Ionizing radiation exposure in complex percutaneous coronary intervention: Defining local diagnostic reference levels in the catheterization laboratory

IF 1.6 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS Revista Portuguesa De Cardiologia Pub Date : 2024-07-01 DOI:10.1016/j.repc.2023.10.010
Hugo Costa , Miguel Espírito-Santo , João Bispo , João Guedes , Graciano Paulo , Jorge Mimoso , Hugo Palmeiro , Rui Baptista Gonçalves , Hugo Vinhas
{"title":"Ionizing radiation exposure in complex percutaneous coronary intervention: Defining local diagnostic reference levels in the catheterization laboratory","authors":"Hugo Costa ,&nbsp;Miguel Espírito-Santo ,&nbsp;João Bispo ,&nbsp;João Guedes ,&nbsp;Graciano Paulo ,&nbsp;Jorge Mimoso ,&nbsp;Hugo Palmeiro ,&nbsp;Rui Baptista Gonçalves ,&nbsp;Hugo Vinhas","doi":"10.1016/j.repc.2023.10.010","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction and objectives</h3><p>Concerns regarding the consequences of ionizing radiation (IR) have been increasing in the field of interventional cardiology (IC). There is little information on reported national and local radiation diagnostic reference levels (DRLs) in catheterization laboratories in Portugal. This study was designed to assess the IR dose exposure during complex percutaneous coronary intervention (PCI), and to set the respective DRLs and future achievable doses (ADs).</p></div><div><h3>Methods</h3><p>This was a retrospective cohort study which took place between 2019 and 2020, including patients who underwent complex PCI. Complex PCI was defined as all procedures that encompass treatment of chronic total occlusions (CTO) or left main coronary artery. DRLs were defined as the 75th percentile of the distribution of the median values of air kerma area product (<em>P</em><sub>KA</sub>) and cumulative air kerma (<em>K</em><sub>a,r</sub>). ADs were set at the 50th percentile of the study dose distribution. Multivariate analysis was performed using linear regression to identify predictors significantly associated with radiation dose (<em>K</em><sub>a,r</sub>).</p></div><div><h3>Results</h3><p>A total of 242 patients were included in the analysis. Most patients underwent a CTO procedure (146, 60.3%). Patients were aged 67.9±11.2 years and mostly male (81.4%). DRLs were set in <em>K</em><sub>a,r</sub> (3012 mGy) and <em>P</em><sub>KA</sub> (162 Gy cm<sup>2</sup>) for complex PCI. ADs were also set in <em>K</em><sub>a,r</sub> (1917 mGy) and <em>P</em><sub>KA</sub> (101 Gy cm<sup>2</sup>). Independent predictors of <em>K</em><sub>a,r</sub> with a positive correlation were <em>P</em><sub>KA</sub> (0.893, p&lt;0.001), fluoroscopy time (0.520, p&lt;0.001) and PCI time (0.521, p&lt;0.001).</p></div><div><h3>Conclusions</h3><p>This study reports the results of IR in complex PCI. DRLs were set for IR dose exposure measured in <em>K</em><sub>a,r</sub> (3012 mGy) and <em>P</em><sub>KA</sub> (162 Gy cm<sup>2</sup>). ADs, values to be achieved in future assessment, were set to <em>K</em><sub>a,r</sub> (1917 mGy) and <em>P</em><sub>KA</sub> (101 Gy cm<sup>2</sup>).</p></div>","PeriodicalId":48985,"journal":{"name":"Revista Portuguesa De Cardiologia","volume":null,"pages":null},"PeriodicalIF":1.6000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0870255124000143/pdfft?md5=7a111485a806376712a0c3b62682cc2a&pid=1-s2.0-S0870255124000143-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista Portuguesa De Cardiologia","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0870255124000143","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction and objectives

Concerns regarding the consequences of ionizing radiation (IR) have been increasing in the field of interventional cardiology (IC). There is little information on reported national and local radiation diagnostic reference levels (DRLs) in catheterization laboratories in Portugal. This study was designed to assess the IR dose exposure during complex percutaneous coronary intervention (PCI), and to set the respective DRLs and future achievable doses (ADs).

