Alejandro Tello, Virgilio Galvis, Samuel Arba-Mosquera, Ruby Morales, Valeria Otoya, Sylvia J Villamizar, Sergio E Serrano
{"title":"Clinical History Method versus Corneal Tomographers in Estimating Corneal Power after Photorefractive Surgery","authors":"Alejandro Tello, Virgilio Galvis, Samuel Arba-Mosquera, Ruby Morales, Valeria Otoya, Sylvia J Villamizar, Sergio E Serrano","doi":"10.31348/2024/23","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>To investigate the concordance between the corneal power determined by various approaches with two tomographers (MS-39® and Galilei G6®) and the clinical history method (CHM) in patients undergoing photorefractive surgery with excimer laser for myopic errors.</p><p><strong>Material and methods: </strong>Prospective cohort study. Patients undergoing keratorefractive surgery, and having pre- and postoperative keratometries, and tomographies, were included.</p><p><strong>Results: </strong>In 90 eyes, the differences in the power estimated by the CHM and the one determined by four approaches with the corneal tomographers, which included measurements of the posterior cornea, did not show statistically significant differences in their averages. However, the 95% limits of agreement were very wide. After obtaining regression formulas to adjust the values of these four variables, the results of the agreement analysis were similar.</p><p><strong>Conclusion: </strong>Although certain values either directly determined or derived from measurements with the Galilei® and MS-39®corneal tomographers, approximated the estimated value of postoperative corneal power according to the CHM, due to the amplitude of their limits of agreement, these calculations must be taken with care, because they may not be accurate in a given eye.</p>","PeriodicalId":39839,"journal":{"name":"Ceska a Slovenska Oftalmologie","volume":"80 Ahead of print","pages":"1-12"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ceska a Slovenska Oftalmologie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31348/2024/23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Aims: To investigate the concordance between the corneal power determined by various approaches with two tomographers (MS-39® and Galilei G6®) and the clinical history method (CHM) in patients undergoing photorefractive surgery with excimer laser for myopic errors.
Material and methods: Prospective cohort study. Patients undergoing keratorefractive surgery, and having pre- and postoperative keratometries, and tomographies, were included.
Results: In 90 eyes, the differences in the power estimated by the CHM and the one determined by four approaches with the corneal tomographers, which included measurements of the posterior cornea, did not show statistically significant differences in their averages. However, the 95% limits of agreement were very wide. After obtaining regression formulas to adjust the values of these four variables, the results of the agreement analysis were similar.
Conclusion: Although certain values either directly determined or derived from measurements with the Galilei® and MS-39®corneal tomographers, approximated the estimated value of postoperative corneal power according to the CHM, due to the amplitude of their limits of agreement, these calculations must be taken with care, because they may not be accurate in a given eye.
期刊介绍:
- Redakce přijímá pouze práce vyhovující po odborné stránce, které jsou na odpovídající profesionální a formální úrovni. - Uveřejněná práce se stává majetkem časopisu, přetisknout její část nebo obrázek lze jen s citací původu. - Rukopis zasílejte v originále a dobře čitelné kopii (je nutná také kopie tabulek, legend, podle možnosti i obrázků). - Listy číslujte v pravém horním rohu a spojujte svorkou, nesešívejte. Přijímáme práce psané na jedné straně kvalitního bílého nelesklého papíru formátu A4 (neprůklepový) na psacím stroji nebo počítači s obvyklými typy.