角膜中央厚度的测量。

Georgiana Camburu, Mihail Zemba, Călin Petru Tătaru, Victor Lorin Purcărea
{"title":"角膜中央厚度的测量。","authors":"Georgiana Camburu,&nbsp;Mihail Zemba,&nbsp;Călin Petru Tătaru,&nbsp;Victor Lorin Purcărea","doi":"10.22336/rjo.2023.29","DOIUrl":null,"url":null,"abstract":"<p><p>We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, refractive surgery and keratoplasty. We measured the CCT using OCT, Topographer (TOPO), Ultrasonography Pachymeter (US), Specular Microscope (MS), and Non-contact Tonometer (TONO). In the analysis of the data collected from 59 patients we found the following mean values: US - 554.51 ± 29.849 μm, OCT - 548.73 ± 31.080 μm, TOPO - 553.76 ± 29.845 μm, MS - 564 ± 32.637 μm, and TONO - 538.9 ± 35.657 μm. Our results confirmed the strong correlation between techniques. <b>Abbreviations:</b> OCT = Optical Coherence Tomography, CCT = Central Corneal Thickness, TOPO = Topographer, US = Ultrasonography Pachymeter, MS = Specular Microscope, TONO = Non-contact Tonometer.</p>","PeriodicalId":21385,"journal":{"name":"Romanian journal of ophthalmology","volume":"67 2","pages":"168-174"},"PeriodicalIF":0.0000,"publicationDate":"2023-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10385715/pdf/","citationCount":"0","resultStr":"{\"title\":\"The measurement of Central Corneal Thickness.\",\"authors\":\"Georgiana Camburu,&nbsp;Mihail Zemba,&nbsp;Călin Petru Tătaru,&nbsp;Victor Lorin Purcărea\",\"doi\":\"10.22336/rjo.2023.29\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, refractive surgery and keratoplasty. We measured the CCT using OCT, Topographer (TOPO), Ultrasonography Pachymeter (US), Specular Microscope (MS), and Non-contact Tonometer (TONO). In the analysis of the data collected from 59 patients we found the following mean values: US - 554.51 ± 29.849 μm, OCT - 548.73 ± 31.080 μm, TOPO - 553.76 ± 29.845 μm, MS - 564 ± 32.637 μm, and TONO - 538.9 ± 35.657 μm. Our results confirmed the strong correlation between techniques. <b>Abbreviations:</b> OCT = Optical Coherence Tomography, CCT = Central Corneal Thickness, TOPO = Topographer, US = Ultrasonography Pachymeter, MS = Specular Microscope, TONO = Non-contact Tonometer.</p>\",\"PeriodicalId\":21385,\"journal\":{\"name\":\"Romanian journal of ophthalmology\",\"volume\":\"67 2\",\"pages\":\"168-174\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10385715/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Romanian journal of ophthalmology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.22336/rjo.2023.29\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Romanian journal of ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22336/rjo.2023.29","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

我们旨在比较五种不同的测量角膜中央厚度的设备。角膜中央厚度(CCT)是眼科的一个重要参数。它涉及各种眼部疾病的管理,如青光眼、圆锥角膜、隐形眼镜佩戴、角膜营养不良、屈光手术和角膜移植术。我们使用OCT、Topographer(TOPO)、超声Pachymeter(US)、Specular Microscope(MS)和非接触式Tonometer(TONO)测量CCT。在对59名患者的数据分析中,我们发现以下平均值:US-554.51±29.849μm,OCT-548.73±31.080μm,TOPO-553.76±29.845μm、MS-564±32.637μm和TONO-538.9±35.657μm。我们的研究结果证实了技术之间的强烈相关性。缩写:OCT=光学相干断层扫描,CCT=中央角膜厚度,TOPO=地形图,US=超声测厚仪,MS=镜面显微镜,TONO=非接触式测厚仪。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

摘要图片

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
The measurement of Central Corneal Thickness.

We aimed to compare five different devices that measure Central Corneal Thickness. The Central Corneal Thickness (CCT) is an important parameter in ophthalmology. It is involved in the management of various eye conditions such as: glaucoma, keratoconus, contact lens wearing, corneal dystrophies, refractive surgery and keratoplasty. We measured the CCT using OCT, Topographer (TOPO), Ultrasonography Pachymeter (US), Specular Microscope (MS), and Non-contact Tonometer (TONO). In the analysis of the data collected from 59 patients we found the following mean values: US - 554.51 ± 29.849 μm, OCT - 548.73 ± 31.080 μm, TOPO - 553.76 ± 29.845 μm, MS - 564 ± 32.637 μm, and TONO - 538.9 ± 35.657 μm. Our results confirmed the strong correlation between techniques. Abbreviations: OCT = Optical Coherence Tomography, CCT = Central Corneal Thickness, TOPO = Topographer, US = Ultrasonography Pachymeter, MS = Specular Microscope, TONO = Non-contact Tonometer.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Rare case report: a 26-year-old man with Eales' disease. Apert Syndrome - caveats of squint management. Assessment of the efficacy of 0.1% cyclosporine A cationic emulsion in the treatment of dry eye disease during COVID-19 pandemic. Dynamics of the association between visual and auditory functional changes in glaucoma. Preliminary results. Decreased perifoveal ganglion cell complex thickness - a first sign for macular damage in patients using hydroxychloroquine.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1