Sysmex XN-V 血液分析仪对猫科动物标本的验证

IF 1.2 4区 农林科学 Q3 VETERINARY SCIENCES Veterinary clinical pathology Pub Date : 2024-09-18 DOI:10.1111/vcp.13377
Maud Guerlin, Fanny Granat, Margot Grebert, Jean-Pierre Braun, Anne Geffré, Nathalie Bourgès-Abella, Catherine Trumel
{"title":"Sysmex XN-V 血液分析仪对猫科动物标本的验证","authors":"Maud Guerlin,&nbsp;Fanny Granat,&nbsp;Margot Grebert,&nbsp;Jean-Pierre Braun,&nbsp;Anne Geffré,&nbsp;Nathalie Bourgès-Abella,&nbsp;Catherine Trumel","doi":"10.1111/vcp.13377","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>The Sysmex XN-V is derived from the new Sysmex XN series of human hematology analyzers. The main changes from the previously validated XT-2000iV analyzer include an optic-fluorescent analysis for platelets and a nucleated red blood cell (NRBC) count.</p>\n </section>\n \n <section>\n \n <h3> Objective</h3>\n \n <p>We aimed to validate the Sysmex XN-V for feline blood following the American College for Veterinary Clinical Pathology and International Council for Standardization in Hematology recommendations.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Feline EDTA blood specimens were analyzed on the Sysmex XN-V to evaluate repeatability, linearity, comparison with the XT-2000iV analyzer and manual methods, stability, and to verify the previously established Sysmex XT-2000iV RIs.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Repeatability was excellent for most variables. Visually determined linearity was excellent or good for most variables except eosinophils and platelet variables. The correlation between the XN-V and XT-2000iV analyzers was good (≥0.82) for all variables except reticulocyte indices. Correlations between the Sysmex XN-V and manual differential counts were good to excellent for most variables, acceptable for neutrophils, and fair for monocytes and NRBC. The previously established Sysmex XT-2000iV RIs can be used to interpret results from the Sysmex XN-V analyzer for most variables except red cell distribution width and reticulocyte variables. The RI for platelet variables could not be evaluated because of platelet clumps. Changes in the Sysmex XN-V measurements after storage at 4 and 24°C were similar to those described for the Sysmex XT-2000iV analyzer.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>The performance of the Sysmex XN-V analyzer was good and compared favorably with the Sysmex XT-2000iV analyzer.</p>\n </section>\n </div>","PeriodicalId":23593,"journal":{"name":"Veterinary clinical pathology","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2024-09-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1111/vcp.13377","citationCount":"0","resultStr":"{\"title\":\"Validation of the Sysmex XN-V hematology analyzer for feline specimens\",\"authors\":\"Maud Guerlin,&nbsp;Fanny Granat,&nbsp;Margot Grebert,&nbsp;Jean-Pierre Braun,&nbsp;Anne Geffré,&nbsp;Nathalie Bourgès-Abella,&nbsp;Catherine Trumel\",\"doi\":\"10.1111/vcp.13377\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>The Sysmex XN-V is derived from the new Sysmex XN series of human hematology analyzers. The main changes from the previously validated XT-2000iV analyzer include an optic-fluorescent analysis for platelets and a nucleated red blood cell (NRBC) count.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Objective</h3>\\n \\n <p>We aimed to validate the Sysmex XN-V for feline blood following the American College for Veterinary Clinical Pathology and International Council for Standardization in Hematology recommendations.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>Feline EDTA blood specimens were analyzed on the Sysmex XN-V to evaluate repeatability, linearity, comparison with the XT-2000iV analyzer and manual methods, stability, and to verify the previously established Sysmex XT-2000iV RIs.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Repeatability was excellent for most variables. Visually determined linearity was excellent or good for most variables except eosinophils and platelet variables. The correlation between the XN-V and XT-2000iV analyzers was good (≥0.82) for all variables except reticulocyte indices. Correlations between the Sysmex XN-V and manual differential counts were good to excellent for most variables, acceptable for neutrophils, and fair for monocytes and NRBC. The previously established Sysmex XT-2000iV RIs can be used to interpret results from the Sysmex XN-V analyzer for most variables except red cell distribution width and reticulocyte variables. The RI for platelet variables could not be evaluated because of platelet clumps. Changes in the Sysmex XN-V measurements after storage at 4 and 24°C were similar to those described for the Sysmex XT-2000iV analyzer.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>The performance of the Sysmex XN-V analyzer was good and compared favorably with the Sysmex XT-2000iV analyzer.</p>\\n </section>\\n </div>\",\"PeriodicalId\":23593,\"journal\":{\"name\":\"Veterinary clinical pathology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-09-18\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1111/vcp.13377\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Veterinary clinical pathology\",\"FirstCategoryId\":\"97\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/vcp.13377\",\"RegionNum\":4,\"RegionCategory\":\"农林科学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"VETERINARY SCIENCES\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary clinical pathology","FirstCategoryId":"97","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/vcp.13377","RegionNum":4,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

