便携式血红蛋白仪(Verio Q Red)与常规临床实践中临床实验室结果的比较。

IF 1.8 4区 医学 Q2 MEDICINE, GENERAL & INTERNAL Atencion Primaria Pub Date : 2024-09-12 DOI:10.1016/j.aprim.2024.103080
Boris Trenado-Luengo , Rosa García-Sierra , Verónica Moreno Gómez , Marina Montenegro Calvo , Jordi Anguita Lapido , Pere Torán-Monserrat
{"title":"便携式血红蛋白仪(Verio Q Red)与常规临床实践中临床实验室结果的比较。","authors":"Boris Trenado-Luengo ,&nbsp;Rosa García-Sierra ,&nbsp;Verónica Moreno Gómez ,&nbsp;Marina Montenegro Calvo ,&nbsp;Jordi Anguita Lapido ,&nbsp;Pere Torán-Monserrat","doi":"10.1016/j.aprim.2024.103080","DOIUrl":null,"url":null,"abstract":"<div><h3>Objective</h3><p>To analyze the concordance between a hemoglobinometer with capillary blood sample and the clinical laboratory with a venous blood sample.</p></div><div><h3>Design</h3><p>Cross-sectional concordance study.</p></div><div><h3>Location</h3><p>Care Centre Primary Bufalà-Canyet Primary and Care Centre Primary Dalt la Vila Primary Care Center, Badalona, Barcelona.</p></div><div><h3>Participants</h3><p>Random selection of participants who attended routine blood tests. Over 18<!--> <!-->years old. No abandonment or loss was obtained.</p></div><div><h3>Main interventions and measurements</h3><p>Sex, age, and reason for the blood test were collected from the medical history. Venous blood is drawn, and simultaneously, capillary blood is collected.</p></div><div><h3>Results</h3><p>120 individuals are included, with an average age of 58.9<!--> <!-->years. The Bland-Almant graph showed differences within the confidence intervals for hemoglobin and hematocrit. The mean differences between the values of the Verio Q Red hemoglobinometer and those of the clinical laboratory were −0.42<!--> <!-->mg/dl for hemoglobin and −1.25% for hematocrit. The intraclass correlation coefficient showed excellent correlation for hemoglobin and hematocrit between the Verio Q Red hemoglobinometer and the clinical laboratory. Pearson's correlation for hemoglobin was 0.737 and for hematocrit 0.787.</p></div><div><h3>Conclusions</h3><p>The Verio Q hemoglobinometer is a valid tool for the early diagnosis of anemias and can be very useful in primary care consultations.</p></div>","PeriodicalId":55435,"journal":{"name":"Atencion Primaria","volume":"57 2","pages":"Article 103080"},"PeriodicalIF":1.8000,"publicationDate":"2024-09-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0212656724002221/pdfft?md5=2145bcd59571a4d2a58341635066ed51&pid=1-s2.0-S0212656724002221-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Comparación de un hemoglobinómetro portátil (Verio Q Red) con los resultados del laboratorio clínico en la práctica clínica habitual\",\"authors\":\"Boris Trenado-Luengo ,&nbsp;Rosa García-Sierra ,&nbsp;Verónica Moreno Gómez ,&nbsp;Marina Montenegro Calvo ,&nbsp;Jordi Anguita Lapido ,&nbsp;Pere Torán-Monserrat\",\"doi\":\"10.1016/j.aprim.2024.103080\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objective</h3><p>To analyze the concordance between a hemoglobinometer with capillary blood sample and the clinical laboratory with a venous blood sample.</p></div><div><h3>Design</h3><p>Cross-sectional concordance study.</p></div><div><h3>Location</h3><p>Care Centre Primary Bufalà-Canyet Primary and Care Centre Primary Dalt la Vila Primary Care Center, Badalona, Barcelona.</p></div><div><h3>Participants</h3><p>Random selection of participants who attended routine blood tests. Over 18<!--> <!-->years old. No abandonment or loss was obtained.</p></div><div><h3>Main interventions and measurements</h3><p>Sex, age, and reason for the blood test were collected from the medical history. Venous blood is drawn, and simultaneously, capillary blood is collected.</p></div><div><h3>Results</h3><p>120 individuals are included, with an average age of 58.9<!--> <!-->years. The Bland-Almant graph showed differences within the confidence intervals for hemoglobin and hematocrit. The mean differences between the values of the Verio Q Red hemoglobinometer and those of the clinical laboratory were −0.42<!--> <!-->mg/dl for hemoglobin and −1.25% for hematocrit. The intraclass correlation coefficient showed excellent correlation for hemoglobin and hematocrit between the Verio Q Red hemoglobinometer and the clinical laboratory. Pearson's correlation for hemoglobin was 0.737 and for hematocrit 0.787.</p></div><div><h3>Conclusions</h3><p>The Verio Q hemoglobinometer is a valid tool for the early diagnosis of anemias and can be very useful in primary care consultations.</p></div>\",\"PeriodicalId\":55435,\"journal\":{\"name\":\"Atencion Primaria\",\"volume\":\"57 2\",\"pages\":\"Article 103080\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2024-09-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S0212656724002221/pdfft?md5=2145bcd59571a4d2a58341635066ed51&pid=1-s2.0-S0212656724002221-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Atencion Primaria\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0212656724002221\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICINE, GENERAL & INTERNAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Atencion Primaria","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0212656724002221","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0

