使用便携式平面 HPGe 探测器估算嵌入组织中的 241Am 和 239Pu 放射性活度。

Pub Date : 2024-06-03 DOI:10.1093/rpd/ncae105
Manohari Murugan, Sugumar Ponraj, Mathiyarasu Ramasamy, Ponraju Durairaj, Venkatraman Balasubramanian
{"title":"使用便携式平面 HPGe 探测器估算嵌入组织中的 241Am 和 239Pu 放射性活度。","authors":"Manohari Murugan, Sugumar Ponraj, Mathiyarasu Ramasamy, Ponraju Durairaj, Venkatraman Balasubramanian","doi":"10.1093/rpd/ncae105","DOIUrl":null,"url":null,"abstract":"<p><p>This paper describes a procedure for the estimation of 241Am and 239Pu activity present in the human tissue by measuring the depth of contaminant using a portable Planar High Purity Germanium detector (HPGe). The ratios of photopeak counts of X-rays or gammas obtained with the detector coupled to collimator are calculated for the estimation of depth of the contaminant and the optimum one is determined. Since Minimum Detectable Activities (MDA) for the detector coupled to a collimator are higher than that of bare detector, activity must be estimated using bare detector, after locating the contaminant. Two methods are described for the estimation of plutonium coexisting with 241Am: (i) Abundance and isotopic correction for 239Pu (ii) and 239Pu:241Am ratio. The procedure to estimate 239Pu when plutonium isotopes alone are present is also established. An optimum monitoring period to detect the minimum value of intake for both radionuclides corresponding to chelation therapy and excision is also derived.</p>","PeriodicalId":0,"journal":{"name":"","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2024-06-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Estimation of 241Am and 239Pu activity embedded in the tissue using portable planar HPGe detector.\",\"authors\":\"Manohari Murugan, Sugumar Ponraj, Mathiyarasu Ramasamy, Ponraju Durairaj, Venkatraman Balasubramanian\",\"doi\":\"10.1093/rpd/ncae105\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>This paper describes a procedure for the estimation of 241Am and 239Pu activity present in the human tissue by measuring the depth of contaminant using a portable Planar High Purity Germanium detector (HPGe). The ratios of photopeak counts of X-rays or gammas obtained with the detector coupled to collimator are calculated for the estimation of depth of the contaminant and the optimum one is determined. Since Minimum Detectable Activities (MDA) for the detector coupled to a collimator are higher than that of bare detector, activity must be estimated using bare detector, after locating the contaminant. Two methods are described for the estimation of plutonium coexisting with 241Am: (i) Abundance and isotopic correction for 239Pu (ii) and 239Pu:241Am ratio. The procedure to estimate 239Pu when plutonium isotopes alone are present is also established. An optimum monitoring period to detect the minimum value of intake for both radionuclides corresponding to chelation therapy and excision is also derived.</p>\",\"PeriodicalId\":0,\"journal\":{\"name\":\"\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0,\"publicationDate\":\"2024-06-03\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"\",\"FirstCategoryId\":\"93\",\"ListUrlMain\":\"https://doi.org/10.1093/rpd/ncae105\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"","FirstCategoryId":"93","ListUrlMain":"https://doi.org/10.1093/rpd/ncae105","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本文介绍了一种利用便携式平面高纯锗探测器(HPGe)测量污染物深度,从而估算人体组织中 241Am 和 239Pu 放射性活度的方法。通过计算探测器与准直器耦合获得的 X 射线或伽马射线的光峰计数比率,来估算污染物的深度,并确定最佳比率。由于与准直器耦合的探测器的最小可探测放射性活度(MDA)高于裸探测器,因此在确定污染物位置后,必须使用裸探测器估算放射性活度。本报告介绍了估算与 241Am 共存的钚的两种方法:(i) 239Pu 的丰度和同位素校正 (ii) 以及 239Pu:241Am 比率。此外,还制定了在只有钚同位素存在时估算 239Pu 的程序。此外,还得出了一个最佳监测期,以检测与螯合疗法和切除术相对应的两种放射性核素的最小摄入值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
Estimation of 241Am and 239Pu activity embedded in the tissue using portable planar HPGe detector.

This paper describes a procedure for the estimation of 241Am and 239Pu activity present in the human tissue by measuring the depth of contaminant using a portable Planar High Purity Germanium detector (HPGe). The ratios of photopeak counts of X-rays or gammas obtained with the detector coupled to collimator are calculated for the estimation of depth of the contaminant and the optimum one is determined. Since Minimum Detectable Activities (MDA) for the detector coupled to a collimator are higher than that of bare detector, activity must be estimated using bare detector, after locating the contaminant. Two methods are described for the estimation of plutonium coexisting with 241Am: (i) Abundance and isotopic correction for 239Pu (ii) and 239Pu:241Am ratio. The procedure to estimate 239Pu when plutonium isotopes alone are present is also established. An optimum monitoring period to detect the minimum value of intake for both radionuclides corresponding to chelation therapy and excision is also derived.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1