{"title":"Diluted contrast media combined with high-resolution C-arm CT for evaluating apposition of flow diversions in a porcine model.","authors":"Shuhai Long, Sun Yu, Chengcheng Shi, Ji Ma, Zhen Li, Jingliang Cheng, Yuncai Ran, Xinwei Han, Tengfei Li","doi":"10.1177/15910199241261763","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>In intracranial stenting, good stent apposition is crucial, and high-resolution C-arm computer tomography (CT) is utilized to assess whether stent apposition is complete. This study was aimed at finding optimal hypertonic or isotonic contrast media injection concentration for high-resolution C-arm CT to assess apposition of flow diversion (FD) after carotid artery stenting in swine.</p><p><strong>Methods: </strong>Twelve FD stents were implanted into the left carotid artery of Bama swine through the endovascular method. During high-resolution C-arm CT scanning, 6%, 8%, 10%, 12%, and 14% dilution percentages of hypertonic contrast media (iopromide 370 mg/ml) and 10%, 12%, 14%, 16%, and 18% dilution percentages of isotonic contrast media (iodixanol 320 mg/ml) were separately injected. A radiologist and a neuro-interventional specialist evaluated and qualitatively scored the post-processed images, and intravascular ultrasound (IVUS) was used to verify the accuracy of these images.</p><p><strong>Results: </strong>Overall, 12 FD stents were implanted into the left common carotid artery of 12 swine, with a technical success rate of 100%. The best reconstructed images used to observe stent apposition were achieved with iopromide diluted to a concentration of 12% (all <i>P</i> < .05) or iodixanol diluted to a concentration of 16% (all <i>P</i> < .05). Malapposition was noted in one case, and good apposition was noted in 11 cases. These results were consistent with IVUS observations.</p><p><strong>Conclusion: </strong>Injecting iopromide or iodixanol diluted to 12% or 16% during high-resolution C-arm CT scanning, respectively, can help evaluate FD stent apposition and obtain optimal image quality.</p>","PeriodicalId":49174,"journal":{"name":"Interventional Neuroradiology","volume":" ","pages":"15910199241261763"},"PeriodicalIF":1.5000,"publicationDate":"2024-06-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11571160/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Interventional Neuroradiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15910199241261763","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: In intracranial stenting, good stent apposition is crucial, and high-resolution C-arm computer tomography (CT) is utilized to assess whether stent apposition is complete. This study was aimed at finding optimal hypertonic or isotonic contrast media injection concentration for high-resolution C-arm CT to assess apposition of flow diversion (FD) after carotid artery stenting in swine.
Methods: Twelve FD stents were implanted into the left carotid artery of Bama swine through the endovascular method. During high-resolution C-arm CT scanning, 6%, 8%, 10%, 12%, and 14% dilution percentages of hypertonic contrast media (iopromide 370 mg/ml) and 10%, 12%, 14%, 16%, and 18% dilution percentages of isotonic contrast media (iodixanol 320 mg/ml) were separately injected. A radiologist and a neuro-interventional specialist evaluated and qualitatively scored the post-processed images, and intravascular ultrasound (IVUS) was used to verify the accuracy of these images.
Results: Overall, 12 FD stents were implanted into the left common carotid artery of 12 swine, with a technical success rate of 100%. The best reconstructed images used to observe stent apposition were achieved with iopromide diluted to a concentration of 12% (all P < .05) or iodixanol diluted to a concentration of 16% (all P < .05). Malapposition was noted in one case, and good apposition was noted in 11 cases. These results were consistent with IVUS observations.
Conclusion: Injecting iopromide or iodixanol diluted to 12% or 16% during high-resolution C-arm CT scanning, respectively, can help evaluate FD stent apposition and obtain optimal image quality.
期刊介绍:
Interventional Neuroradiology (INR) is a peer-reviewed clinical practice journal documenting the current state of interventional neuroradiology worldwide. INR publishes original clinical observations, descriptions of new techniques or procedures, case reports, and articles on the ethical and social aspects of related health care. Original research published in INR is related to the practice of interventional neuroradiology...