评估非压电便携式超声设备在产后护理中的准确性和可靠性。

IF 2.1 3区 医学 Q2 OBSTETRICS & GYNECOLOGY Archives of Gynecology and Obstetrics Pub Date : 2024-10-04 DOI:10.1007/s00404-024-07744-3
Ruben Plöger, Charlotte Behning, Adeline Walter, Agnes Wittek, Ulrich Gembruch, Brigitte Strizek, Florian Recker
{"title":"评估非压电便携式超声设备在产后护理中的准确性和可靠性。","authors":"Ruben Plöger, Charlotte Behning, Adeline Walter, Agnes Wittek, Ulrich Gembruch, Brigitte Strizek, Florian Recker","doi":"10.1007/s00404-024-07744-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The early diagnosis of hemorrhage via postpartum ultrasound is crucial to initiate therapy and, thus, prevent maternal death. In these critical situations rapid availability and simple transport of ultrasound devices is vital, paving the way for a  new generation of portable handheld ultrasound devices (PUD) consisting of transducers and tablets or smart phones. However, evidence to confirm the diagnostic accuracy of these new devices is still scarce.</p><p><strong>Methods: </strong>The accuracy and reliability of these new devices in relation to established standard ultrasound devices is analyses in this pilot study by comparing diagnoses and by applying statistical analysis via Bland-Altman plots, intraclass correlation coefficients (ICC), and Pearson correlation coefficients (PCC). One hundred patients of a university hospital were included in this study.</p><p><strong>Results: </strong>In all cases, the same diagnosis was made regardless of the applied ultrasound device, confirming high accuracy. There was a high correlation (PCC 0.951) and excellent agreement (ICC 0.974) in the assessment of the cavum, while the assessment of the diameters of the uterus showed only a good correlation and a good agreement. Subgroup analysis for maternal weight, mode of delivery and day after delivery was performed  CONCLUSION: The same diagnosis independent of the used devices and excellent results of the cavum assessment promote the use of PUDs in a clinical setting. The slightly lower accuracy in the measurement of the uterus may be caused by the PUD's small acoustic window, reflecting one of its weaknesses. Therefore, the patient may benefit from the short time to diagnosis and the unbound location of examination, either in the delivery room, on the ward, or at home.</p>","PeriodicalId":8330,"journal":{"name":"Archives of Gynecology and Obstetrics","volume":" ","pages":""},"PeriodicalIF":2.1000,"publicationDate":"2024-10-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluating the accuracy and reliability of non-piezo portable ultrasound devices in postpartum care.\",\"authors\":\"Ruben Plöger, Charlotte Behning, Adeline Walter, Agnes Wittek, Ulrich Gembruch, Brigitte Strizek, Florian Recker\",\"doi\":\"10.1007/s00404-024-07744-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>The early diagnosis of hemorrhage via postpartum ultrasound is crucial to initiate therapy and, thus, prevent maternal death. In these critical situations rapid availability and simple transport of ultrasound devices is vital, paving the way for a  new generation of portable handheld ultrasound devices (PUD) consisting of transducers and tablets or smart phones. However, evidence to confirm the diagnostic accuracy of these new devices is still scarce.</p><p><strong>Methods: </strong>The accuracy and reliability of these new devices in relation to established standard ultrasound devices is analyses in this pilot study by comparing diagnoses and by applying statistical analysis via Bland-Altman plots, intraclass correlation coefficients (ICC), and Pearson correlation coefficients (PCC). One hundred patients of a university hospital were included in this study.</p><p><strong>Results: </strong>In all cases, the same diagnosis was made regardless of the applied ultrasound device, confirming high accuracy. There was a high correlation (PCC 0.951) and excellent agreement (ICC 0.974) in the assessment of the cavum, while the assessment of the diameters of the uterus showed only a good correlation and a good agreement. Subgroup analysis for maternal weight, mode of delivery and day after delivery was performed  CONCLUSION: The same diagnosis independent of the used devices and excellent results of the cavum assessment promote the use of PUDs in a clinical setting. The slightly lower accuracy in the measurement of the uterus may be caused by the PUD's small acoustic window, reflecting one of its weaknesses. Therefore, the patient may benefit from the short time to diagnosis and the unbound location of examination, either in the delivery room, on the ward, or at home.</p>\",\"PeriodicalId\":8330,\"journal\":{\"name\":\"Archives of Gynecology and Obstetrics\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2024-10-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Archives of Gynecology and Obstetrics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s00404-024-07744-3\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"OBSTETRICS & GYNECOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Archives of Gynecology and Obstetrics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00404-024-07744-3","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OBSTETRICS & GYNECOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

