{"title":"基于重复控制的医用 X 射线应用 PPSS-LCC 转换器的输出电压纹波抑制功能","authors":"Chudi Lin;Liqun He;Mingdi Fan;Xiaohui Li;Shengfang Fan;Cheng Wang;Yaosuo Xue","doi":"10.1109/JESTIE.2024.3435005","DOIUrl":null,"url":null,"abstract":"The primary-parallel-secondary-series \n<italic>LCC</i>\n resonant high voltage (HV) power supply with single-phase ac input has been widely applied for medical X-ray application. It requires extremely low ripple (<0.5%)>LCC</i>\n converter at the same time. The dc-bus capacitance can, therefore, be remarkably reduced. This strategy has the advantages of no additional hardware costs, directly implemented in digital controller, and insensitive to accuracy of system parameters. Experiments are carried out on a practical 120 kV/900 W HV power supply and confirm the effectiveness of the proposed strategy. The peak–peak output voltage ripple is suppressed to only 0.36%, while the DC capacitance is significantly reduced to 1/12. Comparison of X-ray does prove that the image quality of X-ray machine is consequently greatly improved.","PeriodicalId":100620,"journal":{"name":"IEEE Journal of Emerging and Selected Topics in Industrial Electronics","volume":"5 4","pages":"1688-1697"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Output Voltage Ripple Suppression of PPSS-LCC Converter for Medical X-Ray Application Based on Repetitive Control\",\"authors\":\"Chudi Lin;Liqun He;Mingdi Fan;Xiaohui Li;Shengfang Fan;Cheng Wang;Yaosuo Xue\",\"doi\":\"10.1109/JESTIE.2024.3435005\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"The primary-parallel-secondary-series \\n<italic>LCC</i>\\n resonant high voltage (HV) power supply with single-phase ac input has been widely applied for medical X-ray application. It requires extremely low ripple (<0.5%)>LCC</i>\\n converter at the same time. The dc-bus capacitance can, therefore, be remarkably reduced. This strategy has the advantages of no additional hardware costs, directly implemented in digital controller, and insensitive to accuracy of system parameters. Experiments are carried out on a practical 120 kV/900 W HV power supply and confirm the effectiveness of the proposed strategy. The peak–peak output voltage ripple is suppressed to only 0.36%, while the DC capacitance is significantly reduced to 1/12. Comparison of X-ray does prove that the image quality of X-ray machine is consequently greatly improved.\",\"PeriodicalId\":100620,\"journal\":{\"name\":\"IEEE Journal of Emerging and Selected Topics in Industrial Electronics\",\"volume\":\"5 4\",\"pages\":\"1688-1697\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-07-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"IEEE Journal of Emerging and Selected Topics in Industrial Electronics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://ieeexplore.ieee.org/document/10614095/\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Journal of Emerging and Selected Topics in Industrial Electronics","FirstCategoryId":"1085","ListUrlMain":"https://ieeexplore.ieee.org/document/10614095/","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
单相交流输入的初级-并联-次级-串联 LCC 谐振高压(HV)电源已广泛应用于医用 X 射线。它同时要求极低的纹波(LCC 转换器)。因此,直流母线电容可显著降低。该策略具有无需额外硬件成本、可直接在数字控制器中实现、对系统参数精度不敏感等优点。我们在一个 120 kV/900 W 高压电源上进行了实验,证实了所提策略的有效性。峰-峰输出电压纹波被抑制到仅 0.36%,而直流电容则显著降低到 1/12。X 射线对比证明,X 射线机的图像质量因此得到了极大改善。
Output Voltage Ripple Suppression of PPSS-LCC Converter for Medical X-Ray Application Based on Repetitive Control
The primary-parallel-secondary-series
LCC
resonant high voltage (HV) power supply with single-phase ac input has been widely applied for medical X-ray application. It requires extremely low ripple (<0.5%)>LCC
converter at the same time. The dc-bus capacitance can, therefore, be remarkably reduced. This strategy has the advantages of no additional hardware costs, directly implemented in digital controller, and insensitive to accuracy of system parameters. Experiments are carried out on a practical 120 kV/900 W HV power supply and confirm the effectiveness of the proposed strategy. The peak–peak output voltage ripple is suppressed to only 0.36%, while the DC capacitance is significantly reduced to 1/12. Comparison of X-ray does prove that the image quality of X-ray machine is consequently greatly improved.