Mechanical effect of taper position in abutment hole and screw taper angles on implant system and peri-implant tissue: a finite element analysis.

IF 1.7 4区 医学 Q3 COMPUTER SCIENCE, INTERDISCIPLINARY APPLICATIONS Computer Methods in Biomechanics and Biomedical Engineering Pub Date : 2024-11-04 DOI:10.1080/10255842.2024.2423253
Hu Hou, Jianguo Zhang, Song Huang, Fengling Hu, Youcheng Yu, Liang Song
{"title":"Mechanical effect of taper position in abutment hole and screw taper angles on implant system and peri-implant tissue: a finite element analysis.","authors":"Hu Hou, Jianguo Zhang, Song Huang, Fengling Hu, Youcheng Yu, Liang Song","doi":"10.1080/10255842.2024.2423253","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The study aimed to investigate the mechanical effects of the taper position in the abutment hole and the screw taper angles on the implant system and peri-implant tissue using finite element analysis.</p><p><strong>Methods: </strong>Four taper positions (L1, L2, L3, L4) in the abutment hole were established using 3D software and five screw taper angles (30°, 60°, 90°, 120°, 180°) were set.</p><p><strong>Result: </strong>Taper position significantly affects the stresses in the implant system. The 30° and 180° angles (L4 position) showed less stress than other angles.</p><p><strong>Conclusion: </strong>Elevated taper position and reasonable taper angle are beneficial in reducing the stress in the implant system.</p>","PeriodicalId":50640,"journal":{"name":"Computer Methods in Biomechanics and Biomedical Engineering","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-11-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computer Methods in Biomechanics and Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1080/10255842.2024.2423253","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"COMPUTER SCIENCE, INTERDISCIPLINARY APPLICATIONS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: The study aimed to investigate the mechanical effects of the taper position in the abutment hole and the screw taper angles on the implant system and peri-implant tissue using finite element analysis.

Methods: Four taper positions (L1, L2, L3, L4) in the abutment hole were established using 3D software and five screw taper angles (30°, 60°, 90°, 120°, 180°) were set.

Result: Taper position significantly affects the stresses in the implant system. The 30° and 180° angles (L4 position) showed less stress than other angles.

Conclusion: Elevated taper position and reasonable taper angle are beneficial in reducing the stress in the implant system.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
基台孔锥位和螺钉锥角对种植体系统和种植体周围组织的力学影响:有限元分析。
目的:该研究旨在利用有限元分析方法研究基台孔锥度位置和螺钉锥度角度对种植体系统和种植体周围组织的力学影响:方法:使用三维软件确定基台孔中的四个锥度位置(L1、L2、L3、L4),并设定五个螺钉锥度角度(30°、60°、90°、120°、180°):结果:锥度位置对种植体系统的应力有很大影响。结果:锥度位置对种植体系统的应力影响很大,30° 和 180°(L4 位置)的应力小于其他角度:结论:抬高锥体位置和合理的锥体角度有利于降低种植体系统的应力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
4.10
自引率
6.20%
发文量
179
审稿时长
4-8 weeks
期刊介绍: The primary aims of Computer Methods in Biomechanics and Biomedical Engineering are to provide a means of communicating the advances being made in the areas of biomechanics and biomedical engineering and to stimulate interest in the continually emerging computer based technologies which are being applied in these multidisciplinary subjects. Computer Methods in Biomechanics and Biomedical Engineering will also provide a focus for the importance of integrating the disciplines of engineering with medical technology and clinical expertise. Such integration will have a major impact on health care in the future.
期刊最新文献
Accurate detection of gait events using neural networks and IMU data mimicking real-world smartphone usage. Exploring coronavirus sequence motifs through convolutional neural network for accurate identification of COVID-19. Coexistence of horizontal bone loss and dehiscence with the bundle and conventional fiber post: a finite element analysis. Effects of a soft back exoskeleton on lower lumbar spine loads during manual materials handling: a musculoskeletal modelling study. Mechanical effect of taper position in abutment hole and screw taper angles on implant system and peri-implant tissue: a finite element analysis.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1