摘要

地点巴塞罗那巴达洛纳市 Bufalà-Canyet 初级保健中心和 Dalt la Vila 初级保健中心。参与者随机选择参加常规血液检查的参与者。年龄在 18 岁以上。从病史中收集性别、年龄和验血原因。抽取静脉血,同时采集毛细血管血。结果纳入 120 人,平均年龄 58.9 岁。布兰德-阿尔曼特图显示血红蛋白和血细胞比容在置信区间内存在差异。Verio Q Red 血红蛋白仪和临床实验室的血红蛋白和血细胞比容平均值分别为-0.42 mg/dl和-1.25%。类内相关系数显示,Verio Q Red 血红蛋白仪和临床实验室之间在血红蛋白和血细胞比容方面的相关性极佳。结论 Verio Q 血红蛋白仪是早期诊断贫血症的有效工具,在初级保健咨询中非常有用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Comparación de un hemoglobinómetro portátil (Verio Q Red) con los resultados del laboratorio clínico en la práctica clínica habitual

Objective

To analyze the concordance between a hemoglobinometer with capillary blood sample and the clinical laboratory with a venous blood sample.

Design

Cross-sectional concordance study.

Location

Care Centre Primary Bufalà-Canyet Primary and Care Centre Primary Dalt la Vila Primary Care Center, Badalona, Barcelona.

Participants

Random selection of participants who attended routine blood tests. Over 18 years old. No abandonment or loss was obtained.

Main interventions and measurements

Sex, age, and reason for the blood test were collected from the medical history. Venous blood is drawn, and simultaneously, capillary blood is collected.

Results

120 individuals are included, with an average age of 58.9 years. The Bland-Almant graph showed differences within the confidence intervals for hemoglobin and hematocrit. The mean differences between the values of the Verio Q Red hemoglobinometer and those of the clinical laboratory were −0.42 mg/dl for hemoglobin and −1.25% for hematocrit. The intraclass correlation coefficient showed excellent correlation for hemoglobin and hematocrit between the Verio Q Red hemoglobinometer and the clinical laboratory. Pearson's correlation for hemoglobin was 0.737 and for hematocrit 0.787.

Conclusions

The Verio Q hemoglobinometer is a valid tool for the early diagnosis of anemias and can be very useful in primary care consultations.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Atencion Primaria
Atencion Primaria 医学-医学:内科
CiteScore
2.90
自引率
8.00%
发文量
156
审稿时长
33 days
期刊介绍: Atención Primaria es una revista que publica trabajos de investigación relativos al ámbito de la atención primaria de salud. Desde el punto de vista conceptual, Atención Primaria asume el nuevo modelo de atención primaria de salud, orientado no sólo a la curación de la enfermedad, sino también a su prevención y a la promoción de la salud, tanto en el plano individual como en el de la familia y la comunidad. En estos nuevos aspectos que definen el modelo de atención primaria de salud es en los que se centran los trabajos de investigación que publica Atención Primaria, la primera revista de originales española creada para recoger y difundir la producción científica realizada desde los centros de atención primaria de salud sobre cuestiones como protocolización de la asistencia, programas de prevención, seguimiento y control de pacientes crónicos, organización y gestión de la asistencia primaria, entre otros.
期刊最新文献
[Suicide prevention in the workplace]. [Pharmacological treatment of type 2 diabetes mellitus]. [Analysis of skills and perceptions regarding mHealth in the management of chronic patients by primary care professionals]. Calidad de la atención en podología universitaria. Análisis del cuestionario de satisfacción del paciente Tratamiento analgésico y adecuación terapéutica en los pacientes con enfermedad crónica avanzada. Oportunidades de mejora
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1