Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer.

IF 4.1 1区 医学 Q1 OPHTHALMOLOGY Eye and Vision Pub Date : 2025-02-01 DOI:10.1186/s40662-024-00422-0
Xin Li, Chak Seng Lei, Rui Ning, Luze Liu, Aodong Chen, Xinning Yang, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Jinhai Huang
{"title":"Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer.","authors":"Xin Li, Chak Seng Lei, Rui Ning, Luze Liu, Aodong Chen, Xinning Yang, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Jinhai Huang","doi":"10.1186/s40662-024-00422-0","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To assess the repeatability and reproducibility of the Colombo IOL biometer (Moptim, China), which utilizes spectral-domain optical coherence tomography (SD-OCT), in measuring ocular parameters of normal subjects and to compare its agreement with the swept-source optical coherence tomography (SS-OCT)-based IOLMaster 700 biometer (Carl Zeiss Meditec AG, Germany).</p><p><strong>Methods: </strong>This prospective study included 91 eyes from 91 normal subjects. The evaluated parameters were axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), flattest and steepest meridian keratometry (Kf and Ks), mean keratometry (Km), astigmatism (AST) magnitude, white-to-white (WTW) distance, and pupil diameter (PD). The within-subject standard deviation (S<sub>w</sub>), test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to determine the repeatability and reproducibility. Paired t-tests and Bland-Altman plots with 95% limits of agreement (LoA) were employed to assess the agreement.</p><p><strong>Results: </strong>With respect to intraobserver repeatability, the S<sub>w</sub> and TRT values of all evaluated parameters were low. Except J<sub>45</sub> and PD, the ICCs were all higher than 0.928. The reproducibility S<sub>w</sub> and TRT values of Colombo IOL were also low, and ICCs were not lower than 0.900. Comparing Colombo IOL and IOLMaster 700, the 95% LoA of AL, CCT, AQD, ACD, LT, Kf, Ks, Km, AST, J<sub>0</sub>, J<sub>45</sub>, WTW and PD ranged from - 0.08 to 0.03 mm, - 21.58 to 5.09 μm, 0.01 to 0.15 mm, - 0.01 to 0.14 mm, - 0.05 to 0.10 mm, - 0.14 to 0.59 D, - 0.31 to 0.40 D, - 0.13 to 0.40 D, - 0.68 to 0.32 D, - 0.09 to 0.34 D, - 0.07 to 0.25 D, 0.11 to 1.47 mm, and - 0.97 to 2.31 mm, respectively.</p><p><strong>Conclusion: </strong>The new SD-OCT-based Colombo IOL biometer demonstrates excellent repeatability and reproducibility. Moreover, it generally agrees well with the SS-OCT-based IOLMaster 700, except for the WTW and PD measurements.</p>","PeriodicalId":12194,"journal":{"name":"Eye and Vision","volume":"12 1","pages":"6"},"PeriodicalIF":4.1000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11786401/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Eye and Vision","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40662-024-00422-0","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: To assess the repeatability and reproducibility of the Colombo IOL biometer (Moptim, China), which utilizes spectral-domain optical coherence tomography (SD-OCT), in measuring ocular parameters of normal subjects and to compare its agreement with the swept-source optical coherence tomography (SS-OCT)-based IOLMaster 700 biometer (Carl Zeiss Meditec AG, Germany).

Methods: This prospective study included 91 eyes from 91 normal subjects. The evaluated parameters were axial length (AL), central corneal thickness (CCT), aqueous depth (AQD), anterior chamber depth (ACD), lens thickness (LT), flattest and steepest meridian keratometry (Kf and Ks), mean keratometry (Km), astigmatism (AST) magnitude, white-to-white (WTW) distance, and pupil diameter (PD). The within-subject standard deviation (Sw), test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to determine the repeatability and reproducibility. Paired t-tests and Bland-Altman plots with 95% limits of agreement (LoA) were employed to assess the agreement.

Results: With respect to intraobserver repeatability, the Sw and TRT values of all evaluated parameters were low. Except J45 and PD, the ICCs were all higher than 0.928. The reproducibility Sw and TRT values of Colombo IOL were also low, and ICCs were not lower than 0.900. Comparing Colombo IOL and IOLMaster 700, the 95% LoA of AL, CCT, AQD, ACD, LT, Kf, Ks, Km, AST, J0, J45, WTW and PD ranged from - 0.08 to 0.03 mm, - 21.58 to 5.09 μm, 0.01 to 0.15 mm, - 0.01 to 0.14 mm, - 0.05 to 0.10 mm, - 0.14 to 0.59 D, - 0.31 to 0.40 D, - 0.13 to 0.40 D, - 0.68 to 0.32 D, - 0.09 to 0.34 D, - 0.07 to 0.25 D, 0.11 to 1.47 mm, and - 0.97 to 2.31 mm, respectively.

Conclusion: The new SD-OCT-based Colombo IOL biometer demonstrates excellent repeatability and reproducibility. Moreover, it generally agrees well with the SS-OCT-based IOLMaster 700, except for the WTW and PD measurements.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
Eye and Vision
Eye and Vision OPHTHALMOLOGY-
CiteScore
8.60
自引率
2.40%
发文量
89
审稿时长
15 weeks
期刊介绍: Eye and Vision is an open access, peer-reviewed journal for ophthalmologists and visual science specialists. It welcomes research articles, reviews, methodologies, commentaries, case reports, perspectives and short reports encompassing all aspects of eye and vision. Topics of interest include but are not limited to: current developments of theoretical, experimental and clinical investigations in ophthalmology, optometry and vision science which focus on novel and high-impact findings on central issues pertaining to biology, pathophysiology and etiology of eye diseases as well as advances in diagnostic techniques, surgical treatment, instrument updates, the latest drug findings, results of clinical trials and research findings. It aims to provide ophthalmologists and visual science specialists with the latest developments in theoretical, experimental and clinical investigations in eye and vision.
期刊最新文献
Influence of preoperative variables on the 3-month functional outcomes of the Vivity extended depth-of-focus intraocular lens: a prospective case series. Real-world study of phakic refractive lens for correction of high myopia. Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer. Topographic correlation of microperimetry with foveal microstructure characteristics in idiopathic epiretinal membrane patients with an ectopic inner foveal layer. Whole-genome sequencing identifies novel loci for keratoconus and facilitates risk stratification in a Han Chinese population.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1