摘要

背景Sysmex XN-V 是新型 Sysmex XN 系列人体血液分析仪的衍生产品。与之前经过验证的 XT-2000iV 分析仪相比,它的主要变化包括对血小板和有核红细胞(NRBC)计数的光学荧光分析。目的我们旨在根据美国兽医临床病理学院和国际血液学标准化委员会的建议,对 Sysmex XN-V 猫科动物血液分析仪进行验证。方法用 Sysmex XN-V 分析猫 EDTA 血液标本,以评估可重复性、线性、与 XT-2000iV 分析仪和手动方法的比较、稳定性,并验证之前建立的 Sysmex XT-2000iV RI。除嗜酸性粒细胞和血小板变量外,大多数变量的目测线性度为优或良。除网织红细胞指数外,XN-V 和 XT-2000iV 分析仪对所有变量的相关性都很好(≥0.82)。在大多数变量上,Sysmex XN-V 与人工差分计数的相关性良好至极佳,中性粒细胞的相关性尚可,单核细胞和非红细胞的相关性尚可。先前建立的 Sysmex XT-2000iV RI 可用来解释 Sysmex XN-V 分析仪的大多数变量结果,但红细胞分布宽度和网织红细胞变量除外。由于存在血小板团块,因此无法评估血小板变量的 RI。结论 Sysmex XN-V 分析仪的性能良好,与 Sysmex XT-2000iV 分析仪相比毫不逊色。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Validation of the Sysmex XN-V hematology analyzer for feline specimens

Background

The Sysmex XN-V is derived from the new Sysmex XN series of human hematology analyzers. The main changes from the previously validated XT-2000iV analyzer include an optic-fluorescent analysis for platelets and a nucleated red blood cell (NRBC) count.

Objective

We aimed to validate the Sysmex XN-V for feline blood following the American College for Veterinary Clinical Pathology and International Council for Standardization in Hematology recommendations.

Methods

Feline EDTA blood specimens were analyzed on the Sysmex XN-V to evaluate repeatability, linearity, comparison with the XT-2000iV analyzer and manual methods, stability, and to verify the previously established Sysmex XT-2000iV RIs.

Results

Repeatability was excellent for most variables. Visually determined linearity was excellent or good for most variables except eosinophils and platelet variables. The correlation between the XN-V and XT-2000iV analyzers was good (≥0.82) for all variables except reticulocyte indices. Correlations between the Sysmex XN-V and manual differential counts were good to excellent for most variables, acceptable for neutrophils, and fair for monocytes and NRBC. The previously established Sysmex XT-2000iV RIs can be used to interpret results from the Sysmex XN-V analyzer for most variables except red cell distribution width and reticulocyte variables. The RI for platelet variables could not be evaluated because of platelet clumps. Changes in the Sysmex XN-V measurements after storage at 4 and 24°C were similar to those described for the Sysmex XT-2000iV analyzer.

Conclusions

The performance of the Sysmex XN-V analyzer was good and compared favorably with the Sysmex XT-2000iV analyzer.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Veterinary clinical pathology
Veterinary clinical pathology 农林科学-兽医学
CiteScore
1.70
自引率
16.70%
发文量
133
审稿时长
18-36 weeks
期刊介绍: Veterinary Clinical Pathology is the official journal of the American Society for Veterinary Clinical Pathology (ASVCP) and the European Society of Veterinary Clinical Pathology (ESVCP). The journal''s mission is to provide an international forum for communication and discussion of scientific investigations and new developments that advance the art and science of laboratory diagnosis in animals. Veterinary Clinical Pathology welcomes original experimental research and clinical contributions involving domestic, laboratory, avian, and wildlife species in the areas of hematology, hemostasis, immunopathology, clinical chemistry, cytopathology, surgical pathology, toxicology, endocrinology, laboratory and analytical techniques, instrumentation, quality assurance, and clinical pathology education.
期刊最新文献
What is your diagnosis: Acute hemolysis in a Limousin bull. New strategies to classify canine pleural effusions and the diagnostic value of acute phase proteins, amylase, and adenosine deaminase in pleural exudates. What is your diagnosis? Intra-abdominal mass in a dog. What's your diagnosis? Circumferential small intestinal mass in a cat. Detection of renal tubular transporter proteins in canine urinary extracellular vesicles using liquid chromatography-tandem mass spectrometry.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1