{"title":"Analisis Perbandingan Dosis Keluaran Berkas Radiasi Pesawat Co-60 Merek Theratron Phoenix Dari Perubahan Nilai Panjang Dan Lebar Persegi Panjang Pada Luas Lapangan Yang Sama","authors":"Alya Nabilla, Dian Milvita, Mursiyatun Mursiyatun","doi":"10.20961/PROSIDINGSNFA.V5I0.46589","DOIUrl":null,"url":null,"abstract":"Abstract: Research has been carried out on the comparative analysis of the radiation dose of the Co-60 with the theratron phoenix by two stages, first to analyze the output dose according to the recommendations of the TRS 398 IAEA protocol with a field of 10 cm x 10 cm, a depth of 5 cm and SSD 80. cm, which aims to calculate the output factor, the second is to analyze the ratio of the output dose of changes in the value of length and width of the rectangle on the same field with five variations of the field, the results are compared according to the quality assurance by the AAPM TG 40. The results showed that the error value in the 10 cm x 10 cm, the error value was 0.05%, 5 cm x 6 cm, the error value for each change value was 0.426%, 0, 106%, 0.530%. 5 cm x 8 cm : 0.211%, 0.210%, and 0.419%. 5 cm x 13 cm : 0.207%, 0.311%, 0.104%. 6 cm x 12 cm : 0.206%. 6 cm x 19 cm : 0.610%, 0.713%, 0.102%. The value is still within the tolerance limit of the Co-60 AAPM TG 40 quality assurance protocol 2%.Abstrak: Telah dilakukan penelitian tentang analisis perbandingan dosis keluaran berkas radiasi pesawat Co-60 merek theratron phoenix, penelitian dilakukan dalam dua tahap, pertama menganalisis dosis keluaran sesuai dengan rekomendasi protokol TRS 398 IAEA dengan luas lapangan 10 cm x 10 cm, kedalaman 5 cm dan SSD 80 cm, yang bertujuan untuk menghitung faktor keluaran yang akan digunakan pada tahap selanjutnya, kedua menganalisis perbandingan dosis keluaran dari perubahan nilai panjang dan lebar persegi panjang pada luas lapangan yang sama dengan lima variasi luas lapangan, hasilnya dibandingkan sesuai dengan quality assurance yang dikeluarkan protokol AAPM TG 40 untuk Co-60. Hasil penelitian menunjukkan nilai error atau deviasi pada luas lapangan 10 cm x 10 cm nilai error sebesar 0,05%, luas lapangan 5 cm x 6 cm nilai error terhadap masing-masing nilai perubahan sebesar 0,426%, 0, 106%, 0,530%. Luas lapangan 5 cm x 8 cm nilai error sebesar 0,211%, 0,210%, dan 0,419%. Luas lapangan 5 cm x 13 cm nilai error sebesar 0,207%, 0,311%, 0,104%. Luas lapangan 6 cm x 12 cm nilai error sebesar 0,206%. Luas lapangan 6 cm x 19 cm nilai error sebesar 0,610%, 0,713%, 0,102%. Nilai masih berada dalam batas toleransi quality assurance Co-60 protokol AAPM TG 40 yaitu 2%.","PeriodicalId":117408,"journal":{"name":"Prosiding SNFA (Seminar Nasional Fisika dan Aplikasinya)","volume":"47 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-12-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Prosiding SNFA (Seminar Nasional Fisika dan Aplikasinya)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.20961/PROSIDINGSNFA.V5I0.46589","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

摘要

摘要:本文分两个阶段对Co-60与theratron phoenix的辐射剂量进行了对比分析,首先根据TRS 398 IAEA议定书推荐的10 cm × 10 cm、5 cm深、SSD 80的输出剂量进行了分析。二是分析同一场上矩形长宽值随场的五种变化而变化的输出剂量之比,并根据AAPM TG 40的质量保证对结果进行比较。结果表明,误差值在10 cm × 10 cm,误差值分别为0.05%、5 cm × 6 cm,各变化值的误差值分别为0.426%、0、106%、0.530%。5厘米× 8厘米:0.211%、0.210%和0.419%。5厘米× 13厘米:0.207%,0.311%,0.104%。6厘米× 12厘米:0.206%。6厘米× 19厘米:0.610%,0.713%,0.102%。该值仍在Co-60 AAPM TG 40质量保证协议2%的公差范围内。Abstrak:【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】【翻译】kedua menganalis perbandingan dosis keluaran dari perubahan nilai panjang danlebar persegi panjang padluas lapangan yang sama dengan lima variasi luas lapangan hasilya dibandingkan sesuai dengan质量保证yang dikeluarkan protokol AAPM TG 40 untuk Co-60。