Clinical value of portable 12‐lead electrocardiography devices in patients with heart disease: A validation study

IF 1.2 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS Journal of electrocardiology Pub Date : 2025-01-01 Epub Date: 2024-11-22 DOI:10.1016/j.jelectrocard.2024.153835
Xiuyan Lu , Qiuhuan Wang , Xiujie Sun , Yibing Shao , Wenbo Jiang
{"title":"Clinical value of portable 12‐lead electrocardiography devices in patients with heart disease: A validation study","authors":"Xiuyan Lu ,&nbsp;Qiuhuan Wang ,&nbsp;Xiujie Sun ,&nbsp;Yibing Shao ,&nbsp;Wenbo Jiang","doi":"10.1016/j.jelectrocard.2024.153835","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><div>The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.</div></div><div><h3>Materials and methods</h3><div>This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired <em>t-</em>tests were used to compare the continuous variables. Agreement was evaluated via Bland–Altman plots.</div></div><div><h3>Results</h3><div>Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland–Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (<em>p</em> &lt; 0.001) but not clinically significant.</div></div><div><h3>Conclusions</h3><div>ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.</div></div>","PeriodicalId":15606,"journal":{"name":"Journal of electrocardiology","volume":"88 ","pages":"Article 153835"},"PeriodicalIF":1.2000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of electrocardiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0022073624003054","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/22 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Objective

The present study was conducted to assess the accuracy and reliability of portable 12‑lead electrocardiography (ECG) devices in patients with heart disease.

Materials and methods

This single-center, prospective, blinded study enrolled 62 patients between September and October 2023 from the Heart Center of a Class III hospital. In sequential tests on each patient, heart rate (HR) and the PR, QT, QTc and QRS intervals of ECG recordings obtained with a portable 12‑lead device (Weheal, CN) were compared with those obtained via conventional 12‑lead ECG. ECG parameters were read in batches by 3 blinded electrophysiologists. Two-tailed paired t-tests were used to compare the continuous variables. Agreement was evaluated via Bland–Altman plots.

Results

Sixty-two patients were included. HR and the QT, QTc and QRS intervals from the portable 12‑lead electrocardiogram recordings were essentially the same as those obtained via conventional ECG. Bland–Altman analysis revealed no significant differences in these values, indicating suitable agreement between the 2 measurements. The PR interval was 176.89 ± 29.53 ms in the portable group and 161.56 ± 17.78 ms in the standard group, which was statistically (p < 0.001) but not clinically significant.

Conclusions

ECG recordings obtained with a portable 12‑lead device (Weheal, CN) allow for accurate HR, PR, QT, QTc and QRS assessments. Considering its simplicity, this approach has advantages over conventional ECG and can provide an alternative for evaluating patients outside the hospital. How to improve patients' acceptance of portable ECG machines still needs further research.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
便携式12导联心电图仪在心脏病患者中的临床价值:一项验证性研究。
目的:本研究旨在评估便携式12导联心电图(ECG)设备在心脏病患者中的准确性和可靠性。材料和方法:这项单中心、前瞻性、盲法研究于2023年9月至10月从某三甲医院心脏中心招募了62例患者。在对每位患者进行的顺序试验中,将使用便携式12导联仪(Weheal, CN)获得的心电图记录的心率(HR)、PR、QT、QTc和QRS间期与常规12导联心电图记录的心电图进行比较。3名盲法电生理学家分批读取心电参数。采用双尾配对t检验比较连续变量。通过Bland-Altman图评估一致性。结果:纳入62例患者。便携式12导联心电图记录的HR、QT、QTc和QRS间期与常规心电图记录基本相同。Bland-Altman分析显示这些值没有显著差异,表明2个测量值之间存在合适的一致性。便携式组的PR间期为176.89±29.53 ms,标准组的PR间期为161.56±17.78 ms,差异有统计学意义(p)。结论:使用便携式12导联仪(Weheal, CN)获得的心电图记录可以准确评估HR、PR、QT、QTc和QRS。考虑到它的简单性,这种方法比传统的心电图有优势,可以为医院外的患者提供评估的替代方法。如何提高患者对便携式心电机的接受程度还有待进一步研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of electrocardiology
Journal of electrocardiology 医学-心血管系统
CiteScore
2.70
自引率
7.70%
发文量
152
审稿时长
38 days
期刊介绍: The Journal of Electrocardiology is devoted exclusively to clinical and experimental studies of the electrical activities of the heart. It seeks to contribute significantly to the accuracy of diagnosis and prognosis and the effective treatment, prevention, or delay of heart disease. Editorial contents include electrocardiography, vectorcardiography, arrhythmias, membrane action potential, cardiac pacing, monitoring defibrillation, instrumentation, drug effects, and computer applications.
期刊最新文献
Not just a Wenckebach, not yet a complete heart block Early electrocardiographic repolarization changes are associated with subclinical cancer therapy-related cardiac dysfunction in lymphoma patients: A machine learning-assisted longitudinal study 3D ECG: a new simplified view Case report: An irregular wide QRS rhythm with flutter waves and a positive precordial concordance Atrial anatomy and function in patients aged 75 years or older with advanced interatrial block. Data from a prospective study with matched controls
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1