{"title":"Penentuan Nilai Faktor Mesin Pesawat Sinar-X Radiografi Digital Merek Shimadzu di RSUD Dr. Kanujoso Djatiwibowo Balikpapan","authors":"Mamba'ul Fitriyana, S. Muliyono, Kadek Subagiada","doi":"10.30872/ppj.v1i1.563","DOIUrl":null,"url":null,"abstract":"The research on the determining machine factor (k) of Shimadzu digital radiograph X-ray machine have been done at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan. Machine factor should be required in the calculation of radiation exposure which produced by the X-ray machine. Measurement data that have been used as independent variables were current-time, focus distance to the detector and tube voltage variations of 80 kV to 100 kV, while the dependent variable was the S value. Determination of the k value was conducted by weighted linear regression between V2 and using the results of research by Seibert and Morin (2011) for the condition of calibrated X-ray machine. Thus, the reseach has obtained the machine factor of the Shimadzu digital radiograph X-ray machine at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan.","PeriodicalId":221598,"journal":{"name":"Progressive Physics Journal","volume":"67 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-01-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Progressive Physics Journal","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.30872/ppj.v1i1.563","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
岛津数字x光机的决定机器因子(k)的研究已在RSUD Dr. Kanujoso Djatiwibowo Balikpapan进行。在计算x光机产生的辐射暴露量时,应考虑机器因素。测量数据以电流时间、到探测器的聚焦距离、80 ~ 100 kV的管电压变化为自变量,因变量为S值。k值的确定采用V2之间的加权线性回归,并采用Seibert和Morin(2011)对标定x射线机条件的研究结果。因此,该研究获得了RSUD Dr. Kanujoso Djatiwibowo Balikpapan的岛津数字x射线机的机器因子。
Penentuan Nilai Faktor Mesin Pesawat Sinar-X Radiografi Digital Merek Shimadzu di RSUD Dr. Kanujoso Djatiwibowo Balikpapan
The research on the determining machine factor (k) of Shimadzu digital radiograph X-ray machine have been done at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan. Machine factor should be required in the calculation of radiation exposure which produced by the X-ray machine. Measurement data that have been used as independent variables were current-time, focus distance to the detector and tube voltage variations of 80 kV to 100 kV, while the dependent variable was the S value. Determination of the k value was conducted by weighted linear regression between V2 and using the results of research by Seibert and Morin (2011) for the condition of calibrated X-ray machine. Thus, the reseach has obtained the machine factor of the Shimadzu digital radiograph X-ray machine at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan.