H. Herlina, M. A. Zulfikar, Lasmaryna Sirumapea, B. Buchari
{"title":"铂和氢氧化钴修饰铂电极作为阿莫西林电化学氧化传感器","authors":"H. Herlina, M. A. Zulfikar, Lasmaryna Sirumapea, B. Buchari","doi":"10.19109/alkimia.v6i1.12014","DOIUrl":null,"url":null,"abstract":"Antibiotics have recently gained in popularity due to their usage in medical treatment and the process of removing them from the environment. Amoxicillin is one of the antibiotics that constitutes the study's subject. On a Pt disc electrode and a cobalt hydroxide modified-Pt electrode, the scan rate and pH in the electrochemical oxidation of amoxicillin were investigated. The voltammetry measurement's current peak revealed that changing the electrode surface could increase the electrochemical response and sensitivity of the working electrodes. The Pt/Co(OH)2 modification working electrode had a high sensitivity in the electro-oxidation determination of amoxicillin, with a linear range of the sensor of 20 to 80 M and a limit of detection of 7.15 M for the Pt disc electrode and 3.64 M for the cobalt hydroxide modified Pt electrode. The findings of determination in real samples with electro-oxidation using a modified electrode were in good agreement with a confidence level of 95 percent, according to the comparing method with HPLC.","PeriodicalId":196624,"journal":{"name":"ALKIMIA : Jurnal Ilmu Kimia dan Terapan","volume":"21 3","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-09-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Platinum and Cobalt hydroxide – modified Platinum Electrode as Sensor for Electrochemical Oxidation of Amoxicillin\",\"authors\":\"H. Herlina, M. A. Zulfikar, Lasmaryna Sirumapea, B. Buchari\",\"doi\":\"10.19109/alkimia.v6i1.12014\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Antibiotics have recently gained in popularity due to their usage in medical treatment and the process of removing them from the environment. Amoxicillin is one of the antibiotics that constitutes the study's subject. On a Pt disc electrode and a cobalt hydroxide modified-Pt electrode, the scan rate and pH in the electrochemical oxidation of amoxicillin were investigated. The voltammetry measurement's current peak revealed that changing the electrode surface could increase the electrochemical response and sensitivity of the working electrodes. The Pt/Co(OH)2 modification working electrode had a high sensitivity in the electro-oxidation determination of amoxicillin, with a linear range of the sensor of 20 to 80 M and a limit of detection of 7.15 M for the Pt disc electrode and 3.64 M for the cobalt hydroxide modified Pt electrode. The findings of determination in real samples with electro-oxidation using a modified electrode were in good agreement with a confidence level of 95 percent, according to the comparing method with HPLC.\",\"PeriodicalId\":196624,\"journal\":{\"name\":\"ALKIMIA : Jurnal Ilmu Kimia dan Terapan\",\"volume\":\"21 3\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-09-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"ALKIMIA : Jurnal Ilmu Kimia dan Terapan\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.19109/alkimia.v6i1.12014\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"ALKIMIA : Jurnal Ilmu Kimia dan Terapan","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.19109/alkimia.v6i1.12014","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
抗生素最近因其在医疗中的使用以及从环境中去除它们的过程而受到欢迎。阿莫西林是构成该研究主题的抗生素之一。在Pt圆盘电极和氢氧化钴修饰的Pt电极上,研究了阿莫西林电化学氧化的扫描速率和pH值。伏安法测量的电流峰值表明,改变电极表面可以提高工作电极的电化学响应和灵敏度。Pt/Co(OH)2修饰工作电极对阿莫西林电氧化测定具有较高的灵敏度,传感器线性范围为20 ~ 80 M, Pt圆盘电极的检出限为7.15 M,氢氧化钴修饰Pt电极的检出限为3.64 M。采用改良电极对实际样品进行电氧化测定的结果与HPLC法比较,置信水平为95%。
Platinum and Cobalt hydroxide – modified Platinum Electrode as Sensor for Electrochemical Oxidation of Amoxicillin
Antibiotics have recently gained in popularity due to their usage in medical treatment and the process of removing them from the environment. Amoxicillin is one of the antibiotics that constitutes the study's subject. On a Pt disc electrode and a cobalt hydroxide modified-Pt electrode, the scan rate and pH in the electrochemical oxidation of amoxicillin were investigated. The voltammetry measurement's current peak revealed that changing the electrode surface could increase the electrochemical response and sensitivity of the working electrodes. The Pt/Co(OH)2 modification working electrode had a high sensitivity in the electro-oxidation determination of amoxicillin, with a linear range of the sensor of 20 to 80 M and a limit of detection of 7.15 M for the Pt disc electrode and 3.64 M for the cobalt hydroxide modified Pt electrode. The findings of determination in real samples with electro-oxidation using a modified electrode were in good agreement with a confidence level of 95 percent, according to the comparing method with HPLC.