{"title":"Successful Bronchoscopic Lung Volume Reduction with New Lung Tensioning Device Coil: A Case Report.","authors":"Jonas Herth, Jasmin Wani, Daniel Franzen","doi":"10.1159/000540286","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Bronchoscopic lung volume reduction (BLVR) using endobronchial coil treatment is a widely studied therapeutic option in patients with end-stage chronic obstructive pulmonary disease (COPD) and pulmonary emphysema. However, patient responses were inconsistent, and, from 2020, production discontinuation rendered the treatment unavailable. In the meantime, a next-generation lung tensioning coil (FreeFlow Coil 4; FreeFlow Medical, Inc., Fremont, CA, USA) has been developed by the inventor of the lung volume reduction coil implant technology. This case study presents the first documented successful BLVR using FreeFlow Coils 4.</p><p><strong>Case presentation: </strong>A 68-year-old male patient with COPD, classified as Global Initiative for Chronic Obstructive Lung Disease (GOLD) stage 3, and homogenous emphysema with incomplete interlobar fissures was treated using the new developed FreeFlow Coil 4 in both upper lobes. Apart from a suspected coil-associated opacity, no adverse events occurred. At the 18-week follow-up, a significant improvement in lung function and quality of life was observed, as measured by forced expiratory volume in 1 s (+150 mL), residual volume (-0.50 L), 6-min walking distance (+75 m), and the total score of the St. George's Respiratory Questionnaire (SGRQ) (-35 points).</p><p><strong>Conclusion: </strong>BLVR with FreeFlow Coils 4 has shown promising outcomes, significantly improving both pulmonary function and quality of life in 1 patient. However, larger studies and randomized controlled trials are imperative to investigate the efficacy and safety profile of the new coil system.</p>","PeriodicalId":21048,"journal":{"name":"Respiration","volume":" ","pages":"72-76"},"PeriodicalIF":3.5000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Respiration","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1159/000540286","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/24 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Bronchoscopic lung volume reduction (BLVR) using endobronchial coil treatment is a widely studied therapeutic option in patients with end-stage chronic obstructive pulmonary disease (COPD) and pulmonary emphysema. However, patient responses were inconsistent, and, from 2020, production discontinuation rendered the treatment unavailable. In the meantime, a next-generation lung tensioning coil (FreeFlow Coil 4; FreeFlow Medical, Inc., Fremont, CA, USA) has been developed by the inventor of the lung volume reduction coil implant technology. This case study presents the first documented successful BLVR using FreeFlow Coils 4.
Case presentation: A 68-year-old male patient with COPD, classified as Global Initiative for Chronic Obstructive Lung Disease (GOLD) stage 3, and homogenous emphysema with incomplete interlobar fissures was treated using the new developed FreeFlow Coil 4 in both upper lobes. Apart from a suspected coil-associated opacity, no adverse events occurred. At the 18-week follow-up, a significant improvement in lung function and quality of life was observed, as measured by forced expiratory volume in 1 s (+150 mL), residual volume (-0.50 L), 6-min walking distance (+75 m), and the total score of the St. George's Respiratory Questionnaire (SGRQ) (-35 points).
Conclusion: BLVR with FreeFlow Coils 4 has shown promising outcomes, significantly improving both pulmonary function and quality of life in 1 patient. However, larger studies and randomized controlled trials are imperative to investigate the efficacy and safety profile of the new coil system.
期刊介绍:
''Respiration'' brings together the results of both clinical and experimental investigations on all aspects of the respiratory system in health and disease. Clinical improvements in the diagnosis and treatment of chest and lung diseases are covered, as are the latest findings in physiology, biochemistry, pathology, immunology and pharmacology. The journal includes classic features such as editorials that accompany original articles in clinical and basic science research, reviews and letters to the editor. Further sections are: Technical Notes, The Eye Catcher, What’s Your Diagnosis?, The Opinion Corner, New Drugs in Respiratory Medicine, New Insights from Clinical Practice and Guidelines. ''Respiration'' is the official journal of the Swiss Society for Pneumology (SGP) and also home to the European Association for Bronchology and Interventional Pulmonology (EABIP), which occupies a dedicated section on Interventional Pulmonology in the journal. This modern mix of different features and a stringent peer-review process by a dedicated editorial board make ''Respiration'' a complete guide to progress in thoracic medicine.