A comparison of white-to-white measurements obtained by Anterion AS-OCT versus three optical devices in refractive surgery candidates.

IF 1.4 4区 医学 Q3 OPHTHALMOLOGY European Journal of Ophthalmology Pub Date : 2024-09-02 DOI:10.1177/11206721241281876
Ramin Salouti, Ali Azizi, Hesam Hashemian, Masoud Khorrami-Nejad, Qassim Alkhafaji, Kia Salouti, Maryam Ghoreyshi, M Hossein Nowroozzadeh
{"title":"A comparison of white-to-white measurements obtained by Anterion AS-OCT versus three optical devices in refractive surgery candidates.","authors":"Ramin Salouti, Ali Azizi, Hesam Hashemian, Masoud Khorrami-Nejad, Qassim Alkhafaji, Kia Salouti, Maryam Ghoreyshi, M Hossein Nowroozzadeh","doi":"10.1177/11206721241281876","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and objectives: </strong>To assess the agreement between Anterion AS-OCT and three optical devices in measuring the white-to-white (WTW) diameter in candidates for refractive surgery.</p><p><strong>Methods: </strong>In this cross-sectional prospective study, 129 right eyes of 129 normal individuals underwent consecutive scans with the Anterion, the Pentacam AXL, the IOLMaster 700, and the Orbscan IIz. Mean difference (MD), 95% limits of agreement (LoA), and concordance correlation coefficient (CCC) were calculated to assess agreement and interchangeability.</p><p><strong>Results: </strong>The mean age of participants was 30.4 ± 5.9 (range: 21-47) years. The mean WTW distance measured by the different instruments was as follows: 12.00 ± 0.42 mm (range, 10.57 to 13.05) using the Anterion, 11.87 ± 0.34 (11.10 to 12.80) using the Pentacam, 12.12 ± 0.44 (11.00 to 13.30) using the IOLMaster, and 11.73 ± 0.37 (11.00 to 13.10) using the Orbscan. The MD and 95% LoA for Anterion vs. Pentacam, IOLMaster, or Orbscan were [0.11; -0.31 to 0.54 mm], [-0.13; -0.93 to 0.66], and [0.25; -0.28 to 0.78], respectively. The corresponding CCCs were 0.803, 0.514, and 0.631.</p><p><strong>Conclusions: </strong>This study found weak agreement between Anterion and Pentacam AXL, IOLMaster 700, and Orbscan IIz devices regarding WTW distance measurements in refractive surgery candidates. Therefore, it is not recommended to use Anterion's WTW measurements interchangeably with the other three devices, particularly for phakic intraocular lens sizing.</p>","PeriodicalId":12000,"journal":{"name":"European Journal of Ophthalmology","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/11206721241281876","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background and objectives: To assess the agreement between Anterion AS-OCT and three optical devices in measuring the white-to-white (WTW) diameter in candidates for refractive surgery.

Methods: In this cross-sectional prospective study, 129 right eyes of 129 normal individuals underwent consecutive scans with the Anterion, the Pentacam AXL, the IOLMaster 700, and the Orbscan IIz. Mean difference (MD), 95% limits of agreement (LoA), and concordance correlation coefficient (CCC) were calculated to assess agreement and interchangeability.

Results: The mean age of participants was 30.4 ± 5.9 (range: 21-47) years. The mean WTW distance measured by the different instruments was as follows: 12.00 ± 0.42 mm (range, 10.57 to 13.05) using the Anterion, 11.87 ± 0.34 (11.10 to 12.80) using the Pentacam, 12.12 ± 0.44 (11.00 to 13.30) using the IOLMaster, and 11.73 ± 0.37 (11.00 to 13.10) using the Orbscan. The MD and 95% LoA for Anterion vs. Pentacam, IOLMaster, or Orbscan were [0.11; -0.31 to 0.54 mm], [-0.13; -0.93 to 0.66], and [0.25; -0.28 to 0.78], respectively. The corresponding CCCs were 0.803, 0.514, and 0.631.

Conclusions: This study found weak agreement between Anterion and Pentacam AXL, IOLMaster 700, and Orbscan IIz devices regarding WTW distance measurements in refractive surgery candidates. Therefore, it is not recommended to use Anterion's WTW measurements interchangeably with the other three devices, particularly for phakic intraocular lens sizing.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用 Anterion AS-OCT 和三种光学设备对屈光手术候选者进行白对白测量的比较。
背景和目的:评估 Anterion AS-OCT 和三种光学设备在测量屈光手术候选者的白内障直径方面的一致性:评估 Anterion AS-OCT 和三种光学设备在测量屈光手术候选者的白内障直径(WTW)方面的一致性:在这项横断面前瞻性研究中,129 名正常人的 129 只右眼接受了 Anterion、Pentacam AXL、IOLMaster 700 和 Orbscan IIz 的连续扫描。计算了平均差(MD)、95% 的一致性极限(LoA)和一致性相关系数(CCC),以评估一致性和互换性:参与者的平均年龄为 30.4 ± 5.9(21-47)岁。不同仪器测量的 WTW 距离平均值如下使用Anterion测量的平均WTW距离为12.00 ± 0.42 mm(范围:10.57至13.05),使用Pentacam测量的平均WTW距离为11.87 ± 0.34 mm(范围:11.10至12.80),使用IOLMaster测量的平均WTW距离为12.12 ± 0.44 mm(范围:11.00至13.30),使用Orbscan测量的平均WTW距离为11.73 ± 0.37 mm(范围:11.00至13.10)。Anterion与Pentacam、IOLMaster或Orbscan的MD和95% LoA分别为[0.11;-0.31至0.54 mm]、[-0.13;-0.93至0.66]和[0.25;-0.28至0.78]。相应的 CCC 分别为 0.803、0.514 和 0.631:本研究发现,Anterion 与 Pentacam AXL、IOLMaster 700 和 Orbscan IIz 设备在屈光手术候选者 WTW 距离测量方面的一致性较弱。因此,不建议将 Anterion 的 WTW 测量值与其他三种设备的测量值互换使用,尤其是在隐形眼镜的选型方面。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.60
自引率
0.00%
发文量
372
审稿时长
3-8 weeks
期刊介绍: The European Journal of Ophthalmology was founded in 1991 and is issued in print bi-monthly. It publishes only peer-reviewed original research reporting clinical observations and laboratory investigations with clinical relevance focusing on new diagnostic and surgical techniques, instrument and therapy updates, results of clinical trials and research findings.
期刊最新文献
Acute infectious endophthalmitis after chandelier-assisted scleral buckling surgery. Bagel sign and how to avoid it in DMEK surgery. Changes in choroidal thickness and lamina cribrosa position in subjects with overdipper pattern of arterial pressure. Characteristics of submacular hemorrhage with bacillary layer detachment and intrabacillary hemorrhage. Comparison of 3-month clinical outcomes between two enhanced monofocal intraocular lenses: A single-center prospective study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1