Diluted contrast media combined with high-resolution C-arm CT for evaluating apposition of flow diversions in a porcine model.

IF 1.5 4区 医学 Q4 CLINICAL NEUROLOGY Interventional Neuroradiology Pub Date : 2024-06-17 DOI:10.1177/15910199241261763
Shuhai Long, Sun Yu, Chengcheng Shi, Ji Ma, Zhen Li, Jingliang Cheng, Yuncai Ran, Xinwei Han, Tengfei Li
{"title":"Diluted contrast media combined with high-resolution C-arm CT for evaluating apposition of flow diversions in a porcine model.","authors":"Shuhai Long, Sun Yu, Chengcheng Shi, Ji Ma, Zhen Li, Jingliang Cheng, Yuncai Ran, Xinwei Han, Tengfei Li","doi":"10.1177/15910199241261763","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>In intracranial stenting, good stent apposition is crucial, and high-resolution C-arm computer tomography (CT) is utilized to assess whether stent apposition is complete. This study was aimed at finding optimal hypertonic or isotonic contrast media injection concentration for high-resolution C-arm CT to assess apposition of flow diversion (FD) after carotid artery stenting in swine.</p><p><strong>Methods: </strong>Twelve FD stents were implanted into the left carotid artery of Bama swine through the endovascular method. During high-resolution C-arm CT scanning, 6%, 8%, 10%, 12%, and 14% dilution percentages of hypertonic contrast media (iopromide 370 mg/ml) and 10%, 12%, 14%, 16%, and 18% dilution percentages of isotonic contrast media (iodixanol 320 mg/ml) were separately injected. A radiologist and a neuro-interventional specialist evaluated and qualitatively scored the post-processed images, and intravascular ultrasound (IVUS) was used to verify the accuracy of these images.</p><p><strong>Results: </strong>Overall, 12 FD stents were implanted into the left common carotid artery of 12 swine, with a technical success rate of 100%. The best reconstructed images used to observe stent apposition were achieved with iopromide diluted to a concentration of 12% (all <i>P</i> < .05) or iodixanol diluted to a concentration of 16% (all <i>P</i> < .05). Malapposition was noted in one case, and good apposition was noted in 11 cases. These results were consistent with IVUS observations.</p><p><strong>Conclusion: </strong>Injecting iopromide or iodixanol diluted to 12% or 16% during high-resolution C-arm CT scanning, respectively, can help evaluate FD stent apposition and obtain optimal image quality.</p>","PeriodicalId":49174,"journal":{"name":"Interventional Neuroradiology","volume":" ","pages":"15910199241261763"},"PeriodicalIF":1.5000,"publicationDate":"2024-06-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11571160/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Interventional Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15910199241261763","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: In intracranial stenting, good stent apposition is crucial, and high-resolution C-arm computer tomography (CT) is utilized to assess whether stent apposition is complete. This study was aimed at finding optimal hypertonic or isotonic contrast media injection concentration for high-resolution C-arm CT to assess apposition of flow diversion (FD) after carotid artery stenting in swine.

Methods: Twelve FD stents were implanted into the left carotid artery of Bama swine through the endovascular method. During high-resolution C-arm CT scanning, 6%, 8%, 10%, 12%, and 14% dilution percentages of hypertonic contrast media (iopromide 370 mg/ml) and 10%, 12%, 14%, 16%, and 18% dilution percentages of isotonic contrast media (iodixanol 320 mg/ml) were separately injected. A radiologist and a neuro-interventional specialist evaluated and qualitatively scored the post-processed images, and intravascular ultrasound (IVUS) was used to verify the accuracy of these images.

Results: Overall, 12 FD stents were implanted into the left common carotid artery of 12 swine, with a technical success rate of 100%. The best reconstructed images used to observe stent apposition were achieved with iopromide diluted to a concentration of 12% (all P < .05) or iodixanol diluted to a concentration of 16% (all P < .05). Malapposition was noted in one case, and good apposition was noted in 11 cases. These results were consistent with IVUS observations.

Conclusion: Injecting iopromide or iodixanol diluted to 12% or 16% during high-resolution C-arm CT scanning, respectively, can help evaluate FD stent apposition and obtain optimal image quality.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
稀释造影剂结合高分辨率 C 型臂 CT 评估猪模型中血流分流的位置。
背景:在颅内支架植入术中,良好的支架贴合至关重要,而高分辨率 C 臂计算机断层扫描(CT)可用于评估支架是否完全贴合。本研究旨在为高分辨率 C 型臂 CT 寻找最佳的高渗或等渗造影剂注射浓度,以评估猪颈动脉支架术后血流分流(FD)的固定情况:方法:通过血管内方法将 12 个 FD 支架植入巴马猪的左颈动脉。在进行高分辨率 C 型臂 CT 扫描时,分别注入稀释百分比为 6%、8%、10%、12% 和 14%的高渗造影剂(碘普罗米370 毫克/毫升)和稀释百分比为 10%、12%、14%、16% 和 18%的等渗造影剂(碘克沙醇 320 毫克/毫升)。一名放射科医生和一名神经介入专家对处理后的图像进行了评估和定性评分,并使用血管内超声(IVUS)验证了这些图像的准确性:总共在 12 头猪的左侧颈总动脉中植入了 12 个 FD 支架,技术成功率为 100%。用于观察支架贴合情况的最佳重建图像是用稀释至 12% 浓度的碘普罗米特绘制的(均为 P P 结论):在高分辨率 C 型臂 CT 扫描过程中注射分别稀释至 12% 或 16% 的碘普罗米特或碘克沙醇有助于评估 FD 支架的位置并获得最佳图像质量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Interventional Neuroradiology
Interventional Neuroradiology CLINICAL NEUROLOGY-RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
CiteScore
3.60
自引率
11.80%
发文量
192
审稿时长
6-12 weeks
期刊介绍: Interventional Neuroradiology (INR) is a peer-reviewed clinical practice journal documenting the current state of interventional neuroradiology worldwide. INR publishes original clinical observations, descriptions of new techniques or procedures, case reports, and articles on the ethical and social aspects of related health care. Original research published in INR is related to the practice of interventional neuroradiology...
期刊最新文献
Circulating miRNA profiles as predictive biomarkers for aneurysm healing following endovascular treatment: a prospective study. "Chopperlysis": The effect of helicopter transport on reperfusion and outcomes in large vessel occlusion strokes. DWI lesions after intracranial aneurysm treatment with contour or WEB-does the device matter? Analysis of selective neurocritical care admission costs following elective endovascular treatment of unruptured intracranial aneurysms. Recognition of dural to pial supply in high-grade dural arteriovenous fistula: A technical note.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1