{"title":"A Comparison of Pumping Infusion-Induced Contamination With Different Syringe Types.","authors":"Yutaka Kawakami, Takashi Tagami","doi":"10.1097/NAN.0000000000000560","DOIUrl":null,"url":null,"abstract":"<p><p>Health care providers use several types of syringes in their daily routine to provide or safely deliver therapeutic agents to patients. Pump infusion with a syringe (PIS) are used for the rapid delivery of large amounts of fluid or blood to patients in critical conditions, such as hypovolemic shock. Patients often experience bacterial infections, such as catheter-related bloodstream infections, specifically when they are in critical condition in the intensive care unit (ICU) after surgery or undergoing resuscitation in an emergency department. A previous study has shown that PIS may cause intraluminal contamination. This study compared PIS-induced contamination among different types of available syringes, including disposable plastic syringes produced by several companies, glass syringes, and syringes with plastic covers. The authors found that plastic syringes caused approximately equivalent PIS-induced contamination and that glass syringes caused substantially more contamination than plastic syringes. However, syringes equipped with a plastic cover exhibited no contamination. Furthermore, disinfection with ethanol completely prevented PIS-induced contamination. This study supports the evidence that PIS may cause bacterial contamination and that a thorough aseptic technique is needed, especially when using glass syringes. These findings highlight the need to develop alternative devices for rapid infusion.</p>","PeriodicalId":46291,"journal":{"name":"Journal of Infusion Nursing","volume":null,"pages":null},"PeriodicalIF":2.9000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Infusion Nursing","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/NAN.0000000000000560","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/30 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"NURSING","Score":null,"Total":0}
引用次数: 0
Abstract
Health care providers use several types of syringes in their daily routine to provide or safely deliver therapeutic agents to patients. Pump infusion with a syringe (PIS) are used for the rapid delivery of large amounts of fluid or blood to patients in critical conditions, such as hypovolemic shock. Patients often experience bacterial infections, such as catheter-related bloodstream infections, specifically when they are in critical condition in the intensive care unit (ICU) after surgery or undergoing resuscitation in an emergency department. A previous study has shown that PIS may cause intraluminal contamination. This study compared PIS-induced contamination among different types of available syringes, including disposable plastic syringes produced by several companies, glass syringes, and syringes with plastic covers. The authors found that plastic syringes caused approximately equivalent PIS-induced contamination and that glass syringes caused substantially more contamination than plastic syringes. However, syringes equipped with a plastic cover exhibited no contamination. Furthermore, disinfection with ethanol completely prevented PIS-induced contamination. This study supports the evidence that PIS may cause bacterial contamination and that a thorough aseptic technique is needed, especially when using glass syringes. These findings highlight the need to develop alternative devices for rapid infusion.
医疗服务提供者在日常工作中使用多种类型的注射器为患者提供或安全输送治疗药物。带注射器的泵输液(PIS)用于向低血容量休克等危重病人快速输送大量液体或血液。患者经常会受到细菌感染,如导管相关血流感染,特别是在手术后重症监护室(ICU)或急诊科接受抢救时。之前的一项研究表明,PIS 可能会导致管腔内污染。本研究比较了不同类型注射器(包括多家公司生产的一次性塑料注射器、玻璃注射器和带塑料盖的注射器)的 PIS 引起的污染。作者发现,塑料注射器造成的 PIS 引起的污染大致相当,而玻璃注射器造成的污染要比塑料注射器多得多。但是,装有塑料盖的注射器没有污染。此外,用乙醇消毒可完全防止 PIS 引起的污染。这项研究证实了 PIS 可能会导致细菌污染,因此需要采用彻底的无菌技术,尤其是在使用玻璃注射器时。这些发现凸显了开发快速输液替代设备的必要性。
期刊介绍:
Journal of Infusion Nursing, the official publication of the Infusion Nurses Society (INS), seeks to promote excellence in infusion nursing by presenting new research, clinical reviews, case studies, and professional development information relevant to the practice of infusion therapy. Articles selected for publication represent the broad scope of the infusion specialty and draw on the expertise of all healthcare providers who participate in the delivery of infusion therapy.