In vitro comparison of the effects of direct attachment shape and location on forces and moments generated by thermoplastic aligners during simulated torque movement

IF 1.9 Q2 DENTISTRY, ORAL SURGERY & MEDICINE International Orthodontics Pub Date : 2025-06-01 Epub Date: 2025-02-04 DOI:10.1016/j.ortho.2025.100982
Janine Magura , Amin Akbari , Megann Lear , Alexandra Bojrab , George Eckert , Jie Chen , R. Scott Conley , Hakan Turkkahraman
{"title":"In vitro comparison of the effects of direct attachment shape and location on forces and moments generated by thermoplastic aligners during simulated torque movement","authors":"Janine Magura ,&nbsp;Amin Akbari ,&nbsp;Megann Lear ,&nbsp;Alexandra Bojrab ,&nbsp;George Eckert ,&nbsp;Jie Chen ,&nbsp;R. Scott Conley ,&nbsp;Hakan Turkkahraman","doi":"10.1016/j.ortho.2025.100982","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><div>The aim of this study was to investigate the effects of direct attachment shape and location on the forces and moments generated by thermoplastic aligners during simulated maxillary central incisor torque.</div></div><div><h3>Materials and methods</h3><div>A total of 7 typodonts were digitally printed with different attachment design and locations. Five clear aligners were fabricated for each typodont and placed on an orthodontic force tester (OFT) with the maxillary central incisor rotated palatally 1° around the incisal edge. Forces and moments were measured 2 times by the load cells. Analysis of variance (Anova) was used to determine the effects of group, tooth, and the group-by-tooth interaction on the outcomes; A two-sided 5% significance level was used for all tests. Analyses were performed using SAS version 9.4 (SAS Institute Inc., Cary, NC, USA).</div></div><div><h3>Results</h3><div>The two-way interactions between group and tooth were significant (<em>P</em> <!-->&lt;<!--> <!-->0.001) for all outcomes except facial/palatal moment around the X axis (M<sub>x</sub>) (<em>P</em> <!-->&gt;<!--> <!-->0.05). The horizontal ellipsoid attachment at the middle third generated the highest M<sub>x</sub> (−25.74<!--> <!-->±<!--> <!-->8.93<!--> <!-->Nmm) with facial crown/palatal root, while the rectangular attachment at the middle third yielded the lowest Mx (3.31<!--> <!-->±<!--> <!-->12.92<!--> <!-->Nmm).</div></div><div><h3>Conclusions</h3><div>Changing attachment shape and location had a significant effect on the forces and moments generated during simulated maxillary incisor torque. The best design for the incisor torque movement was found to be the horizontal ellipsoid attachment at the middle third. Besides the desired torque moment, all attachment shape and location combinations produced a vertical extrusive force as a side effect.</div></div>","PeriodicalId":45449,"journal":{"name":"International Orthodontics","volume":"23 2","pages":"Article 100982"},"PeriodicalIF":1.9000,"publicationDate":"2025-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Orthodontics","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1761722725000178","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/2/4 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

The aim of this study was to investigate the effects of direct attachment shape and location on the forces and moments generated by thermoplastic aligners during simulated maxillary central incisor torque.

Materials and methods

A total of 7 typodonts were digitally printed with different attachment design and locations. Five clear aligners were fabricated for each typodont and placed on an orthodontic force tester (OFT) with the maxillary central incisor rotated palatally 1° around the incisal edge. Forces and moments were measured 2 times by the load cells. Analysis of variance (Anova) was used to determine the effects of group, tooth, and the group-by-tooth interaction on the outcomes; A two-sided 5% significance level was used for all tests. Analyses were performed using SAS version 9.4 (SAS Institute Inc., Cary, NC, USA).

Results

The two-way interactions between group and tooth were significant (P < 0.001) for all outcomes except facial/palatal moment around the X axis (Mx) (P > 0.05). The horizontal ellipsoid attachment at the middle third generated the highest Mx (−25.74 ± 8.93 Nmm) with facial crown/palatal root, while the rectangular attachment at the middle third yielded the lowest Mx (3.31 ± 12.92 Nmm).

Conclusions

Changing attachment shape and location had a significant effect on the forces and moments generated during simulated maxillary incisor torque. The best design for the incisor torque movement was found to be the horizontal ellipsoid attachment at the middle third. Besides the desired torque moment, all attachment shape and location combinations produced a vertical extrusive force as a side effect.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在体外比较直接连接的形状和位置对热塑性对准器在模拟扭矩运动中产生的力和力矩的影响
本研究的目的是研究在模拟上颌中切牙扭矩时,直接附着的形状和位置对热塑性矫直器产生的力和力矩的影响。材料和方法采用不同的附件设计和位置,对7个排字件进行数字印刷。为每个印型牙制作5个明确的对准器,并放置在正畸力测试仪(OFT)上,上颌中切牙沿切缘腭侧旋转1°。力和力矩由称重传感器测量2次。采用方差分析(Anova)来确定组、牙和组间相互作用对结果的影响;所有检验均采用双侧5%显著性水平。使用SAS 9.4版(SAS Institute Inc., Cary, NC, USA)进行分析。结果组与牙的双向交互作用显著(P <;0.001),除了X轴周围的面部/腭力矩(Mx) (P >;0.05)。面冠/腭根水平椭球状附着体的Mx值最高(- 25.74±8.93 Nmm),矩形附着体的Mx值最低(3.31±12.92 Nmm)。结论改变附着体形状和位置对上颌切牙模拟转矩产生的力和力矩有显著影响。结果表明,在中三分之一位置采用水平椭球体附着体是实现切牙力矩运动的最佳设计。除了所需的扭矩外,所有附件形状和位置组合都会产生垂直挤压力作为副作用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
International Orthodontics
International Orthodontics DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
2.50
自引率
13.30%
发文量
71
审稿时长
26 days
期刊介绍: Une revue de référence dans le domaine de orthodontie et des disciplines frontières Your reference in dentofacial orthopedics International Orthodontics adresse aux orthodontistes, aux dentistes, aux stomatologistes, aux chirurgiens maxillo-faciaux et aux plasticiens de la face, ainsi quà leurs assistant(e)s. International Orthodontics is addressed to orthodontists, dentists, stomatologists, maxillofacial surgeons and facial plastic surgeons, as well as their assistants.
期刊最新文献
Accuracy of 3D-printed customized orthodontic bracket: A systematic review and meta-analysis of in vitro studies Clinical outcomes of orthodontic traction–based treatment strategies for impacted maxillary central incisors: A systematic review CBCT evaluation of midpalatal suture maturation for expansion planning in a Turkish population aged 8–30 years: A single-center retrospective cross-sectional study Effects of maxillary incisor intrusion with clear aligners under ideal retention: A finite element study using local coordinate systems Shear bond strength of metal orthodontic brackets bonded to artificially demineralized enamel treated with three remineralizing agents bonded with AgNP-modified adhesive: An in vitro study
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1