{"title":"[A preliminary study of risk factors for the stress fracture in treated with clavicle hook plate].","authors":"Peng-Fei Nie, Yuan-Lin Xu, Yong-Fu Li, Lun Zhang, Qian-Qian Zhou, Jian-Nyu Luo, Jian Guo","doi":"10.12200/j.issn.1003-0034.20230132","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To investigate the effects of bone density, plate bending degree and proximal screw type on the stress fracture of clavicle hook.</p><p><strong>Methods: </strong>Three sows weighing between 45 and 50 kg were selected, from which a total of 40 rivs were collected. The 15 ribs of sows were divided into 3 groups according to bone density and bone hardness with 5 rivs in each group. And then the 3 groups were fixed with 6-hole collarbone hook plates and 3 locking screws. Measure the maximum torsion force when the ribs were fractured by force. The same size 15 rids were divided into 3 groups, named forward bending group, 0° group(the angle between the plate surface and the rib surface) and reverse bending group. All fixed with 6-hole collarbone hook plates and locking screws to measure the maximum torsion force of rib stress fracture. Then the same size 10 rids were divided into 2 groups, the normal screw group and the locking screw group with 5 ribs in each group. Both groups were fixed with 6-hole collarbone hook plates and screws. The normal screw group was a normal screw, fixed in proximal end, and two locking screws. The locking screw group was fixed by locking screws. Measure the maximum torsion force of the two groups when the ribs fracture by force.</p><p><strong>Results: </strong>In the bone density experiment, the torque force of hard bone group (104.51±6.27) N was greater than the normal bone group (75.04±3.81) N(<i>t</i>=8.979, <i>P</i><0.05). The force of normal bone group was greater than the osteoporosis group (49.99±2.12) N(<i>t</i>=12.832, <i>P</i><0.05). In the bending collarbone hook experiment, the order of the torque force generated by each group as follow:the forward bending group (343.59±6.18) N greater than the 0° group (106.01±5.29) N(<i>t</i>=65.279, <i>P</i><0.05) greater than the reverse bending group (95.82±4.12) N(<i>t</i>=3.398, <i>P</i><0.05). The force of the normal screw group (98.68±0.70) N was greater than the locking screw group (50.20±0.95) N(<i>t</i>=91.484, <i>P</i><0.05). The data comparisons of each group were statistically significant.</p><p><strong>Conclusion: </strong>Bone density, plate bending degree and proximal screw type had an impact on stress fracture of clavicle hook plate. Higher bone density, forward bending of the steel plate, and ordinary screws in proximal end can reduce the rates of stress fractures of clavicle hooks.</p>","PeriodicalId":23964,"journal":{"name":"Zhongguo gu shang = China journal of orthopaedics and traumatology","volume":"38 1","pages":"61-5"},"PeriodicalIF":0.0000,"publicationDate":"2025-01-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Zhongguo gu shang = China journal of orthopaedics and traumatology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.12200/j.issn.1003-0034.20230132","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To investigate the effects of bone density, plate bending degree and proximal screw type on the stress fracture of clavicle hook.

Methods: Three sows weighing between 45 and 50 kg were selected, from which a total of 40 rivs were collected. The 15 ribs of sows were divided into 3 groups according to bone density and bone hardness with 5 rivs in each group. And then the 3 groups were fixed with 6-hole collarbone hook plates and 3 locking screws. Measure the maximum torsion force when the ribs were fractured by force. The same size 15 rids were divided into 3 groups, named forward bending group, 0° group(the angle between the plate surface and the rib surface) and reverse bending group. All fixed with 6-hole collarbone hook plates and locking screws to measure the maximum torsion force of rib stress fracture. Then the same size 10 rids were divided into 2 groups, the normal screw group and the locking screw group with 5 ribs in each group. Both groups were fixed with 6-hole collarbone hook plates and screws. The normal screw group was a normal screw, fixed in proximal end, and two locking screws. The locking screw group was fixed by locking screws. Measure the maximum torsion force of the two groups when the ribs fracture by force.

Results: In the bone density experiment, the torque force of hard bone group (104.51±6.27) N was greater than the normal bone group (75.04±3.81) N(t=8.979, P<0.05). The force of normal bone group was greater than the osteoporosis group (49.99±2.12) N(t=12.832, P<0.05). In the bending collarbone hook experiment, the order of the torque force generated by each group as follow:the forward bending group (343.59±6.18) N greater than the 0° group (106.01±5.29) N(t=65.279, P<0.05) greater than the reverse bending group (95.82±4.12) N(t=3.398, P<0.05). The force of the normal screw group (98.68±0.70) N was greater than the locking screw group (50.20±0.95) N(t=91.484, P<0.05). The data comparisons of each group were statistically significant.

Conclusion: Bone density, plate bending degree and proximal screw type had an impact on stress fracture of clavicle hook plate. Higher bone density, forward bending of the steel plate, and ordinary screws in proximal end can reduce the rates of stress fractures of clavicle hooks.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
0.50
自引率
0.00%
发文量
189
期刊最新文献
[Preliminary application of human-computer interaction CT imaging AI recognition and positioning technology in the treatment of type C1 distal radius fractures]. [Guiding significance of intra-articular sagittal reduction in the treatment of tibial plateau fractures]. [Meta-analysis of the efficacy of plate fixation and external fixator fixation in the treatment of AO-C type distal radius fractures]. [Preoperative diagnostic efficacy of novel blood markers white blood cell ratio and fibrinogen levels in periprosthetic joint infection]. [Short-term clinical efficacy of unilateral external fixator combined with percutaneous Kirschner wire fixation in the treatment of type C1 distal radius fractures in elderly patients].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1