Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System

IF 0.6 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING World Journal of Nuclear Medicine Pub Date : 2024-01-22 DOI:10.1055/s-0044-1778709
Manoj Kumar Singh, Krishan Kant Agarwal, Manish Vishwakarma, Hemant Patel
{"title":"Technical NEMA NU2–2018 Performance Assessment of Time-of-Flight-Integrated Digital PET-CT System","authors":"Manoj Kumar Singh, Krishan Kant Agarwal, Manish Vishwakarma, Hemant Patel","doi":"10.1055/s-0044-1778709","DOIUrl":null,"url":null,"abstract":"\n Aim The objective of this study includes the NEMA (National Electrical Manufacturer Association) NU2–2018 performance evaluation of the uMIvista PET-CT (positron emission tomography-computed tomography) system.\n Methods The latest NEMA NU2–2018 guidelines have been followed for the evaluation of performance parameters of this PET-CT scanner: axial, tangential, and radial spatial resolution, sensitivity, counting losses, scatter, randomness, random and counting loss correction, image quality, time and energy resolution, image uniformity, and image registration alignment post installation of country first uMIvista PET-CT.\n Results The measured NEMA sensitivity of the uMIvista PET scanner was 12.053 cps/kBq. The spatial resolutions of the PET were measured as tangential, radial, and transaxial spatial resolutions at 10 mm, with 3.01 mm, 2.95 mm, and 2.93 mm, respectively; at 100 mm, with 3.17 mm, 3.42 mm, and 3.05 mm, respectively; and at 200 mm, with 3.65 mm, 4.54 mm, and 3.17 mm, respectively, at full-width half-maximum (FWHM); while at full-width tenths-maximum (FWTM), the values at 10 mm were 5.79 mm, 5.57 mm, and 5.69 mm, respectively, and at 100 mm were 5.59 mm, 5.96 mm, and 5.91 mm, respectively. The measured time-of-flight (TOF) timing resolution was 302.294 ps and the measured energy resolution was 11.76% with FWHM and FWTM.\n Conclusion The NEMA NU2–2018 performances of this TOF-integrated digital PET-CT system are extremely good in all parameters.","PeriodicalId":23742,"journal":{"name":"World Journal of Nuclear Medicine","volume":null,"pages":null},"PeriodicalIF":0.6000,"publicationDate":"2024-01-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"World Journal of Nuclear Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/s-0044-1778709","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Aim The objective of this study includes the NEMA (National Electrical Manufacturer Association) NU2–2018 performance evaluation of the uMIvista PET-CT (positron emission tomography-computed tomography) system. Methods The latest NEMA NU2–2018 guidelines have been followed for the evaluation of performance parameters of this PET-CT scanner: axial, tangential, and radial spatial resolution, sensitivity, counting losses, scatter, randomness, random and counting loss correction, image quality, time and energy resolution, image uniformity, and image registration alignment post installation of country first uMIvista PET-CT. Results The measured NEMA sensitivity of the uMIvista PET scanner was 12.053 cps/kBq. The spatial resolutions of the PET were measured as tangential, radial, and transaxial spatial resolutions at 10 mm, with 3.01 mm, 2.95 mm, and 2.93 mm, respectively; at 100 mm, with 3.17 mm, 3.42 mm, and 3.05 mm, respectively; and at 200 mm, with 3.65 mm, 4.54 mm, and 3.17 mm, respectively, at full-width half-maximum (FWHM); while at full-width tenths-maximum (FWTM), the values at 10 mm were 5.79 mm, 5.57 mm, and 5.69 mm, respectively, and at 100 mm were 5.59 mm, 5.96 mm, and 5.91 mm, respectively. The measured time-of-flight (TOF) timing resolution was 302.294 ps and the measured energy resolution was 11.76% with FWHM and FWTM. Conclusion The NEMA NU2–2018 performances of this TOF-integrated digital PET-CT system are extremely good in all parameters.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
技术 NEMA NU2-2018 飞行时间集成数字 PET-CT 系统的性能评估
目的 本研究的目的包括对 uMIvista PET-CT(正电子发射计算机断层扫描)系统进行 NEMA(美国国家电气制造商协会)NU2-2018 性能评估。方法 根据最新的 NEMA NU2-2018 准则对该 PET-CT 扫描仪的性能参数进行评估:轴向、切向和径向空间分辨率、灵敏度、计数损失、散射、随机性、随机和计数损失校正、图像质量、时间和能量分辨率、图像均匀性以及国家首台 uMIvista PET-CT 安装后的图像配准。结果 uMIvista PET 扫描仪的 NEMA 灵敏度为 12.053 cps/kBq。正电子发射计算机断层显像的空间分辨率在 10 毫米处分别为切向空间分辨率 3.01 毫米、径向空间分辨率 2.95 毫米和横轴空间分辨率 2.93 毫米;在 100 毫米处分别为切向空间分辨率 3.17 毫米、径向空间分辨率 3.42 毫米和横轴空间分辨率 3.05 毫米;在 200 毫米处分别为切向空间分辨率 3.65 毫米、径向空间分辨率 4.54 毫米和横轴空间分辨率 3.05 毫米。在全宽半极大值(FWHM)下,10 毫米处的值分别为 5.79 毫米、5.57 毫米和 5.69 毫米,100 毫米处的值分别为 5.59 毫米、5.96 毫米和 5.91 毫米。测量的飞行时间 (TOF) 时间分辨率为 302.294 ps,测量的能量分辨率为 11.76%(FWHM 和 FWTM)。结论 NEMA NU2-2018 集成 TOF 的数字 PET-CT 系统在所有参数方面都表现出色。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
World Journal of Nuclear Medicine
World Journal of Nuclear Medicine RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
自引率
16.70%
发文量
118
审稿时长
48 weeks
期刊最新文献
Portal Venous Tumor Thrombosis and Visceral Organ Metastasis without Skeletal Involvement in mCRPC: Adverse Prognostic Indicators on Dual Tracer PET/CT and Clinical Outcome after 177Lu-PSMA-617 PRLT and Cabazitaxel Therapy Aggressive Metastatic Insulinoma in a Patient of Diabetes Mellitus with Documentation on Dual-Tracer PET-CT ([68Ga]Ga-DOTATATE and [18F]FDG): Clinical Benefits with Combined Chemo-PRRT Approach Trifecta of Tumors: Simultaneous Detection of Three Primary Malignancies by Different Radiotracers of Nuclear Medicine Extensive Multiorgan and Neurological Involvement on F-18 FDG PET-CT in a Case of Rosai-Dorfman Disease Role of Textural Analysis Parameters Derived from FDG PET/CT in Diagnosing Cardiac Sarcoidosis
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1