[临床经验]笼式电极EMT 588 K。阈值-电流消耗-术后并发症[作者简介]。

H W Präuer, M Lampadius
{"title":"[临床经验]笼式电极EMT 588 K。阈值-电流消耗-术后并发症[作者简介]。","authors":"H W Präuer,&nbsp;M Lampadius","doi":"10.1055/s-0028-1096594","DOIUrl":null,"url":null,"abstract":"<p><p>The Lagergren cage electrode Elema EMT 588 K was implanted in 26 male and 21 female patients for permanent cardiac stimulation. 29 of the 47 cage electrodes were fitted with a platinum and 18 with an Elgiloy head. Acute threshold in the platinum group was 0,52 +/- 0,15 volts (x +/- s), the equivalent electrode resistance being 565 +/- 135 ohms. The threshold mean value for the Elgiloy cage electrodes was shown to be 0,36 +/- 0,11 volts, while the resistance values measured respectively 410 +/- 44 and 418 +/- 23 ohms depending on the method of measurement. In the postoperative period (up to 16 months) the following complications were observed: 3 dislocations (6,38%) and one case of rise in threshold (2,12%). The total correction rate (8,5%) corresponds almost exactly to that recorded by the author in a personal series of 400 electrode implantations of various designs. The cage electrode offers no advantages in comparison to other small surface electrodes, however, the relatively higher current consumption caused by low total electrode resistance must be regarded as disadvantages. Experience with the cage electrode has shown that the transitional resistance depends not only on the surface, but also essentially on the material and shape of the electrode tip.</p>","PeriodicalId":22981,"journal":{"name":"Thoraxchirurgie, vaskulare Chirurgie","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1978-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1055/s-0028-1096594","citationCount":"0","resultStr":"{\"title\":\"[Clinical experience with the cage electrode EMT 588 K. Threshold--current consumption--postoperative complications (author's transl)].\",\"authors\":\"H W Präuer,&nbsp;M Lampadius\",\"doi\":\"10.1055/s-0028-1096594\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The Lagergren cage electrode Elema EMT 588 K was implanted in 26 male and 21 female patients for permanent cardiac stimulation. 29 of the 47 cage electrodes were fitted with a platinum and 18 with an Elgiloy head. Acute threshold in the platinum group was 0,52 +/- 0,15 volts (x +/- s), the equivalent electrode resistance being 565 +/- 135 ohms. The threshold mean value for the Elgiloy cage electrodes was shown to be 0,36 +/- 0,11 volts, while the resistance values measured respectively 410 +/- 44 and 418 +/- 23 ohms depending on the method of measurement. In the postoperative period (up to 16 months) the following complications were observed: 3 dislocations (6,38%) and one case of rise in threshold (2,12%). The total correction rate (8,5%) corresponds almost exactly to that recorded by the author in a personal series of 400 electrode implantations of various designs. The cage electrode offers no advantages in comparison to other small surface electrodes, however, the relatively higher current consumption caused by low total electrode resistance must be regarded as disadvantages. Experience with the cage electrode has shown that the transitional resistance depends not only on the surface, but also essentially on the material and shape of the electrode tip.</p>\",\"PeriodicalId\":22981,\"journal\":{\"name\":\"Thoraxchirurgie, vaskulare Chirurgie\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1978-02-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1055/s-0028-1096594\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Thoraxchirurgie, vaskulare Chirurgie\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1055/s-0028-1096594\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Thoraxchirurgie, vaskulare Chirurgie","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1055/s-0028-1096594","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

将Elema EMT 588 K拉格伦笼电极植入26例男性患者和21例女性患者进行永久性心脏刺激。47个笼状电极中有29个配有铂电极,18个配有Elgiloy电极头。铂组急性阈值为0.52 +/- 0.15伏(x +/- s),等效电极电阻为565 +/- 135欧姆。Elgiloy笼电极的阈值平均值显示为0,36 +/- 0,11伏,而电阻值分别测量为410 +/- 44和418 +/- 23欧姆,具体取决于测量方法。术后(16个月)出现以下并发症:脱位3例(6.38%),门限升高1例(2.12%)。总校正率(8.5%)几乎完全符合作者在各种设计的400个电极植入的个人系列中记录的。笼状电极与其他小表面电极相比没有优势,但由于电极总电阻低而导致的相对较高的电流消耗必须被视为缺点。笼状电极的经验表明,过渡电阻不仅取决于表面,而且本质上取决于电极尖端的材料和形状。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[Clinical experience with the cage electrode EMT 588 K. Threshold--current consumption--postoperative complications (author's transl)].

The Lagergren cage electrode Elema EMT 588 K was implanted in 26 male and 21 female patients for permanent cardiac stimulation. 29 of the 47 cage electrodes were fitted with a platinum and 18 with an Elgiloy head. Acute threshold in the platinum group was 0,52 +/- 0,15 volts (x +/- s), the equivalent electrode resistance being 565 +/- 135 ohms. The threshold mean value for the Elgiloy cage electrodes was shown to be 0,36 +/- 0,11 volts, while the resistance values measured respectively 410 +/- 44 and 418 +/- 23 ohms depending on the method of measurement. In the postoperative period (up to 16 months) the following complications were observed: 3 dislocations (6,38%) and one case of rise in threshold (2,12%). The total correction rate (8,5%) corresponds almost exactly to that recorded by the author in a personal series of 400 electrode implantations of various designs. The cage electrode offers no advantages in comparison to other small surface electrodes, however, the relatively higher current consumption caused by low total electrode resistance must be regarded as disadvantages. Experience with the cage electrode has shown that the transitional resistance depends not only on the surface, but also essentially on the material and shape of the electrode tip.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
[Fresh autologus blood transfusion during open heart surgery (author's transl)]. [Longitudinal rupture of the aortic arche (author's transl)]. [Behaviour of ATP and lactate in human papillary muscle during profound hypothermia and injection cardioplegia with magnesium-asparatate-procaine (author's transl)]. [Biochemical and morphological investigations on the human heart in cardioplegia and profound hypothermia (author's transl)]. [Some psychological predictors for psychosis after open heart surgery (author's transl)].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1