A. Masumoto, Hiroyuki Yamamoto, Nobuyuki Takahashi, Tomofumi Takaya
{"title":"Balloon-expandable transcatheter heart valve deformation: A rare complication during transcatheter aortic valve replacement","authors":"A. Masumoto, Hiroyuki Yamamoto, Nobuyuki Takahashi, Tomofumi Takaya","doi":"10.1093/ehjcr/ytae285","DOIUrl":null,"url":null,"abstract":"\n An 80-year-old female underwent transfemoral transcatheter aortic valve replacement, during which an outward bending of the transcatheter heart valve’s stent frame occurred. Because of the difficulty in fixing the deformation by balloon angioplasty, the valve delivered with balloon cushion technique was slowly deployed with underfilling, resulting in no annulus rupture.","PeriodicalId":507701,"journal":{"name":"European Heart Journal - Case Reports","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-06-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Heart Journal - Case Reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/ehjcr/ytae285","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
An 80-year-old female underwent transfemoral transcatheter aortic valve replacement, during which an outward bending of the transcatheter heart valve’s stent frame occurred. Because of the difficulty in fixing the deformation by balloon angioplasty, the valve delivered with balloon cushion technique was slowly deployed with underfilling, resulting in no annulus rupture.