TAW – An Indian-made telecobalt unit","authors":"Shashi Bhushan Sharma, G. Sarma, Mithu Barthakur, Pranjal Goswami, Bhaveshwar Yadav, Shrayanti Goswami","doi":"10.4103/JRMT.JRMT_12_20","DOIUrl":null,"url":null,"abstract":"Aim: The aim of this study was to report commissioning and performance evaluation of Bhabhatron-II TAW (an Indian made machine). Materials and Methods: Commissioning comprises of following quality assurance (QA) (a) electrical QA, (b) mechanical QA, (c) radiation QA and (d) survey of installation. Electrical QA were done for various components of the unit for its functionality. Mechanical QA were done and the measured data were verified as per guidelines of Atomic Energy Regulatory Board (AERB), India and vendor's specified values. Cylindrical ionization chamber, electrometer, ready pack films, water phantom, thermometer and barometer were used during radiation QA and survey of installation procedures. Moreover, the data so generated were fed to treatment planning system (TPS) for dose calculation and also kept in proper tabular manner for clinical use. Results: Couch and collimator parameters were found to comply the tolerance criteria. Source strength of 172.2 roentgen per minute at 1 meter (RMM) was measured with an absolute dose rate of 225.76 cGy/min at dmax for 10cm x 10cm field size at normal treatment distance (NTD). Shutter timer error of 0.004 min was measured and incorporated in the clinical data. A maximum of 3.66 % variation was noted for 300 (15W × 20) physical wedge transmission factor between measured and vendor specified value. Radiation leakage through beam limiting device (BLD) and outside BLD were found to be within the limits prescribed by AERB. Conclusion: Commissioning is a prerequisite process after installation of a radiation therapy machine for medical purpose. Bhabhatron-II machine has several advantages over other telecobalt machine. Features like high source storage capacity of machine head, one pair of asymmetric jaw, closure of jaws to 0 cm x 0 cm during radiation emergency and presence of one motorized wedge filter makes it unique and yet cost effective telecobalt machine..","PeriodicalId":319549,"journal":{"name":"Journal of Radiation Medicine in the Tropics","volume":"13 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiation Medicine in the Tropics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/JRMT.JRMT_12_20","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Aim: The aim of this study was to report commissioning and performance evaluation of Bhabhatron-II TAW (an Indian made machine). Materials and Methods: Commissioning comprises of following quality assurance (QA) (a) electrical QA, (b) mechanical QA, (c) radiation QA and (d) survey of installation. Electrical QA were done for various components of the unit for its functionality. Mechanical QA were done and the measured data were verified as per guidelines of Atomic Energy Regulatory Board (AERB), India and vendor's specified values. Cylindrical ionization chamber, electrometer, ready pack films, water phantom, thermometer and barometer were used during radiation QA and survey of installation procedures. Moreover, the data so generated were fed to treatment planning system (TPS) for dose calculation and also kept in proper tabular manner for clinical use. Results: Couch and collimator parameters were found to comply the tolerance criteria. Source strength of 172.2 roentgen per minute at 1 meter (RMM) was measured with an absolute dose rate of 225.76 cGy/min at dmax for 10cm x 10cm field size at normal treatment distance (NTD). Shutter timer error of 0.004 min was measured and incorporated in the clinical data. A maximum of 3.66 % variation was noted for 300 (15W × 20) physical wedge transmission factor between measured and vendor specified value. Radiation leakage through beam limiting device (BLD) and outside BLD were found to be within the limits prescribed by AERB. Conclusion: Commissioning is a prerequisite process after installation of a radiation therapy machine for medical purpose. Bhabhatron-II machine has several advantages over other telecobalt machine. Features like high source storage capacity of machine head, one pair of asymmetric jaw, closure of jaws to 0 cm x 0 cm during radiation emergency and presence of one motorized wedge filter makes it unique and yet cost effective telecobalt machine..
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
印度制造的远程弹道导弹装置Bhabhatron-II TAW的安装、调试和性能评估
目的:本研究的目的是报告Bhabhatron-II TAW(印度制造的机器)的调试和性能评估。材料和方法:调试包括以下质量保证(QA):(a)电气质量保证(QA), (b)机械质量保证(QA), (c)辐射质量保证(QA)和(d)安装检查。为了保证设备的功能,我们对设备的各个部件进行了电气质量保证。根据印度原子能管理委员会(AERB)的指导方针和供应商的指定值,进行了机械质量保证和测量数据验证。在辐射质量保证和安装程序调查中,使用了圆柱形电离室、静电计、ready pack薄膜、水影、温度计和气压计。并将生成的数据输入治疗计划系统(TPS)进行剂量计算,并以适当的表格形式保存,供临床使用。结果:测得的Couch和准直仪参数符合公差标准。在正常治疗距离(NTD)下,在10cm × 10cm场尺寸下,源强度为172.2伦支/分钟(RMM),绝对剂量率为225.76 cGy/min。测量到的快门定时器误差为0.004 min,并纳入临床资料。当测量值为300 (15W × 20)时,测量值与供应商规定值之间的最大变化为3.66%。通过限束装置和限束装置外的辐射泄漏均在AERB规定的范围内。结论:调试是医疗用放射治疗机安装后的必备环节。与其他远程钴机相比,Bhabhatron-II机有几个优点。特点像机头的高源存储容量,一对不对称颚,关闭颚到0厘米x 0厘米在辐射紧急情况和一个电动楔形过滤器的存在使它独特的,但成本效益的远程钴机。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Commissioning and performance evaluation of varian truebeam linear accelerator Correlation of hepatobiliary ultrasound findings with CD4+ count and viral load in adult patients with human immunodeficiency virus/acquired immune deficiency syndrome in South-East Nigeria Early experience of high-dose-rate interstitial brachytherapy for prostate cancer in Nigeria Acute abdomen in pregnancy: Ovarian torsion diagnosed as hemorrhagic cyst on ultrasonography The correlation between dose homogeneity index and target volume in intracavitary brachytherapy treatment plans of cervical cancer patients
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1