Adrien Thiriet, Vincent Poindron, Laurent Sattler, Jordan Wimmer, Delphine Rolland, Anne-Sophie Korganow, Laurent Mauvieux, Agathe Herb
{"title":"评估用于检测狼疮抗凝物的综合活化部分凝血活酶时间(Cephen LS/Cephen)。","authors":"Adrien Thiriet, Vincent Poindron, Laurent Sattler, Jordan Wimmer, Delphine Rolland, Anne-Sophie Korganow, Laurent Mauvieux, Agathe Herb","doi":"10.1111/ijlh.14349","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>It is recommended to use two chronometric assays of different principles for the diagnosis of lupus anticoagulant (LA), consisting in diluted Russell Viper Venom Time (dRVVT) and activated Partial Thromboplastin Time (aPTT). Yet, there are only a few integrated aPTT assays; this study aims to evaluate one of them: Cephen LS/Cephen (Hyphen Biomed).</p><p><strong>Method: </strong>249 samples of patients were included in this study. Normal reference ranges were determined with platelet-poor plasma (PPP) from healthy blood donors. Performances were then evaluated by comparing Cephen LS/Cephen test results to the results of the laboratory's reference assay for the diagnosis of LA and to clinical data, both on non-anticoagulated and anticoagulated patients' samples (Unfractioned heparin (UFH), Low Molecular Weight Heparin (LMWH), Vitamin K Antagonists (VKA) and apixaban). Interference of UFH, LMWH and VKA were also evaluated thanks to spiking experiment of increasing heparin concentrations or factor deficiency.</p><p><strong>Results: </strong>Cephen LS/Cephen test had 48.6% sensitivity towards LA. Although UFH and VKA seemed to interfere with this assay and were likely to cause false negative, LMWH and apixaban did not. Finally, combination of Cephen LS/ Cephen with dRVVT had 89.0% sensitivity.</p><p><strong>Conclusion: </strong>Cephen LS/Cephen seems relevant for LA diagnosis, in combination with dRVVT, and might be used in patients undergoing LMWH or apixaban therapy.</p>","PeriodicalId":94050,"journal":{"name":"International journal of laboratory hematology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of an integrated activated partial thromboplastin time (Cephen LS/Cephen) for the detection of lupus anticoagulant.\",\"authors\":\"Adrien Thiriet, Vincent Poindron, Laurent Sattler, Jordan Wimmer, Delphine Rolland, Anne-Sophie Korganow, Laurent Mauvieux, Agathe Herb\",\"doi\":\"10.1111/ijlh.14349\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>It is recommended to use two chronometric assays of different principles for the diagnosis of lupus anticoagulant (LA), consisting in diluted Russell Viper Venom Time (dRVVT) and activated Partial Thromboplastin Time (aPTT). Yet, there are only a few integrated aPTT assays; this study aims to evaluate one of them: Cephen LS/Cephen (Hyphen Biomed).</p><p><strong>Method: </strong>249 samples of patients were included in this study. Normal reference ranges were determined with platelet-poor plasma (PPP) from healthy blood donors. Performances were then evaluated by comparing Cephen LS/Cephen test results to the results of the laboratory's reference assay for the diagnosis of LA and to clinical data, both on non-anticoagulated and anticoagulated patients' samples (Unfractioned heparin (UFH), Low Molecular Weight Heparin (LMWH), Vitamin K Antagonists (VKA) and apixaban). Interference of UFH, LMWH and VKA were also evaluated thanks to spiking experiment of increasing heparin concentrations or factor deficiency.</p><p><strong>Results: </strong>Cephen LS/Cephen test had 48.6% sensitivity towards LA. Although UFH and VKA seemed to interfere with this assay and were likely to cause false negative, LMWH and apixaban did not. Finally, combination of Cephen LS/ Cephen with dRVVT had 89.0% sensitivity.</p><p><strong>Conclusion: </strong>Cephen LS/Cephen seems relevant for LA diagnosis, in combination with dRVVT, and might be used in patients undergoing LMWH or apixaban therapy.</p>\",\"PeriodicalId\":94050,\"journal\":{\"name\":\"International journal of laboratory hematology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International journal of laboratory hematology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1111/ijlh.14349\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/8/5 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of laboratory hematology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1111/ijlh.14349","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/5 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
Evaluation of an integrated activated partial thromboplastin time (Cephen LS/Cephen) for the detection of lupus anticoagulant.
Introduction: It is recommended to use two chronometric assays of different principles for the diagnosis of lupus anticoagulant (LA), consisting in diluted Russell Viper Venom Time (dRVVT) and activated Partial Thromboplastin Time (aPTT). Yet, there are only a few integrated aPTT assays; this study aims to evaluate one of them: Cephen LS/Cephen (Hyphen Biomed).
Method: 249 samples of patients were included in this study. Normal reference ranges were determined with platelet-poor plasma (PPP) from healthy blood donors. Performances were then evaluated by comparing Cephen LS/Cephen test results to the results of the laboratory's reference assay for the diagnosis of LA and to clinical data, both on non-anticoagulated and anticoagulated patients' samples (Unfractioned heparin (UFH), Low Molecular Weight Heparin (LMWH), Vitamin K Antagonists (VKA) and apixaban). Interference of UFH, LMWH and VKA were also evaluated thanks to spiking experiment of increasing heparin concentrations or factor deficiency.
Results: Cephen LS/Cephen test had 48.6% sensitivity towards LA. Although UFH and VKA seemed to interfere with this assay and were likely to cause false negative, LMWH and apixaban did not. Finally, combination of Cephen LS/ Cephen with dRVVT had 89.0% sensitivity.
Conclusion: Cephen LS/Cephen seems relevant for LA diagnosis, in combination with dRVVT, and might be used in patients undergoing LMWH or apixaban therapy.