{"title":"在气管导管外部连接吸管作为冷冻活组织切片检查中 Fogarty 导管插入路径的新方法:回顾性比较研究","authors":"Hiroshi Ishimoto , Noriho Sakamoto , Daisuke Okuno , Hirokazu Yura , Mutsumi Ozasa , Ritsuko Miyashita , Takatomo Tokito , Takashi Kido , Shinnosuke Takemoto , Takahiro Takazono , Yasushi Obase , Yuji Ishimatsu , Tomoya Nishino , Hiroshi Mukae","doi":"10.1016/j.resinv.2024.06.001","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Cryobiopsy use is anticipated to become more common in diagnosing lung diseases. In Japan, inserting a Fogarty catheter through a suction channel above the endotracheal tube's cuff for hemostasis is common practice. However, the rigid nature of the endotracheal tube poses challenges to tracheal intubation using a bronchoscope. The endotracheal tube cuff must be removed to prevent interference during Fogarty catheter insertion. To simplify the procedure and enhance safety, we devised and implemented a method of inserting a hemostatic Fogarty catheter with a suction tube externally attached to a softer endotracheal tube. This study aimed to evaluate the sustainability of this Fogarty catheter insertion method using suction tubes.</p></div><div><h3>Methods</h3><p>The hemostatic Fogarty catheter insertion method was retrospectively validated. We compared outcomes between 60 patients who underwent the conventional method with a suction channel above the cuff and 50 patients who underwent the novel approach with an externally attached suction tube.</p></div><div><h3>Results</h3><p>The physicians performing bronchoscopy and inserting the Fogarty catheter in the group in which the suction tube was externally attached for Fogarty catheter insertion had little experience. However, the overall bronchoscopy time was shorter; the two groups showed no significant differences in complications.</p></div><div><h3>Conclusion</h3><p>Regarding cryobiopsy procedures, using an externally attached suction tube for Fogarty catheter insertion was practical and comparable to the conventional method of using a suction channel above the cuff. This method made the procedure more simple and safe.</p></div>","PeriodicalId":20934,"journal":{"name":"Respiratory investigation","volume":"62 4","pages":"Pages 732-737"},"PeriodicalIF":2.4000,"publicationDate":"2024-06-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2212534524000935/pdfft?md5=3d319577c45ee5a1919abd1e45efa291&pid=1-s2.0-S2212534524000935-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Novel method of attaching a suction tube externally to a tracheal tube as an insertion route for a Fogarty catheter in cryobiopsy: Retrospective comparative study\",\"authors\":\"Hiroshi Ishimoto , Noriho Sakamoto , Daisuke Okuno , Hirokazu Yura , Mutsumi Ozasa , Ritsuko Miyashita , Takatomo Tokito , Takashi Kido , Shinnosuke Takemoto , Takahiro Takazono , Yasushi Obase , Yuji Ishimatsu , Tomoya Nishino , Hiroshi Mukae\",\"doi\":\"10.1016/j.resinv.2024.06.001\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>Cryobiopsy use is anticipated to become more common in diagnosing lung diseases. In Japan, inserting a Fogarty catheter through a suction channel above the endotracheal tube's cuff for hemostasis is common practice. However, the rigid nature of the endotracheal tube poses challenges to tracheal intubation using a bronchoscope. The endotracheal tube cuff must be removed to prevent interference during Fogarty catheter insertion. To simplify the procedure and enhance safety, we devised and implemented a method of inserting a hemostatic Fogarty catheter with a suction tube externally attached to a softer endotracheal tube. This study aimed to evaluate the sustainability of this Fogarty catheter insertion method using suction tubes.</p></div><div><h3>Methods</h3><p>The hemostatic Fogarty catheter insertion method was retrospectively validated. We compared outcomes between 60 patients who underwent the conventional method with a suction channel above the cuff and 50 patients who underwent the novel approach with an externally attached suction tube.</p></div><div><h3>Results</h3><p>The physicians performing bronchoscopy and inserting the Fogarty catheter in the group in which the suction tube was externally attached for Fogarty catheter insertion had little experience. However, the overall bronchoscopy time was shorter; the two groups showed no significant differences in complications.</p></div><div><h3>Conclusion</h3><p>Regarding cryobiopsy procedures, using an externally attached suction tube for Fogarty catheter insertion was practical and comparable to the conventional method of using a suction channel above the cuff. This method made the procedure more simple and safe.</p></div>\",\"PeriodicalId\":20934,\"journal\":{\"name\":\"Respiratory investigation\",\"volume\":\"62 4\",\"pages\":\"Pages 732-737\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-06-14\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2212534524000935/pdfft?md5=3d319577c45ee5a1919abd1e45efa291&pid=1-s2.0-S2212534524000935-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Respiratory investigation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2212534524000935\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RESPIRATORY SYSTEM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Respiratory investigation","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2212534524000935","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RESPIRATORY SYSTEM","Score":null,"Total":0}
Novel method of attaching a suction tube externally to a tracheal tube as an insertion route for a Fogarty catheter in cryobiopsy: Retrospective comparative study
Background
Cryobiopsy use is anticipated to become more common in diagnosing lung diseases. In Japan, inserting a Fogarty catheter through a suction channel above the endotracheal tube's cuff for hemostasis is common practice. However, the rigid nature of the endotracheal tube poses challenges to tracheal intubation using a bronchoscope. The endotracheal tube cuff must be removed to prevent interference during Fogarty catheter insertion. To simplify the procedure and enhance safety, we devised and implemented a method of inserting a hemostatic Fogarty catheter with a suction tube externally attached to a softer endotracheal tube. This study aimed to evaluate the sustainability of this Fogarty catheter insertion method using suction tubes.
Methods
The hemostatic Fogarty catheter insertion method was retrospectively validated. We compared outcomes between 60 patients who underwent the conventional method with a suction channel above the cuff and 50 patients who underwent the novel approach with an externally attached suction tube.
Results
The physicians performing bronchoscopy and inserting the Fogarty catheter in the group in which the suction tube was externally attached for Fogarty catheter insertion had little experience. However, the overall bronchoscopy time was shorter; the two groups showed no significant differences in complications.
Conclusion
Regarding cryobiopsy procedures, using an externally attached suction tube for Fogarty catheter insertion was practical and comparable to the conventional method of using a suction channel above the cuff. This method made the procedure more simple and safe.