{"title":"在非扁平光束下对半柔性三维腔体进行剂量学评估,并对不同探测器进行比较。","authors":"Kanakavel Kandasamy, E James Jebaseelan Samuel","doi":"10.4103/jmp.jmp_115_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The goal of this study is to investigate the dosimetric properties of a Semiflex three-dimensional (3D) chamber in an unflatten beam and compare its data from a small to a large field flattening filter-free (FFF) beam with different radiation detectors.</p><p><strong>Methods: </strong>The sensitivity, linearity, reproducibility, dose rate dependency, and energy dependence of a Semiflex 3D detector in flattening filter and filter-free beam were fully investigated. The minimum radiation observed field widths for all detectors were calculated using lateral electronic charged particle equilibrium to investigate dosimetric characteristics such as percentage depth doses (PDDs), profiles, and output factors (OPFs) for Semiflex 3D detector under 6FFF Beam. The Semiflex 3D measured data were compared to that of other detectors employed in this study.</p><p><strong>Results: </strong>The ion chamber has a dosage linearity deviation of +1.2% for <10 MU, a dose-rate dependency deviation of +0.5%, and significantly poorer sensitivity due to its small volume. There is a difference in field sizes between manufacturer specs and derived field sizes. The measured PDD, profiles, and OPFs of the Semiflex 3D chamber were within 1% of each other for all square field sizes set under linac for the 6FFF beam.</p><p><strong>Conclusion: </strong>It was discovered to be an appropriate detector for relative dose measurements for 6 FFF beams with higher dose rates for field sizes more than or equal to 3 cm × 3 cm.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 1","pages":"84-94"},"PeriodicalIF":0.7000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11141746/pdf/","citationCount":"0","resultStr":"{\"title\":\"Dosimetric Evaluation of Semiflex Three-dimensional Chamber under Unflatten Beam in Comparison among Different Detectors.\",\"authors\":\"Kanakavel Kandasamy, E James Jebaseelan Samuel\",\"doi\":\"10.4103/jmp.jmp_115_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>The goal of this study is to investigate the dosimetric properties of a Semiflex three-dimensional (3D) chamber in an unflatten beam and compare its data from a small to a large field flattening filter-free (FFF) beam with different radiation detectors.</p><p><strong>Methods: </strong>The sensitivity, linearity, reproducibility, dose rate dependency, and energy dependence of a Semiflex 3D detector in flattening filter and filter-free beam were fully investigated. The minimum radiation observed field widths for all detectors were calculated using lateral electronic charged particle equilibrium to investigate dosimetric characteristics such as percentage depth doses (PDDs), profiles, and output factors (OPFs) for Semiflex 3D detector under 6FFF Beam. The Semiflex 3D measured data were compared to that of other detectors employed in this study.</p><p><strong>Results: </strong>The ion chamber has a dosage linearity deviation of +1.2% for <10 MU, a dose-rate dependency deviation of +0.5%, and significantly poorer sensitivity due to its small volume. There is a difference in field sizes between manufacturer specs and derived field sizes. The measured PDD, profiles, and OPFs of the Semiflex 3D chamber were within 1% of each other for all square field sizes set under linac for the 6FFF beam.</p><p><strong>Conclusion: </strong>It was discovered to be an appropriate detector for relative dose measurements for 6 FFF beams with higher dose rates for field sizes more than or equal to 3 cm × 3 cm.</p>\",\"PeriodicalId\":51719,\"journal\":{\"name\":\"Journal of Medical Physics\",\"volume\":\"49 1\",\"pages\":\"84-94\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11141746/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jmp.jmp_115_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/3/30 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_115_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/30 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
摘要
目的:本研究的目的是调查 Semiflex 三维(3D)腔室在非扁平化射束中的剂量测定特性,并比较其在小场扁平化无滤光片(FFF)射束和大场扁平化无滤光片(FFF)射束中使用不同辐射探测器的数据:全面研究了Semiflex三维探测器在扁平化滤光片和无滤光片光束中的灵敏度、线性度、再现性、剂量率依赖性和能量依赖性。利用横向电子带电粒子平衡计算出所有探测器的最小辐射观测场宽,以研究 6FFF 光束下 Semiflex 3D 探测器的剂量学特性,如深度剂量百分比(PDD)、剖面和输出因子(OPF)。Semiflex 3D 的测量数据与本研究中使用的其他探测器的数据进行了比较:结果:离子室的剂量线性偏差为+1.2%:研究发现,它是一种合适的检测器,可用于测量 6 FFF 射束的相对剂量,在射野尺寸大于或等于 3 厘米×3 厘米时,剂量率更高。
Dosimetric Evaluation of Semiflex Three-dimensional Chamber under Unflatten Beam in Comparison among Different Detectors.
Purpose: The goal of this study is to investigate the dosimetric properties of a Semiflex three-dimensional (3D) chamber in an unflatten beam and compare its data from a small to a large field flattening filter-free (FFF) beam with different radiation detectors.
Methods: The sensitivity, linearity, reproducibility, dose rate dependency, and energy dependence of a Semiflex 3D detector in flattening filter and filter-free beam were fully investigated. The minimum radiation observed field widths for all detectors were calculated using lateral electronic charged particle equilibrium to investigate dosimetric characteristics such as percentage depth doses (PDDs), profiles, and output factors (OPFs) for Semiflex 3D detector under 6FFF Beam. The Semiflex 3D measured data were compared to that of other detectors employed in this study.
Results: The ion chamber has a dosage linearity deviation of +1.2% for <10 MU, a dose-rate dependency deviation of +0.5%, and significantly poorer sensitivity due to its small volume. There is a difference in field sizes between manufacturer specs and derived field sizes. The measured PDD, profiles, and OPFs of the Semiflex 3D chamber were within 1% of each other for all square field sizes set under linac for the 6FFF beam.
Conclusion: It was discovered to be an appropriate detector for relative dose measurements for 6 FFF beams with higher dose rates for field sizes more than or equal to 3 cm × 3 cm.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.