Osteoarthritic Human Synovial Fluid Alters CoCrMo Electrochemical Properties on a Patient-Specific Basis

IF 2.3 3区 医学 Q2 ORTHOPEDICS Journal of Orthopaedic Research® Pub Date : 2025-04-07 DOI:10.1002/jor.26079
Bailey Bond, Madison N. Brown, Michael A. Kurtz, Alayna Robinson, Marc J. Mihalko, John R. Crockarell Jr., Christopher T. Holland, James L. Guyton, William M. Mihalko
{"title":"Osteoarthritic Human Synovial Fluid Alters CoCrMo Electrochemical Properties on a Patient-Specific Basis","authors":"Bailey Bond,&nbsp;Madison N. Brown,&nbsp;Michael A. Kurtz,&nbsp;Alayna Robinson,&nbsp;Marc J. Mihalko,&nbsp;John R. Crockarell Jr.,&nbsp;Christopher T. Holland,&nbsp;James L. Guyton,&nbsp;William M. Mihalko","doi":"10.1002/jor.26079","DOIUrl":null,"url":null,"abstract":"<div>\n \n <p>Cobalt-chromium-molybdenum (CoCrMo) femoral components are widely used in total knee arthroplasty (TKA). However, recent retrospective clinical trials associate moderate adverse local tissue reactions with CoCrMo release in the knee. Additionally, gaps persist in our understanding of the fundamental corrosion processes that occur at the CoCrMo-synovial fluid interface. In this study, we investigated the electrochemical behavior of CoCrMo in human synovial fluid obtained at the time of primary TKA, using CoCrMo in phosphate-buffered saline (PBS) as a comparison. Synovial fluid was collected from 118 patients immediately before arthroplasty, then transferred to a three-electrode electrochemical cell with a wrought CoCrMo alloy working electrode. To quantify electrochemical properties, open circuit potential (OCP), electrochemical impedance spectroscopy, and linear polarization tests were run. Generally, the properties varied on a patient-by-patient basis and significantly differed (<i>p</i> &lt; 0.05) from comparison tests performed in PBS. In human synovial fluid, we measured OCPs between a range of −0.38 and 0.15 V and corrosion potentials (<i>E</i><sub>corr</sub>) between −0.95 and −0.2 V. Additionally, we reported instantaneous corrosion rates (1/<i>R<sub>p</sub></i>) spanning nearly four orders of magnitude. The variability we documented suggested that the electrochemical properties of CoCrMo implants may depend on the patient's local physiological environment, influenced by the biological and chemical components of synovial fluid.</p>\n <p><b>Statement of Clinical Significance:</b> This study shows that human synovial fluid affects the electrochemical behavior of CoCrMo, evidenced by the variation on a patient-by-patient basis. This study may have implications for long-term biological response to orthopaedic implants, including total knee arthroplasty.</p></div>","PeriodicalId":16650,"journal":{"name":"Journal of Orthopaedic Research®","volume":"43 6","pages":"1144-1154"},"PeriodicalIF":2.3000,"publicationDate":"2025-04-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orthopaedic Research®","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/jor.26079","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Cobalt-chromium-molybdenum (CoCrMo) femoral components are widely used in total knee arthroplasty (TKA). However, recent retrospective clinical trials associate moderate adverse local tissue reactions with CoCrMo release in the knee. Additionally, gaps persist in our understanding of the fundamental corrosion processes that occur at the CoCrMo-synovial fluid interface. In this study, we investigated the electrochemical behavior of CoCrMo in human synovial fluid obtained at the time of primary TKA, using CoCrMo in phosphate-buffered saline (PBS) as a comparison. Synovial fluid was collected from 118 patients immediately before arthroplasty, then transferred to a three-electrode electrochemical cell with a wrought CoCrMo alloy working electrode. To quantify electrochemical properties, open circuit potential (OCP), electrochemical impedance spectroscopy, and linear polarization tests were run. Generally, the properties varied on a patient-by-patient basis and significantly differed (p < 0.05) from comparison tests performed in PBS. In human synovial fluid, we measured OCPs between a range of −0.38 and 0.15 V and corrosion potentials (Ecorr) between −0.95 and −0.2 V. Additionally, we reported instantaneous corrosion rates (1/Rp) spanning nearly four orders of magnitude. The variability we documented suggested that the electrochemical properties of CoCrMo implants may depend on the patient's local physiological environment, influenced by the biological and chemical components of synovial fluid.

Statement of Clinical Significance: This study shows that human synovial fluid affects the electrochemical behavior of CoCrMo, evidenced by the variation on a patient-by-patient basis. This study may have implications for long-term biological response to orthopaedic implants, including total knee arthroplasty.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
骨关节炎患者滑液改变CoCrMo的电化学特性。
钴铬钼(CoCrMo)股骨假体广泛应用于全膝关节置换术(TKA)。然而,最近的回顾性临床试验将膝关节CoCrMo释放与中度局部不良组织反应联系起来。此外,我们对cocrmo -滑液界面发生的基本腐蚀过程的理解仍然存在空白。在这项研究中,我们研究了CoCrMo在初次TKA时获得的人体滑液中的电化学行为,并以磷酸盐缓冲盐水(PBS)中的CoCrMo作为比较。在关节成形术前立即收集118例患者的滑液,然后将其转移到具有变形CoCrMo合金工作电极的三电极电化学电池中。为了量化电化学性能,进行了开路电位(OCP)、电化学阻抗谱和线性极化测试。一般来说,这些特性因患者而异,在-0.95和-0.2 V之间差异显著(p < 0.05)。此外,我们报告的瞬时腐蚀速率(1/Rp)跨越了近四个数量级。我们记录的差异表明,CoCrMo植入物的电化学特性可能取决于患者的局部生理环境,受到滑液生物和化学成分的影响。临床意义声明:本研究表明,人体滑液会影响CoCrMo的电化学行为,这可以通过患者之间的差异来证明。这项研究可能对包括全膝关节置换术在内的骨科植入物的长期生物学反应有影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Orthopaedic Research®
Journal of Orthopaedic Research® 医学-整形外科
CiteScore
6.10
自引率
3.60%
发文量
261
审稿时长
3-6 weeks
期刊介绍: The Journal of Orthopaedic Research is the forum for the rapid publication of high quality reports of new information on the full spectrum of orthopaedic research, including life sciences, engineering, translational, and clinical studies.
期刊最新文献
Bio-Compositional and Microstructural Changes in Rabbit Knee Collateral Ligaments Eight Weeks After Anterior Cruciate Ligament Transection Physeal Allograft Transfer for Physeal Bars: A Safety and Feasibility Study in a Domestic Swine Model High Demand Loading Conditions and Their Effect on Polyethylene Stresses in Lumbar Total Joint Replacement: Implications for Spine Wear Protocols The Effect of Hip Arthroplasty on Gait Function: Comparison of Ceramic-On-Ceramic Hip Resurfacing, Metal-On-Metal Hip Resurfacing, and Total Hip Arthroplasty. 2025 International Consensus Meeting on Musculoskeletal Infection: Summary From the Biofilm Workgroup on Biofilm Formation, Persistence, and Host-Environment Interactions.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1