Initial Multicenter Experience With the New Castor Arch Branched Device in Europe: A Middle-term Results Study.

IF 1.7 2区 医学 Q3 PERIPHERAL VASCULAR DISEASE Journal of Endovascular Therapy Pub Date : 2024-12-12 DOI:10.1177/15266028241304307
Michał Żołnierczuk, Paweł Rynio, Anita Rybicka, Jerzy Głowiński, Artur Milnerowicz, Kornel Pormanczuk, Andrzej Kostka, Grzegorz Oszkinis, Jacek Hobot, Piotr Gutowski, Arkadiusz Kazimierczak
{"title":"Initial Multicenter Experience With the New Castor Arch Branched Device in Europe: A Middle-term Results Study.","authors":"Michał Żołnierczuk, Paweł Rynio, Anita Rybicka, Jerzy Głowiński, Artur Milnerowicz, Kornel Pormanczuk, Andrzej Kostka, Grzegorz Oszkinis, Jacek Hobot, Piotr Gutowski, Arkadiusz Kazimierczak","doi":"10.1177/15266028241304307","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Published reports demonstrate that the use of Castor stent-graft is a promising treatment of aortic pathologies with the need to land in zone 2. However, there is a lack of publications on the medium-term results of Castor in the European population. This research evaluates the mid-term results of the Castor stent-graft in several Polish centers for the treatment of aortic pathologies and enriches the current knowledge of Castor stent-grafts.</p><p><strong>Material and methods: </strong>Twenty-one patients from 5 polish vascular surgery centers in subacute phase of type B aortic dissection (TBAD) with at least 1 classic risk factor of aneurysmatic degeneration, thoracic aortic aneurysms (TAAs) exceeding 5.5 cm, subacute phase of intramural hematomas (IMHs) and penetrating aortic ulcerations (PAUs) with at least 1 indication to endovascular treatment. Structure of the aortic arch and factors of aneurysmal degeneration for TBAD were highlighted. Early complications, long-term complications, and reinterventions after Castor device implantation are reported in the study.</p><p><strong>Results: </strong>Twenty (95.24%) patients survived, of which a technical success involving implantation of the Castor device in the correct location with a patent branch for the left subclavian artery, and no endoleaks were noted in 19 (90.48%) patients. After the procedure, there were 2 (9.52%) cases of type I endoleak (in which technical success was not achieved) and 2 (9.52%) cases of bird beak. In addition, 1 case of access site hematoma, 1 case of pseudoaneurysm, and 1 case of access iliac artery rupture requiring implantation of a Viabahn device were observed. During the follow-up period (mean 14 months; range = 1-40 months), 1 patient required reintervention due to type I endoleak and 1 patient due to left subclavian artery (LSA) branch thrombosis. Two patients required subsequent branched endovascular aortic repair procedure due to unfavorable remodeling and fast aneurysm formation in visceral aorta, regardless of Castor results.</p><p><strong>Conclusion: </strong>The Castor device is an easy-to-use stent-graft with good medium-term results. It is an excellent option in cases requiring LSA revascularization during TEVAR.</p><p><strong>Clinical impact: </strong>In this study, we investigate the medium-term results of the Castor stent graft in patients with type B aortic dissection, thoracic aortic aneurysm, intramural hematoma and penetrating aortic ulceration. The medium-term results of treatment with the Castor device are not yet well studied. Currently, there are only a small number of publications on the safety, complications and success of Castor device implantation.</p>","PeriodicalId":50210,"journal":{"name":"Journal of Endovascular Therapy","volume":" ","pages":"15266028241304307"},"PeriodicalIF":1.7000,"publicationDate":"2024-12-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Endovascular Therapy","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1177/15266028241304307","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Published reports demonstrate that the use of Castor stent-graft is a promising treatment of aortic pathologies with the need to land in zone 2. However, there is a lack of publications on the medium-term results of Castor in the European population. This research evaluates the mid-term results of the Castor stent-graft in several Polish centers for the treatment of aortic pathologies and enriches the current knowledge of Castor stent-grafts.

Material and methods: Twenty-one patients from 5 polish vascular surgery centers in subacute phase of type B aortic dissection (TBAD) with at least 1 classic risk factor of aneurysmatic degeneration, thoracic aortic aneurysms (TAAs) exceeding 5.5 cm, subacute phase of intramural hematomas (IMHs) and penetrating aortic ulcerations (PAUs) with at least 1 indication to endovascular treatment. Structure of the aortic arch and factors of aneurysmal degeneration for TBAD were highlighted. Early complications, long-term complications, and reinterventions after Castor device implantation are reported in the study.

