{"title":"“tct - 700s”x射线计算机断层扫描性能评价。2]。","authors":"Y Hayakawa,&nbsp;M Wakoh,&nbsp;K Yamamoto,&nbsp;K Nishikawa,&nbsp;K Enomoto,&nbsp;K Tamura,&nbsp;C Hagiwara,&nbsp;H Ueno,&nbsp;H Fujimori,&nbsp;Y Kohsuge","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Unlabelled: </strong>The performance of the rotate/rotate CT scanner TCT-700 S (Toshiba) was evaluated on the basis of the item entitled \"User's initial acceptance tests (phase II)\" set forth in \"Standard of performance evaluation for x-ray computed tomography (second recommendation)\" proposed by Takenaka et al. (1982). This second report in the series describes measurements of exposure and the relation between exposure and such other performance aspects as noise and contrast resolution. The exposure-performance recommendation proposes measurements of average exposure in the scanning field and a radial exposure profile.</p><p><strong>Results: </strong>1. In the case of multiple scanning, average exposures in the field were uniform because all surfaces of the perpendicular object were oriented toward the incidence of the x-ray beam. At a minimum scanning speed of 1.8 second, however, a discrepancy of 3 times occurred in exposures observed reciprocally from the left and right sides of the object. This phenomenon indicates that only one-half of the object surface is oriented toward the x-ray beam and that the x-ray tube and detectors rotate in mutually opposite directions. In the case of multiple scannings, when slice thicknesses were 10 mm, 5 mm and 2 mm, average exposures were almost uniform. But with a slice thickness of 1 mm, average exposure was twice as great. This measurement indicates inadequate collimation in the case of a 1 mm slice thickness. These average exposures were roughly proportional to the mAs (product of tube current and scan speed) in the range between a minimum of 100 mAs and a maximum of 1200 mAs. 2. Noise increased in proportion to decrease of average exposure.(ABSTRACT TRUNCATED AT 250 WORDS)</p>","PeriodicalId":76540,"journal":{"name":"Shika gakuho. Dental science reports","volume":"89 11","pages":"1699-707"},"PeriodicalIF":0.0000,"publicationDate":"1989-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Performance evaluation of X-ray computed tomography \\\"TCT-700 S\\\". 2].\",\"authors\":\"Y Hayakawa,&nbsp;M Wakoh,&nbsp;K Yamamoto,&nbsp;K Nishikawa,&nbsp;K Enomoto,&nbsp;K Tamura,&nbsp;C Hagiwara,&nbsp;H Ueno,&nbsp;H Fujimori,&nbsp;Y Kohsuge\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Unlabelled: </strong>The performance of the rotate/rotate CT scanner TCT-700 S (Toshiba) was evaluated on the basis of the item entitled \\\"User's initial acceptance tests (phase II)\\\" set forth in \\\"Standard of performance evaluation for x-ray computed tomography (second recommendation)\\\" proposed by Takenaka et al. (1982). This second report in the series describes measurements of exposure and the relation between exposure and such other performance aspects as noise and contrast resolution. The exposure-performance recommendation proposes measurements of average exposure in the scanning field and a radial exposure profile.</p><p><strong>Results: </strong>1. In the case of multiple scanning, average exposures in the field were uniform because all surfaces of the perpendicular object were oriented toward the incidence of the x-ray beam. At a minimum scanning speed of 1.8 second, however, a discrepancy of 3 times occurred in exposures observed reciprocally from the left and right sides of the object. This phenomenon indicates that only one-half of the object surface is oriented toward the x-ray beam and that the x-ray tube and detectors rotate in mutually opposite directions. In the case of multiple scannings, when slice thicknesses were 10 mm, 5 mm and 2 mm, average exposures were almost uniform. But with a slice thickness of 1 mm, average exposure was twice as great. This measurement indicates inadequate collimation in the case of a 1 mm slice thickness. These average exposures were roughly proportional to the mAs (product of tube current and scan speed) in the range between a minimum of 100 mAs and a maximum of 1200 mAs. 2. Noise increased in proportion to decrease of average exposure.(ABSTRACT TRUNCATED AT 250 WORDS)</p>\",\"PeriodicalId\":76540,\"journal\":{\"name\":\"Shika gakuho. Dental science reports\",\"volume\":\"89 11\",\"pages\":\"1699-707\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1989-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Shika gakuho. Dental science reports\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Shika gakuho. Dental science reports","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

未标记:旋转/旋转CT扫描仪tct - 700s(东芝)的性能根据Takenaka等人(1982年)提出的“x射线计算机断层扫描性能评估标准(第二次建议)”中规定的题为“用户初始验收测试(第二阶段)”的项目进行了评估。本系列的第二份报告描述了曝光的测量以及曝光与其他性能方面(如噪声和对比度分辨率)之间的关系。曝光性能建议建议测量扫描场的平均曝光和径向曝光剖面。结果:1。在多次扫描的情况下,由于垂直物体的所有表面都朝向x射线束的入射方向,因此场中的平均曝光是均匀的。然而,在1.8秒的最小扫描速度下,从物体的左右两侧分别观察到的曝光出现了3倍的差异。这种现象表明,只有一半的物体表面面向x射线束,x射线管和探测器在相互相反的方向上旋转。在多次扫描的情况下,当切片厚度为10 mm、5 mm和2 mm时,平均曝光几乎均匀。但当切片厚度为1毫米时,平均曝光量为其两倍。这一测量表明,在1毫米的切片厚度的情况下,准直不足。这些平均暴露大致与最小100 mAs和最大1200 mAs之间的mAs(管电流和扫描速度的乘积)成正比。2. 噪声的增加与平均暴露量的减少成正比。(摘要删节250字)
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[Performance evaluation of X-ray computed tomography "TCT-700 S". 2].

Unlabelled: The performance of the rotate/rotate CT scanner TCT-700 S (Toshiba) was evaluated on the basis of the item entitled "User's initial acceptance tests (phase II)" set forth in "Standard of performance evaluation for x-ray computed tomography (second recommendation)" proposed by Takenaka et al. (1982). This second report in the series describes measurements of exposure and the relation between exposure and such other performance aspects as noise and contrast resolution. The exposure-performance recommendation proposes measurements of average exposure in the scanning field and a radial exposure profile.

Results: 1. In the case of multiple scanning, average exposures in the field were uniform because all surfaces of the perpendicular object were oriented toward the incidence of the x-ray beam. At a minimum scanning speed of 1.8 second, however, a discrepancy of 3 times occurred in exposures observed reciprocally from the left and right sides of the object. This phenomenon indicates that only one-half of the object surface is oriented toward the x-ray beam and that the x-ray tube and detectors rotate in mutually opposite directions. In the case of multiple scannings, when slice thicknesses were 10 mm, 5 mm and 2 mm, average exposures were almost uniform. But with a slice thickness of 1 mm, average exposure was twice as great. This measurement indicates inadequate collimation in the case of a 1 mm slice thickness. These average exposures were roughly proportional to the mAs (product of tube current and scan speed) in the range between a minimum of 100 mAs and a maximum of 1200 mAs. 2. Noise increased in proportion to decrease of average exposure.(ABSTRACT TRUNCATED AT 250 WORDS)

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
[Study of structural changes in canine maxillary trabecular bone after tooth extraction]. [A study of the internal structure of the Japanese edentulous mandible]. [Oriental psychosomatic medical therapy for vertigo]. [Experimental analytical electron microscopic studies on the quantitative analysis of elemental concentrations in biological thin specimens and its application to dental science]. [Experimental studies on healing process of plication technique at postdiscal connective tissue of TMJ in Macaca fuscata].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1