在 Accu-chek Inform II 上验证血糖测试方法

Thi Ngoc Lan Nguyen, Hoang Bich Nga Le, Thanh Huyen Tran, Thi Thuy Quynh Tran
{"title":"在 Accu-chek Inform II 上验证血糖测试方法","authors":"Thi Ngoc Lan Nguyen, Hoang Bich Nga Le, Thanh Huyen Tran, Thi Thuy Quynh Tran","doi":"10.31276/vjst.66(1).68-74","DOIUrl":null,"url":null,"abstract":"Blood glucose testing, including capillary blood glucose, is the recommended test for the diagnosis and monitoring of diabetes treatment. Currently, capillary blood glucose testing is performed on point-of-care testing (POCT) with reasonable cost, fast time, convenience, supporting timely diagnosis and treatment. However, in order to select POCT devices that meet quality standards, it is necessary to strictly control the laboratory. Object and method: Confirmation of the capillary blood sugar testing method on the Accu-chek Inform II (Acc-II) machine including the following indicators: accuracy, precision, linear range, comparison of test results according to the recommendations of the Clinical & Laboratory Standards Institute (CLSI) (EP15A3, EP06A, EP09A3). The authors compared results on the testing machine in the central laboratory (Cobas 8000) of blood glucose testing on the POCT machine with the results of whole capillary blood, whole venous blood, and plasma venous samples. Results: Precision, accuracy, and linearity were confirmed as published. The linear range of the method is from 1.6 to 31 mmol/l. The difference between capillary blood glucose pairs and plasma venous glucose, whole blood venous glucose on Accu-chek Inform II and plasma venous blood glucose on Cobas 8000 machine are 6.89, -8.62, and -10.97%, respectively.","PeriodicalId":18650,"journal":{"name":"Ministry of Science and Technology, Vietnam","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-01-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Validation of the blood glucose testing method on the Accu-chek Inform II\",\"authors\":\"Thi Ngoc Lan Nguyen, Hoang Bich Nga Le, Thanh Huyen Tran, Thi Thuy Quynh Tran\",\"doi\":\"10.31276/vjst.66(1).68-74\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Blood glucose testing, including capillary blood glucose, is the recommended test for the diagnosis and monitoring of diabetes treatment. Currently, capillary blood glucose testing is performed on point-of-care testing (POCT) with reasonable cost, fast time, convenience, supporting timely diagnosis and treatment. However, in order to select POCT devices that meet quality standards, it is necessary to strictly control the laboratory. Object and method: Confirmation of the capillary blood sugar testing method on the Accu-chek Inform II (Acc-II) machine including the following indicators: accuracy, precision, linear range, comparison of test results according to the recommendations of the Clinical & Laboratory Standards Institute (CLSI) (EP15A3, EP06A, EP09A3). The authors compared results on the testing machine in the central laboratory (Cobas 8000) of blood glucose testing on the POCT machine with the results of whole capillary blood, whole venous blood, and plasma venous samples. Results: Precision, accuracy, and linearity were confirmed as published. The linear range of the method is from 1.6 to 31 mmol/l. The difference between capillary blood glucose pairs and plasma venous glucose, whole blood venous glucose on Accu-chek Inform II and plasma venous blood glucose on Cobas 8000 machine are 6.89, -8.62, and -10.97%, respectively.\",\"PeriodicalId\":18650,\"journal\":{\"name\":\"Ministry of Science and Technology, Vietnam\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Ministry of Science and Technology, Vietnam\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.31276/vjst.66(1).68-74\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ministry of Science and Technology, Vietnam","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31276/vjst.66(1).68-74","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

血糖检测(包括毛细血管血糖)是诊断和监测糖尿病治疗的推荐检测方法。目前,毛细血管血糖检测在床旁检测(POCT)上进行,具有成本合理、时间快、方便快捷、支持及时诊断和治疗等优点。然而,为了选择符合质量标准的 POCT 设备,有必要严格控制实验室。对象与方法:根据临床与实验室标准协会(CLSI)的建议(EP15A3、EP06A、EP09A3),对 Accu-chek Inform II(Acc-II)机上的毛细血管血糖检测方法进行确认,包括以下指标:准确度、精密度、线性范围、检测结果比较。作者将中心实验室检测机(Cobas 8000)上的 POCT 机血糖检测结果与全毛细血管血、全静脉血和血浆静脉样本的结果进行了比较。结果精确度、准确度和线性度与公布的结果一致。该方法的线性范围为 1.6 至 31 毫摩尔/升。毛细血管血糖对与血浆静脉血糖、Accu-chek Inform II 全血静脉血糖与 Cobas 8000 血浆静脉血糖的差值分别为 6.89%、-8.62% 和 -10.97%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Validation of the blood glucose testing method on the Accu-chek Inform II
Blood glucose testing, including capillary blood glucose, is the recommended test for the diagnosis and monitoring of diabetes treatment. Currently, capillary blood glucose testing is performed on point-of-care testing (POCT) with reasonable cost, fast time, convenience, supporting timely diagnosis and treatment. However, in order to select POCT devices that meet quality standards, it is necessary to strictly control the laboratory. Object and method: Confirmation of the capillary blood sugar testing method on the Accu-chek Inform II (Acc-II) machine including the following indicators: accuracy, precision, linear range, comparison of test results according to the recommendations of the Clinical & Laboratory Standards Institute (CLSI) (EP15A3, EP06A, EP09A3). The authors compared results on the testing machine in the central laboratory (Cobas 8000) of blood glucose testing on the POCT machine with the results of whole capillary blood, whole venous blood, and plasma venous samples. Results: Precision, accuracy, and linearity were confirmed as published. The linear range of the method is from 1.6 to 31 mmol/l. The difference between capillary blood glucose pairs and plasma venous glucose, whole blood venous glucose on Accu-chek Inform II and plasma venous blood glucose on Cobas 8000 machine are 6.89, -8.62, and -10.97%, respectively.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Quantification of catalpol in root of Rehmannia glutinosa varieties 19 collected in Phu Tho province by high-performance liquid chromatography Morphology, anatomy, and quantitative determination of corosolic acid in Lagerstroemia calyculata Kurz in the Southeast region, Vietnam Prediction of geomechanical changes in faulted rock mass around underground structures subjected to earthquakes Synthesis of TiO2 by hydrothermal method using deep eutectic solvent for application in dye-sensitised solar cell Application of solid-phase extraction materials from ion liquids for the analysis of carbamate pesticide
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1