{"title":"Reproducibility of ultrasound-derived fat fraction in measuring hepatic steatosis.","authors":"Danlei Song, Pingping Wang, Jiahao Han, Huihui Chen, Ruixia Gao, Ling Li, Jia Li","doi":"10.1186/s13244-024-01834-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Steatotic liver disease (SLD) has become the most common cause of chronic liver disease. Nevertheless, the non-invasive quantitative diagnosis of steatosis is still lacking in clinical practice. This study aimed to evaluate the reproducibility of the new parameter for steatosis quantification named ultrasound-derived fat fraction (UDFF).</p><p><strong>Materials and methods: </strong>The UDFF values were independently executed by two operators in two periods. In the process, repeated measurements of the same patient were performed by the same operator under different conditions (liver segments, respiration, positions, and dietary). Finally, the results of some subjects (28) were compared with the MRI-derived proton density fat fraction (PDFF). The concordance analysis was mainly achieved by the intraclass correlation coefficient (ICC) and Bland-Altman.</p><p><strong>Results: </strong>One hundred-five participants were included in the study. UDFF had good reliability in measuring the adult liver (ICC<sub>intra-observer</sub> = 0.96, ICC<sub>inter-observer</sub> = 0.94). Meanwhile, the ICC of the two operators increased over time. The variable measurement states did not influence the UDFF values on the surface, but they affected the coefficient of variation (Cov) of the results. Segment 8 (S8), end-expiratory, supine, and fasting images had the most minor variability. On the other hand, the UDFF value of S8 displayed satisfied consistency with PDFF (mean difference, -0.24 ± 1.44), and the results of both S5 (mean difference: -0.56 ± 3.95) and S8 (mean difference: 0.73 ± 1.87) agreed well with the whole-liver PDFF.</p><p><strong>Conclusion: </strong>UDFF measurements had good reproducibility. Furthermore, the state of S8, end-expiration, supine, and fasting might be the more stable measurement approach.</p><p><strong>Critical relevance statement: </strong>UDFF is the quantitative ultrasound parameter of hepatic steatosis and has good reproducibility. It can show more robust performance under specific measurement conditions (S8, end-expiratory, supine, and fasting).</p><p><strong>Trial registration: </strong>The research protocol was registered at the Chinese Clinical Trial Registry on October 9, 2023 ( http://www.chictr.org.cn/ ). The registration number is ChiCTR 2300076457.</p><p><strong>Key points: </strong>There is a lack of non-invasive quantitative measurement options for hepatic steatosis. UDFF demonstrated excellent reproducibility in measuring hepatic steatosis. S8, end-expiratory, supine, and fasting may be the more stable measuring condition. Training could improve the operators' measurement stability. Variable measurement state affects the repeatability of the UDFF values (Cov).</p>","PeriodicalId":13639,"journal":{"name":"Insights into Imaging","volume":"15 1","pages":"254"},"PeriodicalIF":4.1000,"publicationDate":"2024-10-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11496408/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Insights into Imaging","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s13244-024-01834-1","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: Steatotic liver disease (SLD) has become the most common cause of chronic liver disease. Nevertheless, the non-invasive quantitative diagnosis of steatosis is still lacking in clinical practice. This study aimed to evaluate the reproducibility of the new parameter for steatosis quantification named ultrasound-derived fat fraction (UDFF).
Materials and methods: The UDFF values were independently executed by two operators in two periods. In the process, repeated measurements of the same patient were performed by the same operator under different conditions (liver segments, respiration, positions, and dietary). Finally, the results of some subjects (28) were compared with the MRI-derived proton density fat fraction (PDFF). The concordance analysis was mainly achieved by the intraclass correlation coefficient (ICC) and Bland-Altman.
Results: One hundred-five participants were included in the study. UDFF had good reliability in measuring the adult liver (ICCintra-observer = 0.96, ICCinter-observer = 0.94). Meanwhile, the ICC of the two operators increased over time. The variable measurement states did not influence the UDFF values on the surface, but they affected the coefficient of variation (Cov) of the results. Segment 8 (S8), end-expiratory, supine, and fasting images had the most minor variability. On the other hand, the UDFF value of S8 displayed satisfied consistency with PDFF (mean difference, -0.24 ± 1.44), and the results of both S5 (mean difference: -0.56 ± 3.95) and S8 (mean difference: 0.73 ± 1.87) agreed well with the whole-liver PDFF.
Conclusion: UDFF measurements had good reproducibility. Furthermore, the state of S8, end-expiration, supine, and fasting might be the more stable measurement approach.
Critical relevance statement: UDFF is the quantitative ultrasound parameter of hepatic steatosis and has good reproducibility. It can show more robust performance under specific measurement conditions (S8, end-expiratory, supine, and fasting).
Trial registration: The research protocol was registered at the Chinese Clinical Trial Registry on October 9, 2023 ( http://www.chictr.org.cn/ ). The registration number is ChiCTR 2300076457.
Key points: There is a lack of non-invasive quantitative measurement options for hepatic steatosis. UDFF demonstrated excellent reproducibility in measuring hepatic steatosis. S8, end-expiratory, supine, and fasting may be the more stable measuring condition. Training could improve the operators' measurement stability. Variable measurement state affects the repeatability of the UDFF values (Cov).
期刊介绍:
Insights into Imaging (I³) is a peer-reviewed open access journal published under the brand SpringerOpen. All content published in the journal is freely available online to anyone, anywhere!
I³ continuously updates scientific knowledge and progress in best-practice standards in radiology through the publication of original articles and state-of-the-art reviews and opinions, along with recommendations and statements from the leading radiological societies in Europe.
Founded by the European Society of Radiology (ESR), I³ creates a platform for educational material, guidelines and recommendations, and a forum for topics of controversy.
A balanced combination of review articles, original papers, short communications from European radiological congresses and information on society matters makes I³ an indispensable source for current information in this field.
I³ is owned by the ESR, however authors retain copyright to their article according to the Creative Commons Attribution License (see Copyright and License Agreement). All articles can be read, redistributed and reused for free, as long as the author of the original work is cited properly.
The open access fees (article-processing charges) for this journal are kindly sponsored by ESR for all Members.
The journal went open access in 2012, which means that all articles published since then are freely available online.