{"title":"使用扩散采样器监测十亿分之一的甲醛水平。","authors":"J O Levin, R Lindahl, K Andersson","doi":"10.1080/08940630.1989.10466506","DOIUrl":null,"url":null,"abstract":"<p><p>A diffusive sampler for formaldehyde originally designed for use in personal monitoring of worker exposure has been evaluated for static measurement of low formaldehyde levels in indoor air. The sampler consists of a 37-mm glass fiber filter impregnated with 2,4-dinitrophenylhydrazine and phosphoric acid and mounted in a polystyrene filter cassette. Formaldehyde is sampled by controlled diffusion and subsequent hydrazone formation on the filter. The hydrazone is eluted from the filter with acetonitrile and analyzed by liquid chromatography and a UV detector. The diffusive sampler was evaluated for static (area) sampling in several industries, offices, and homes with formaldehyde levels of 6-200 ppb. Results from diffusive samplers were compared with results from pumped samplers. Even at low wind velocities (less than 0.01 m.s-1) there was good agreement between pumped and diffusive samples. The sensitivity will be 1 ppb in a 24-h sample, making the sampler especially useful for indoor air monitoring of low formaldehyde levels.</p>","PeriodicalId":77731,"journal":{"name":"JAPCA","volume":"39 1","pages":"44-7"},"PeriodicalIF":0.0000,"publicationDate":"1989-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1080/08940630.1989.10466506","citationCount":"20","resultStr":"{\"title\":\"Monitoring of parts-per-billion levels of formaldehyde using a diffusive sampler.\",\"authors\":\"J O Levin, R Lindahl, K Andersson\",\"doi\":\"10.1080/08940630.1989.10466506\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A diffusive sampler for formaldehyde originally designed for use in personal monitoring of worker exposure has been evaluated for static measurement of low formaldehyde levels in indoor air. The sampler consists of a 37-mm glass fiber filter impregnated with 2,4-dinitrophenylhydrazine and phosphoric acid and mounted in a polystyrene filter cassette. Formaldehyde is sampled by controlled diffusion and subsequent hydrazone formation on the filter. The hydrazone is eluted from the filter with acetonitrile and analyzed by liquid chromatography and a UV detector. The diffusive sampler was evaluated for static (area) sampling in several industries, offices, and homes with formaldehyde levels of 6-200 ppb. Results from diffusive samplers were compared with results from pumped samplers. Even at low wind velocities (less than 0.01 m.s-1) there was good agreement between pumped and diffusive samples. The sensitivity will be 1 ppb in a 24-h sample, making the sampler especially useful for indoor air monitoring of low formaldehyde levels.</p>\",\"PeriodicalId\":77731,\"journal\":{\"name\":\"JAPCA\",\"volume\":\"39 1\",\"pages\":\"44-7\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1989-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1080/08940630.1989.10466506\",\"citationCount\":\"20\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"JAPCA\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1080/08940630.1989.10466506\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"JAPCA","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1080/08940630.1989.10466506","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 20
摘要
最初设计用于工人接触情况个人监测的甲醛扩散采样器已被评估用于室内空气中低甲醛水平的静态测量。采样器包括一个37毫米玻璃纤维过滤器浸渍2,4-二硝基苯肼和磷酸,并安装在聚苯乙烯过滤盒。通过控制扩散和随后在过滤器上形成腙来取样甲醛。用乙腈从过滤器中洗脱腙,用液相色谱法和紫外检测器进行分析。在甲醛水平为6-200 ppb的几个工业、办公室和家庭中,对扩散采样器进行了静态(区域)采样评估。将扩散采样器的结果与泵送采样器的结果进行了比较。即使在低风速下(小于0.01 ms -1),泵送样品和扩散样品之间也有很好的一致性。灵敏度将为1 ppb在一个24小时的样品,使采样器特别有用的室内空气监测低甲醛水平。
Monitoring of parts-per-billion levels of formaldehyde using a diffusive sampler.
A diffusive sampler for formaldehyde originally designed for use in personal monitoring of worker exposure has been evaluated for static measurement of low formaldehyde levels in indoor air. The sampler consists of a 37-mm glass fiber filter impregnated with 2,4-dinitrophenylhydrazine and phosphoric acid and mounted in a polystyrene filter cassette. Formaldehyde is sampled by controlled diffusion and subsequent hydrazone formation on the filter. The hydrazone is eluted from the filter with acetonitrile and analyzed by liquid chromatography and a UV detector. The diffusive sampler was evaluated for static (area) sampling in several industries, offices, and homes with formaldehyde levels of 6-200 ppb. Results from diffusive samplers were compared with results from pumped samplers. Even at low wind velocities (less than 0.01 m.s-1) there was good agreement between pumped and diffusive samples. The sensitivity will be 1 ppb in a 24-h sample, making the sampler especially useful for indoor air monitoring of low formaldehyde levels.