Methods

This was a retrospective cohort study which took place between 2019 and 2020, including patients who underwent complex PCI. Complex PCI was defined as all procedures that encompass treatment of chronic total occlusions (CTO) or left main coronary artery. DRLs were defined as the 75th percentile of the distribution of the median values of air kerma area product (PKA) and cumulative air kerma (Ka,r). ADs were set at the 50th percentile of the study dose distribution. Multivariate analysis was performed using linear regression to identify predictors significantly associated with radiation dose (Ka,r).

Results

A total of 242 patients were included in the analysis. Most patients underwent a CTO procedure (146, 60.3%). Patients were aged 67.9±11.2 years and mostly male (81.4%). DRLs were set in Ka,r (3012 mGy) and PKA (162 Gy cm2) for complex PCI. ADs were also set in Ka,r (1917 mGy) and PKA (101 Gy cm2). Independent predictors of Ka,r with a positive correlation were PKA (0.893, p<0.001), fluoroscopy time (0.520, p<0.001) and PCI time (0.521, p<0.001).

Conclusions

This study reports the results of IR in complex PCI. DRLs were set for IR dose exposure measured in Ka,r (3012 mGy) and PKA (162 Gy cm2). ADs, values to be achieved in future assessment, were set to Ka,r (1917 mGy) and PKA (101 Gy cm2).

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
复杂经皮冠状动脉介入治疗中的电离辐射暴露:确定导管室的本地诊断参考水平。
导言和目标:介入心脏病学(IC)领域对电离辐射(IR)后果的担忧与日俱增。有关葡萄牙导管室的国家和地方辐射诊断参考水平(DRLs)的报告信息很少。本研究旨在评估复杂的经皮冠状动脉介入治疗(PCI)过程中的红外辐射剂量,并设定相应的诊断参考水平和未来的可达到剂量(ADs):这是一项回顾性队列研究,在 2019 年至 2020 年期间进行,包括接受复杂 PCI 的患者。复杂 PCI 被定义为包括治疗慢性全闭塞(CTO)或左冠状动脉主干的所有手术。DRL 被定义为空气切迹面积乘积 (PKA) 和累积空气切迹 (Ka,r) 中位值分布的第 75 百分位数。AD 设置为研究剂量分布的第 50 个百分位数。使用线性回归法进行多变量分析,以确定与辐射剂量(Ka,r)显著相关的预测因素:共有 242 名患者纳入分析。大多数患者接受了 CTO 手术(146 例,60.3%)。患者年龄为(67.9±11.2)岁,大部分为男性(81.4%)。复杂 PCI 的 DRL 设置为 Ka,r(3012 mGy)和 PKA(162 Gy.Cm2)。AD也被设定为Ka,r(1917 mGy)和PKA(101 Gy.Cm2)。与 Ka,r 呈正相关的独立预测因子是 PKA(0.893,p 结论:本研究报告了红外线在复杂 PCI 中的应用结果。根据Ka,r(3012 mGy)和PKA(162 Gy.Cm2)测量的红外剂量暴露设定了DRL。AD是未来评估中要达到的值,被设定为Ka,r(1917 mGy)和PKA(101 Gy.Cm2)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Revista Portuguesa De Cardiologia
Revista Portuguesa De Cardiologia CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
2.70
自引率
22.20%
发文量
205
审稿时长
54 days
期刊介绍: The Portuguese Journal of Cardiology, the official journal of the Portuguese Society of Cardiology, was founded in 1982 with the aim of keeping Portuguese cardiologists informed through the publication of scientific articles on areas such as arrhythmology and electrophysiology, cardiovascular surgery, intensive care, coronary artery disease, cardiovascular imaging, hypertension, heart failure and cardiovascular prevention. The Journal is a monthly publication with high standards of quality in terms of scientific content and production. Since 1999 it has been published in English as well as Portuguese, which has widened its readership abroad. It is distributed to all members of the Portuguese Societies of Cardiology, Internal Medicine, Pneumology and Cardiothoracic Surgery, as well as to leading non-Portuguese cardiologists and to virtually all cardiology societies worldwide. It has been referred in Medline since 1987.
期刊最新文献
Cardiovascular benefits of resistance exercise: It's time to prescribe. Cardio-oncology guidelines, structural heart disease and Kounis syndrome in the upcoming guidelines. Response to the letter "Cardio-oncology guidelines, structural heart disease and Kounis syndrome in the upcoming guidelines". Cardiac arrest: It is important not just to survive, but to survive with at least an acceptable quality of life. Diagnostic prediction rules in acute pulmonary embolism: Is it acceptable to compromise safety?
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1