在功能齐全的商用显微镜底座上使用方便的斜面显微镜

IF 2.9 2区 医学 Q2 BIOCHEMICAL RESEARCH METHODS Biomedical optics express Pub Date : 2024-03-14 DOI:10.1364/boe.518856
George Sirinakis, Edward S. Allgeyer, Dmitry Nashchekin, and Daniel St. Johnston
{"title":"在功能齐全的商用显微镜底座上使用方便的斜面显微镜","authors":"George Sirinakis, Edward S. Allgeyer, Dmitry Nashchekin, and Daniel St. Johnston","doi":"10.1364/boe.518856","DOIUrl":null,"url":null,"abstract":"In this work we present an oblique plane microscope designed to work seamlessly with a commercially available microscope base. To support all the functionality offered by the microscope base, where the position of the objective lens is not fixed, we adopted a two-mirror scanning geometry that can compensate for changes to the position of the objective lens during routine microscope operation. We showed that within a ± 1 mm displacement range of the 100X, 1.35 NA objective lens away from its designed position, the PSF size increased by <3% and <11% in the lateral and axial dimensions, respectively, while the error in magnification was <0.5% within volumes extending ± 10 µm about the focal plane. Compared to the more traditional scan-lens/galvo-mirror combination, the two-mirror scanning geometry offers higher light efficiency and a more compact footprint, which could be beneficial to all OPM designs regardless of the use of a commercial base or not.","PeriodicalId":8969,"journal":{"name":"Biomedical optics express","volume":null,"pages":null},"PeriodicalIF":2.9000,"publicationDate":"2024-03-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"User-friendly oblique plane microscopy on a fully functional commercially available microscope base\",\"authors\":\"George Sirinakis, Edward S. Allgeyer, Dmitry Nashchekin, and Daniel St. Johnston\",\"doi\":\"10.1364/boe.518856\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In this work we present an oblique plane microscope designed to work seamlessly with a commercially available microscope base. To support all the functionality offered by the microscope base, where the position of the objective lens is not fixed, we adopted a two-mirror scanning geometry that can compensate for changes to the position of the objective lens during routine microscope operation. We showed that within a ± 1 mm displacement range of the 100X, 1.35 NA objective lens away from its designed position, the PSF size increased by <3% and <11% in the lateral and axial dimensions, respectively, while the error in magnification was <0.5% within volumes extending ± 10 µm about the focal plane. Compared to the more traditional scan-lens/galvo-mirror combination, the two-mirror scanning geometry offers higher light efficiency and a more compact footprint, which could be beneficial to all OPM designs regardless of the use of a commercial base or not.\",\"PeriodicalId\":8969,\"journal\":{\"name\":\"Biomedical optics express\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-03-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biomedical optics express\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1364/boe.518856\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"BIOCHEMICAL RESEARCH METHODS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomedical optics express","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1364/boe.518856","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0

摘要

在这项工作中,我们提出了一种斜平面显微镜,其设计可与市售显微镜底座无缝配合。为了支持显微镜底座提供的所有功能(其中物镜的位置并不固定),我们采用了双镜扫描几何结构,可以在显微镜常规操作过程中补偿物镜位置的变化。我们的研究表明,在 100 倍、1.35 NA 物镜偏离设计位置 ± 1 mm 的位移范围内,PSF 尺寸在横向和轴向分别增加了 3% 和 11%,而在焦平面周围 ± 10 µm 的范围内,放大率误差为 0.5%。与更传统的扫描镜头/巨像镜组合相比,双镜扫描几何具有更高的光效和更紧凑的占地面积,这对所有 OPM 设计都有好处,无论是否使用商用底座。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
User-friendly oblique plane microscopy on a fully functional commercially available microscope base
In this work we present an oblique plane microscope designed to work seamlessly with a commercially available microscope base. To support all the functionality offered by the microscope base, where the position of the objective lens is not fixed, we adopted a two-mirror scanning geometry that can compensate for changes to the position of the objective lens during routine microscope operation. We showed that within a ± 1 mm displacement range of the 100X, 1.35 NA objective lens away from its designed position, the PSF size increased by <3% and <11% in the lateral and axial dimensions, respectively, while the error in magnification was <0.5% within volumes extending ± 10 µm about the focal plane. Compared to the more traditional scan-lens/galvo-mirror combination, the two-mirror scanning geometry offers higher light efficiency and a more compact footprint, which could be beneficial to all OPM designs regardless of the use of a commercial base or not.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Biomedical optics express
Biomedical optics express BIOCHEMICAL RESEARCH METHODS-OPTICS
CiteScore
6.80
自引率
11.80%
发文量
633
审稿时长
1 months
期刊介绍: The journal''s scope encompasses fundamental research, technology development, biomedical studies and clinical applications. BOEx focuses on the leading edge topics in the field, including: Tissue optics and spectroscopy Novel microscopies Optical coherence tomography Diffuse and fluorescence tomography Photoacoustic and multimodal imaging Molecular imaging and therapies Nanophotonic biosensing Optical biophysics/photobiology Microfluidic optical devices Vision research.
期刊最新文献
Through-focus performance and off-axis effects in aspheric monofocal intraocular lenses Development of the international standard for functional near-infrared spectroscopy as a medical equipment A naturalistic approach to investigate the neural correlates of a laundry cycle with and without fragrance Microscope Integrated MHz Optical Coherence Tomography System for Neurosurgery: Development and Clinical In-Vivo Imaging Optical directional differential operation enabled visual chirality detection
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1