Flexible inner surface of polysulfone membranes prevents platelet adhesive protein adsorption and improves antithrombogenicity in vitro.

IF 1.4 4区 医学 Q4 ENGINEERING, BIOMEDICAL International Journal of Artificial Organs Pub Date : 2024-09-04 DOI:10.1177/03913988241269465
Ryo Takatsuji, Masahide Koremoto, Yoko Fujimoto, Yuko Saida, Yoshihiro Hatanaka
{"title":"Flexible inner surface of polysulfone membranes prevents platelet adhesive protein adsorption and improves antithrombogenicity in vitro.","authors":"Ryo Takatsuji, Masahide Koremoto, Yoko Fujimoto, Yuko Saida, Yoshihiro Hatanaka","doi":"10.1177/03913988241269465","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>We investigated whether the condition of the inner surface of hollow fibers affects the blood compatibility of hemodialyzers.</p><p><strong>Methods: </strong>We used scanning probe microscope/atomic force microscopy (SPM/AFM) to investigate the height of the swelling and flexible layers (thickness and softness) on the inner surfaces of the hollow fibers. Next, we tested the blood compatibility between dialyzers comprising a hollow fiber membrane, in which the other dialyzers, except for PVP, were additionally coated using PS membranes coated with other materials. After blood was injected into the dialyzer and plugged, dynamic stimulation was performed by slightly rotating the dialyzer for 4 h, although there was no blood circulation.</p><p><strong>Results: </strong>The vitamin E-coated polysulfone (PS) membrane showed a higher thickness and softness of the flexible layer than the asymmetric cellulose triacetate membrane without polyvinylpyrrolidone (PVP) and the PS membranes with PVP. We found that the dialyzer with vitamin E coating significantly suppressed the decrease in platelets, increase in β-TG, and increase in PF4 compared to those coated with NV polymer. Additionally, as the adsorbed protein on the inner surface, the total protein, fibronectin, and vWF levels were significantly lower in the vitamin E-coated dialyzer.</p><p><strong>Conclusion: </strong>The thickness and softness of the flexible layer of the inner surface of the hollow fiber membrane in vitro affect differences in blood coagulation performance in clinical research. Future clinical trials are required to confirm our results.</p>","PeriodicalId":13932,"journal":{"name":"International Journal of Artificial Organs","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-09-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Artificial Organs","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1177/03913988241269465","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Background: We investigated whether the condition of the inner surface of hollow fibers affects the blood compatibility of hemodialyzers.

Methods: We used scanning probe microscope/atomic force microscopy (SPM/AFM) to investigate the height of the swelling and flexible layers (thickness and softness) on the inner surfaces of the hollow fibers. Next, we tested the blood compatibility between dialyzers comprising a hollow fiber membrane, in which the other dialyzers, except for PVP, were additionally coated using PS membranes coated with other materials. After blood was injected into the dialyzer and plugged, dynamic stimulation was performed by slightly rotating the dialyzer for 4 h, although there was no blood circulation.

Results: The vitamin E-coated polysulfone (PS) membrane showed a higher thickness and softness of the flexible layer than the asymmetric cellulose triacetate membrane without polyvinylpyrrolidone (PVP) and the PS membranes with PVP. We found that the dialyzer with vitamin E coating significantly suppressed the decrease in platelets, increase in β-TG, and increase in PF4 compared to those coated with NV polymer. Additionally, as the adsorbed protein on the inner surface, the total protein, fibronectin, and vWF levels were significantly lower in the vitamin E-coated dialyzer.

Conclusion: The thickness and softness of the flexible layer of the inner surface of the hollow fiber membrane in vitro affect differences in blood coagulation performance in clinical research. Future clinical trials are required to confirm our results.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
聚砜膜的柔性内表面可防止血小板粘附蛋白吸附,提高体外抗血栓形成能力。
背景:我们研究了中空纤维内表面的状况是否会影响血液透析器的血液相容性:我们研究了中空纤维内表面的状况是否会影响血液透析器的血液相容性:我们使用扫描探针显微镜/原子力显微镜(SPM/AFM)研究了中空纤维内表面膨胀层和柔性层的高度(厚度和柔软度)。接下来,我们测试了由中空纤维膜组成的透析器之间的血液相容性,其中除 PVP 外,其他透析器都额外使用了涂有其他材料的 PS 膜。将血液注入透析器并堵塞后,通过轻微旋转透析器进行动态刺激,持续 4 小时,尽管没有血液循环:结果:与不含聚乙烯吡咯烷酮(PVP)的不对称三醋酸纤维素膜和含 PVP 的聚砜(PS)膜相比,涂有维生素 E 的聚砜(PS)膜显示出更高的厚度和柔性层的柔软度。我们发现,与涂有 NV 聚合物的透析器相比,涂有维生素 E 的透析器能明显抑制血小板的减少、β-TG 的增加和 PF4 的增加。此外,由于内表面吸附了蛋白质,涂有维生素 E 的透析器中的总蛋白、纤连蛋白和 vWF 水平明显降低:结论:体外中空纤维膜内表面柔性层的厚度和柔软度会影响临床研究中血液凝固性能的差异。未来的临床试验需要证实我们的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
International Journal of Artificial Organs
International Journal of Artificial Organs 医学-工程:生物医学
CiteScore
3.40
自引率
5.90%
发文量
92
审稿时长
3 months
期刊介绍: The International Journal of Artificial Organs (IJAO) publishes peer-reviewed research and clinical, experimental and theoretical, contributions to the field of artificial, bioartificial and tissue-engineered organs. The mission of the IJAO is to foster the development and optimization of artificial, bioartificial and tissue-engineered organs, for implantation or use in procedures, to treat functional deficits of all human tissues and organs.
期刊最新文献
Performance study of dual heart assisted control system based on SL-SMC physiological combination controller. Depurative capacity toward medium molecules of the dialyzer Toray NV-U® Hydrolink™: A new hydrophilic membrane to perform online hemodiafiltration. Assessment of haemolysis models for a positive-displacement total artificial heart. Dynamic VAD simulations: Performing accurate simulations of ventricular assist devices in interaction with the cardiovascular system. Flexible inner surface of polysulfone membranes prevents platelet adhesive protein adsorption and improves antithrombogenicity in vitro.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1