V. Thakur, A. Jain, P. Sawant, P. Ashokkumar, L. Chaudhari, Probal Chaudhury
{"title":"Development of a desktop radiation monitoring system","authors":"V. Thakur, A. Jain, P. Sawant, P. Ashokkumar, L. Chaudhari, Probal Chaudhury","doi":"10.4103/rpe.rpe_6_21","DOIUrl":null,"url":null,"abstract":"This article presents the design and development of a digital desktop gamma radiation monitoring system using Geiger–Muller tube detector and microcontroller. The radiation detector, battery, signal processing circuits, and the microcontroller board are housed inside a tabletop wooden model with an analog watch and a digital display of radiation dose rate. The alarm level has been set at 1 μSv/h and indicated through buzzer and LED. The system has been calibrated using 137Cs standard source and has a sensitivity of 0.7364 cps/μSv/h. This model can be displayed on any table/desk or kept near a doorway to continuously monitor background radiation level, without drawing public attention, to detect any movement or presence of radiation source.","PeriodicalId":32488,"journal":{"name":"Radiation Protection and Environment","volume":"44 1","pages":"19 - 21"},"PeriodicalIF":0.0000,"publicationDate":"2021-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiation Protection and Environment","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/rpe.rpe_6_21","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
This article presents the design and development of a digital desktop gamma radiation monitoring system using Geiger–Muller tube detector and microcontroller. The radiation detector, battery, signal processing circuits, and the microcontroller board are housed inside a tabletop wooden model with an analog watch and a digital display of radiation dose rate. The alarm level has been set at 1 μSv/h and indicated through buzzer and LED. The system has been calibrated using 137Cs standard source and has a sensitivity of 0.7364 cps/μSv/h. This model can be displayed on any table/desk or kept near a doorway to continuously monitor background radiation level, without drawing public attention, to detect any movement or presence of radiation source.