{"title":"[Usefulness of Copper Filter Addition and Potential for Dose Reduction in Hip-joint Radiography].","authors":"Tomoyoshi Kawabata, Shoko Kamiyama, Shu Onodera","doi":"10.6009/jjrt.2024-1445","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.</p><p><strong>Methods: </strong>We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.</p><p><strong>Results: </strong>The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.</p><p><strong>Conclusion: </strong>It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-09-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1445","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.

Methods: We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.

Results: The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.

Conclusion: It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[在髋关节 X 射线照相术中添加铜滤波器的效用和降低剂量的潜力]。
目的:在这项研究中,我们评估了在髋关节 X 射线成像中添加铜(Cu)滤光片后的图像质量和辐射剂量降低情况:方法:我们测量了在 70 千伏电压下未添加(0 毫米)和添加(0.1/0.2 毫米)铜滤光片时的有效能量,并计算了在入口表面剂量恒定的情况下软组织-骨骼对比度和信号差-噪声比(SDNR)。之后,我们估算了剂量减少率:结果:0 毫米铜的有效能量为 32.07 keV,0.1 毫米铜的有效能量为 37.59 keV,0.2 毫米铜的有效能量为 40.91 keV。随着加铜滤光片厚度的增加,对比度降低,但 SDNR 增加。通过测量 SDNR 计算出的骨骼剂量减少率,最大值为 0.1 mm Cu 34%,0.2 mm Cu 47%:结论:在髋关节 X 射线成像中加入铜滤光片可减少入口表面剂量,同时根据 SDNR 保持图像质量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
[Usefulness of Copper Filter Addition and Potential for Dose Reduction in Hip-joint Radiography]. [Effect of Pulse Wave Synchronization on T1 Value in Cardiac T1 Mapping: Is Pulse Wave Synchronization a Substitute for Electrocardiogram Gating?] [Development of a Head Radiography Simulator System Using Pose Estimation Techniques]. [Artifact Assessment in the Skull Base Region of Head CT Images with Various Tilt Angles Relative to the Orbitomeatal Line]. [Evaluation of Transmitted X-ray Spectrum, Lead Equivalent, and Uniformity of Radiation Protective Clothing Made of Lead-containing and Lead-free Materials].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1