Hasil penelitian menunjukkan nilai误差为10 cm × 10 cm,误差为0.05%,误差为5 cm × 6 cm,误差为0.0426%,误差为0.0106%,误差为0.0530%。Luas lapangan 5 cm x 8 cm nilai误差为0,211%,0,210%,dan 0,419%。Luas lapangan 5 cm x 13 cm nilai误差分别为0,207%,0,311%,0,104%。Luas lapangan 6厘米× 12厘米nilai误差为0.0206%。Luas lapangan 6 cm x 19 cm nilai误差分别为0,610%,0,713%,0,102%。Nilai masih berada dalam batas公差质量保证Co-60 protokol AAPM TG 40 yaitu 2%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Analisis Perbandingan Dosis Keluaran Berkas Radiasi Pesawat Co-60 Merek Theratron Phoenix Dari Perubahan Nilai Panjang Dan Lebar Persegi Panjang Pada Luas Lapangan Yang Sama
Abstract: Research has been carried out on the comparative analysis of the radiation dose of the Co-60 with the theratron phoenix by two stages, first to analyze the output dose according to the recommendations of the TRS 398 IAEA protocol with a field of 10 cm x 10 cm, a depth of 5 cm and SSD 80. cm, which aims to calculate the output factor, the second is to analyze the ratio of the output dose of changes in the value of length and width of the rectangle on the same field with five variations of the field, the results are compared according to the quality assurance by the AAPM TG 40. The results showed that the error value in the 10 cm x 10 cm, the error value was 0.05%, 5 cm x 6 cm, the error value for each change value was 0.426%, 0, 106%, 0.530%. 5 cm x 8 cm : 0.211%, 0.210%, and 0.419%. 5 cm x 13 cm : 0.207%, 0.311%, 0.104%. 6 cm x 12 cm : 0.206%. 6 cm x 19 cm : 0.610%, 0.713%, 0.102%. The value is still within the tolerance limit of the Co-60 AAPM TG 40 quality assurance protocol 2%.Abstrak: Telah dilakukan penelitian tentang analisis perbandingan dosis keluaran berkas radiasi pesawat Co-60 merek theratron phoenix, penelitian dilakukan dalam dua tahap, pertama menganalisis dosis keluaran sesuai dengan rekomendasi protokol TRS 398 IAEA dengan luas lapangan 10 cm x 10 cm, kedalaman 5 cm dan SSD 80 cm, yang bertujuan untuk menghitung faktor keluaran yang akan digunakan pada tahap selanjutnya, kedua menganalisis perbandingan dosis keluaran dari perubahan nilai panjang dan lebar persegi panjang pada luas lapangan yang sama dengan lima variasi luas lapangan, hasilnya dibandingkan sesuai dengan quality assurance yang dikeluarkan protokol AAPM TG 40 untuk Co-60. Hasil penelitian menunjukkan nilai error atau deviasi pada luas lapangan 10 cm x 10 cm nilai error sebesar 0,05%, luas lapangan 5 cm x 6 cm nilai error terhadap masing-masing nilai perubahan sebesar 0,426%, 0, 106%, 0,530%. Luas lapangan 5 cm x 8 cm nilai error sebesar 0,211%, 0,210%, dan 0,419%. Luas lapangan 5 cm x 13 cm nilai error sebesar 0,207%, 0,311%, 0,104%. Luas lapangan 6 cm x 12 cm nilai error sebesar 0,206%. Luas lapangan 6 cm x 19 cm nilai error sebesar 0,610%, 0,713%, 0,102%. Nilai masih berada dalam batas toleransi quality assurance Co-60 protokol AAPM TG 40 yaitu 2%.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Pengembangan Inovasi Modul Digital dengan Model POE2WE Sebagai Salah Satu Alternatif Pembelajaran Daring di Masa New Normal Sistem Kendali Penggerak Motor Stepper Pada Orbital Welding Menggunakan Perangkat Lunak LabVIEW Identification the Need of Electronic-Based Physics Teaching Materials for Increasing Problem Solving Ability in the 21st Century Pengaruh Variasi Dimensi Superkonduktor Berbentuk Persegi Panjang Terhadap Medan Kritis Pada Pada κ = 1,5 Analisis Perbandingan Dosis Keluaran Berkas Radiasi Pesawat Co-60 Merek Theratron Phoenix Dari Perubahan Nilai Panjang Dan Lebar Persegi Panjang Pada Luas Lapangan Yang Sama
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1