Results: Twenty (95.24%) patients survived, of which a technical success involving implantation of the Castor device in the correct location with a patent branch for the left subclavian artery, and no endoleaks were noted in 19 (90.48%) patients. After the procedure, there were 2 (9.52%) cases of type I endoleak (in which technical success was not achieved) and 2 (9.52%) cases of bird beak. In addition, 1 case of access site hematoma, 1 case of pseudoaneurysm, and 1 case of access iliac artery rupture requiring implantation of a Viabahn device were observed. During the follow-up period (mean 14 months; range = 1-40 months), 1 patient required reintervention due to type I endoleak and 1 patient due to left subclavian artery (LSA) branch thrombosis. Two patients required subsequent branched endovascular aortic repair procedure due to unfavorable remodeling and fast aneurysm formation in visceral aorta, regardless of Castor results.

Conclusion: The Castor device is an easy-to-use stent-graft with good medium-term results. It is an excellent option in cases requiring LSA revascularization during TEVAR.

Clinical impact: In this study, we investigate the medium-term results of the Castor stent graft in patients with type B aortic dissection, thoracic aortic aneurysm, intramural hematoma and penetrating aortic ulceration. The medium-term results of treatment with the Castor device are not yet well studied. Currently, there are only a small number of publications on the safety, complications and success of Castor device implantation.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
欧洲新Castor拱形分支装置的初步多中心经验:中期结果研究。
背景:已发表的报告表明,使用Castor支架移植是一种有希望的治疗主动脉病变的方法,需要在2区着陆。然而,缺乏关于Castor在欧洲人口中的中期结果的出版物。本研究评估了Castor支架移植在波兰几个中心治疗主动脉病变的中期结果,丰富了Castor支架移植的现有知识。材料和方法:21例来自5个国家血管外科中心的B型主动脉夹层(TBAD)亚急性期患者,至少有1个典型危险因素为动脉瘤变性,胸主动脉瘤(TAAs)超过5.5 cm,亚急性期壁内血肿(IMHs)和穿透性主动脉溃疡(PAUs),至少有1个血管内治疗指征。强调了主动脉弓的结构和TBAD的动脉瘤变性因素。本研究报告了Castor装置植入后的早期并发症、长期并发症和再干预。结果:20例(95.24%)患者存活,其中19例(90.48%)患者未发生内漏,19例(90.48%)患者成功植入左锁骨下动脉分支的正确位置。术后未取得技术成功的I型漏腔2例(9.52%),鸟喙漏腔2例(9.52%)。此外,观察到1例通路部位血肿,1例假性动脉瘤,1例通路髂动脉破裂需要植入Viabahn装置。随访期间(平均14个月;范围= 1-40个月),1例患者因I型内漏而需要再次干预,1例患者因左锁骨下动脉(LSA)分支血栓形成。无论Castor结果如何,2例患者由于脏主动脉重构不利和快速动脉瘤形成而需要后续的血管内动脉分支修复手术。结论:Castor支架是一种易于使用的支架移植,中期效果良好。在TEVAR期间需要LSA血运重建术的病例中,这是一个很好的选择。临床影响:在本研究中,我们研究Castor支架在B型主动脉夹层、胸主动脉瘤、壁内血肿和穿透性主动脉溃疡患者中的中期效果。Castor装置治疗的中期结果尚未得到很好的研究。目前,关于Castor装置植入的安全性、并发症及成功与否的报道较少。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.30
自引率
15.40%
发文量
203
审稿时长
6-12 weeks
期刊介绍: The Journal of Endovascular Therapy (formerly the Journal of Endovascular Surgery) was established in 1994 as a forum for all physicians, scientists, and allied healthcare professionals who are engaged or interested in peripheral endovascular techniques and technology. An official publication of the International Society of Endovascular Specialists (ISEVS), the Journal of Endovascular Therapy publishes peer-reviewed articles of interest to clinicians and researchers in the field of peripheral endovascular interventions.
期刊最新文献
Transapical to Femoral Through-and-Through Wire Technique for Ruptured Aortic Arch Aneurysm Management: Case Report and Review of Literature. Percutaneous Superficial Temporal Artery Access Facilitating Carotid Artery Stenting Performed From Distal Radial Artery. Stent Implantation and Thromboendarterectomy for the Common Femoral Artery in Real-World Practice. Endovascular Treatment for Isolated Infected Iliac Artery Aneurysms. Initial Clinical Experience With AneuFix Injectable Biocompatible Elastomer for Translumbar Embolization of Type 2 Endoleaks.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1