A comparative assessment of contamination rates in gastrointestinal endoscope reprocessing: sterilization versus high-level disinfection

IF 1.5 Q4 GASTROENTEROLOGY & HEPATOLOGY DEN open Pub Date : 2025-03-09 DOI:10.1002/deo2.70093
Tanyaporn Chantarojanasiri, Rachanikorn Rungrueangmaitree, Siriporn Thongsri, Urasa Jampa-ngern, Thawee Ratanachu-Ek
{"title":"A comparative assessment of contamination rates in gastrointestinal endoscope reprocessing: sterilization versus high-level disinfection","authors":"Tanyaporn Chantarojanasiri,&nbsp;Rachanikorn Rungrueangmaitree,&nbsp;Siriporn Thongsri,&nbsp;Urasa Jampa-ngern,&nbsp;Thawee Ratanachu-Ek","doi":"10.1002/deo2.70093","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Objectives</h3>\n \n <p>This study aimed to evaluate the disparity in culture results between sterilization and high-level disinfection (HLD) for duodenoscopes and linear endoscopic ultrasound (EUS), and to assess the effectiveness of different bacterial contamination detection methods.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>This is a prospective randomized study, including duodenoscopes and linear EUS with adenosine triphosphate bioluminescence assay values below 200 relative light units after manual cleaning which were randomly assigned to undergo either sterilization or HLD in a 1:1 ratio. Following disinfection, all endoscopes were subjected to adenosine triphosphate bioluminescence assay testing and cultures using both swab and liquid samples from endoscope channels.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Totally 752 endoscopes (444 duodenoscopes and 308 linear EUS) were studied. After disinfection, the positive culture rates for the sterilization and HLD groups were 5.9% and 7.2%, respectively (<i>p</i> = 0.460). No significant difference in contamination rates was observed between duodenoscopes and linear EUS (5.9% and 7.5%, respectively; <i>p</i> = 0.379), and no significant association between contamination rates and the presence of biliary stones was seen (7.3% vs. 6.9%; <i>p</i> = 0.613). The detection rate of bacteria from liquid samples taken from endoscope channels was 0.5%, which was significantly lower than the swabbing method (6.0%, <i>p</i> &lt; 0.001).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>This study found no statistically significant difference in contamination rates between sterilization and HLD methods for gastrointestinal endoscope reprocessing. The type of endoscope and the presence of biliary stones did not influence the positive culture rate. The swabbing method showed significantly higher bacterial detection when compared with liquid samples.</p>\n </section>\n </div>","PeriodicalId":93973,"journal":{"name":"DEN open","volume":"5 1","pages":""},"PeriodicalIF":1.5000,"publicationDate":"2025-03-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/deo2.70093","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"DEN open","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/deo2.70093","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"GASTROENTEROLOGY & HEPATOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives

This study aimed to evaluate the disparity in culture results between sterilization and high-level disinfection (HLD) for duodenoscopes and linear endoscopic ultrasound (EUS), and to assess the effectiveness of different bacterial contamination detection methods.

Methods

This is a prospective randomized study, including duodenoscopes and linear EUS with adenosine triphosphate bioluminescence assay values below 200 relative light units after manual cleaning which were randomly assigned to undergo either sterilization or HLD in a 1:1 ratio. Following disinfection, all endoscopes were subjected to adenosine triphosphate bioluminescence assay testing and cultures using both swab and liquid samples from endoscope channels.

Results

Totally 752 endoscopes (444 duodenoscopes and 308 linear EUS) were studied. After disinfection, the positive culture rates for the sterilization and HLD groups were 5.9% and 7.2%, respectively (p = 0.460). No significant difference in contamination rates was observed between duodenoscopes and linear EUS (5.9% and 7.5%, respectively; p = 0.379), and no significant association between contamination rates and the presence of biliary stones was seen (7.3% vs. 6.9%; p = 0.613). The detection rate of bacteria from liquid samples taken from endoscope channels was 0.5%, which was significantly lower than the swabbing method (6.0%, p < 0.001).

Conclusions

This study found no statistically significant difference in contamination rates between sterilization and HLD methods for gastrointestinal endoscope reprocessing. The type of endoscope and the presence of biliary stones did not influence the positive culture rate. The swabbing method showed significantly higher bacterial detection when compared with liquid samples.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
消化道内窥镜再处理过程中污染率的比较评估:消毒与高水平消毒
目的评价十二指肠镜和线性内镜超声(EUS)灭菌和高水平消毒(HLD)培养结果的差异,并评价不同细菌污染检测方法的有效性。方法:这是一项前瞻性随机研究,包括十二指肠镜和线性EUS,人工清洁后三磷酸腺苷生物发光测定值低于200相对光单位,随机分配进行灭菌或按1:1比例进行HLD。消毒后,所有内窥镜均进行三磷酸腺苷生物发光试验,并使用内窥镜通道的拭子和液体样本进行培养。结果共检查752台内窥镜,其中十二指肠镜444台,线性EUS 308台。消毒后,灭菌组和HLD组的阳性培养率分别为5.9%和7.2% (p = 0.460)。十二指肠镜和线性EUS的污染率无显著差异(分别为5.9%和7.5%);P = 0.379),污染率与胆结石存在之间无显著关联(7.3% vs. 6.9%;p = 0.613)。内窥镜通道液样细菌检出率为0.5%,显著低于拭子法(6.0%,p <;0.001)。结论本研究发现灭菌法和HLD法对胃肠道内窥镜再处理的污染率无统计学差异。内镜类型和胆结石的存在对阳性培养率没有影响。与液体样品相比,拭子法的细菌检出率显著高于液体样品。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.30
自引率
0.00%
发文量
0
期刊最新文献
Pancreatitis Following Biliary Endoscopic Retrograde Cholangiopancreatography for Common Bile Duct Stones in Japan: A Multicenter Prospective Cohort Study. Adalimumab-responsive Monogenic Inflammatory Bowel Disease With Pseudopolyposis Characteristic of TGFBR2 Variant in Loeys-Dietz Syndrome Efficacy of the Stent-in-Stent Technique as a Rescue Method for Removing Embedded Metallic Biliary Stents. Enhanced Visualization of the Distal Myotomy Endpoint during Peroral Endoscopic Myotomy Using Indocyanine Green With Infrared Imaging Comparative Analysis of Conventional Ideal Button Versus New Ideal Button ZERO for Percutaneous Endoscopic Gastrostomy Catheter Replacement
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1