Bilateral comparison of 1.018 V and 10 V standards between INRIM (Italy) and the BIPM, November to December 2023 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b)
S Solve, R Chayramy, M Stock, P Durandetto and E Enrico
{"title":"Bilateral comparison of 1.018 V and 10 V standards between INRIM (Italy) and the BIPM, November to December 2023 (part of the ongoing BIPM key comparison BIPM.EM-K11.a and b)","authors":"S Solve, R Chayramy, M Stock, P Durandetto and E Enrico","doi":"10.1088/0026-1394/61/1a/01004","DOIUrl":null,"url":null,"abstract":"As part of the ongoing BIPM key comparison BIPM.EM-K11.a and b, a comparison of the 1.018 V and 10 V voltage reference standards of the BIPM and the INRIM (Italy), from November to December 2023. Two BIPM Zener diode-based travelling standards (Fluke 732B), BIPM_7 (Z7) and BIPM_9 (Z9), were transported by freight to the INRIM and back to the BIPM. In order to keep the Zeners powered during their transportation phase, two additional external batteries were connected in parallel to the internal battery. At the INRIM and the BIPM, the reference standard for DC voltage is a Programmable Josephson Voltage Standard (PJVS). The output electromotive force (EMF) of each travelling standard was measured by direct comparison with the primary standard. The final result of the comparison is presented as the difference between the values assigned to DC voltage standards by the INRIM, at the level of 1.018 V and 10 V, at the INRIM, UINRIM, and those assigned by the BIPM, at the BIPM, UBIPM, at the reference date of the 1st of December 2023. UINRIM - UBIPM = 0.06 V; uc = 0.05 V, at 1.018 V UINRIM - UBIPM = 0.22 V; uc = 0.12 V, at 10 V where uc is the combined standard uncertainty associated with the measured difference, including the uncertainty of the realisation of the volt at the BIPM and at the INRIM, and the uncertainty related to the comparison. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":"24 1","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-05-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/0026-1394/61/1a/01004","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
As part of the ongoing BIPM key comparison BIPM.EM-K11.a and b, a comparison of the 1.018 V and 10 V voltage reference standards of the BIPM and the INRIM (Italy), from November to December 2023. Two BIPM Zener diode-based travelling standards (Fluke 732B), BIPM_7 (Z7) and BIPM_9 (Z9), were transported by freight to the INRIM and back to the BIPM. In order to keep the Zeners powered during their transportation phase, two additional external batteries were connected in parallel to the internal battery. At the INRIM and the BIPM, the reference standard for DC voltage is a Programmable Josephson Voltage Standard (PJVS). The output electromotive force (EMF) of each travelling standard was measured by direct comparison with the primary standard. The final result of the comparison is presented as the difference between the values assigned to DC voltage standards by the INRIM, at the level of 1.018 V and 10 V, at the INRIM, UINRIM, and those assigned by the BIPM, at the BIPM, UBIPM, at the reference date of the 1st of December 2023. UINRIM - UBIPM = 0.06 V; uc = 0.05 V, at 1.018 V UINRIM - UBIPM = 0.22 V; uc = 0.12 V, at 10 V where uc is the combined standard uncertainty associated with the measured difference, including the uncertainty of the realisation of the volt at the BIPM and at the INRIM, and the uncertainty related to the comparison. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/. The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
期刊介绍:
Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them.
The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics.
In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.