导言:通过产后超声波对大出血进行早期诊断对于启动治疗,从而防止产妇死亡至关重要。在这些危急情况下,超声波设备的快速可用性和简便运输至关重要,这为由传感器和平板电脑或智能手机组成的新一代便携式手持超声波设备(PUD)铺平了道路。然而,证实这些新设备诊断准确性的证据仍然很少:本试验研究通过比较诊断结果,并通过布兰德-阿尔特曼图、类内相关系数(ICC)和皮尔逊相关系数(PCC)进行统计分析,分析了这些新设备与现有标准超声设备的准确性和可靠性。一家大学医院的 100 名患者参与了这项研究:在所有病例中,无论使用哪种超声设备,诊断结果都是相同的,证明了诊断的高度准确性。宫腔评估的相关性高(PCC 0.951),一致性好(ICC 0.974),而子宫直径评估的相关性和一致性较好。对产妇体重、分娩方式和产后天数进行了分组分析 结论:与所使用的设备无关的相同诊断结果和出色的宫腔评估结果促进了 PUD 在临床中的应用。子宫测量的准确性稍低可能是由于 PUD 的声窗较小,这反映了其弱点之一。因此,诊断时间短、检查地点不受限制(可在产房、病房或家中进行)可能会使患者受益。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Evaluating the accuracy and reliability of non-piezo portable ultrasound devices in postpartum care.

Introduction: The early diagnosis of hemorrhage via postpartum ultrasound is crucial to initiate therapy and, thus, prevent maternal death. In these critical situations rapid availability and simple transport of ultrasound devices is vital, paving the way for a  new generation of portable handheld ultrasound devices (PUD) consisting of transducers and tablets or smart phones. However, evidence to confirm the diagnostic accuracy of these new devices is still scarce.

Methods: The accuracy and reliability of these new devices in relation to established standard ultrasound devices is analyses in this pilot study by comparing diagnoses and by applying statistical analysis via Bland-Altman plots, intraclass correlation coefficients (ICC), and Pearson correlation coefficients (PCC). One hundred patients of a university hospital were included in this study.

Results: In all cases, the same diagnosis was made regardless of the applied ultrasound device, confirming high accuracy. There was a high correlation (PCC 0.951) and excellent agreement (ICC 0.974) in the assessment of the cavum, while the assessment of the diameters of the uterus showed only a good correlation and a good agreement. Subgroup analysis for maternal weight, mode of delivery and day after delivery was performed  CONCLUSION: The same diagnosis independent of the used devices and excellent results of the cavum assessment promote the use of PUDs in a clinical setting. The slightly lower accuracy in the measurement of the uterus may be caused by the PUD's small acoustic window, reflecting one of its weaknesses. Therefore, the patient may benefit from the short time to diagnosis and the unbound location of examination, either in the delivery room, on the ward, or at home.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.70
自引率
15.40%
发文量
493
审稿时长
1 months
期刊介绍: Founded in 1870 as "Archiv für Gynaekologie", Archives of Gynecology and Obstetrics has a long and outstanding tradition. Since 1922 the journal has been the Organ of the Deutsche Gesellschaft für Gynäkologie und Geburtshilfe. "The Archives of Gynecology and Obstetrics" is circulated in over 40 countries world wide and is indexed in "PubMed/Medline" and "Science Citation Index Expanded/Journal Citation Report". The journal publishes invited and submitted reviews; peer-reviewed original articles about clinical topics and basic research as well as news and views and guidelines and position statements from all sub-specialties in gynecology and obstetrics.
期刊最新文献
The effects of the intrapartum care model given in line with the recommendations of the World Health Organization (WHO) on the mother's maternal behavior towards her baby, breastfeeding self-efficacy, breastfeeding success, and hospital discharge readiness: a randomized controlled trial. Association between placental site and successful induction of labor among postdate primiparous women. Clinical comparison of laparoscopic and open surgical approaches for uterus-preserving myomectomy: a retrospective analysis on patient-reported outcome, postoperative morbidity and pregnancy outcomes. Effect of mother's active pushing at cesarean delivery: a randomized controlled trial. Relationship of female pelvic floor muscle function and body composition: cross-sectional study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1