{"title":"在连续冷却装置条件下,滑动切片机和旋转切片机的切片厚度相同。","authors":"Mayuna Fukuzawa, Raia Kushibiki, Yuki Kanehira, Akifumi Ishizawa, Moe Kameda, Sayaka Kobayashi, Yoshimi Nishijima, Masanao Saio","doi":"10.1016/j.acthis.2024.152208","DOIUrl":null,"url":null,"abstract":"<div><div>To date, no report has compared section thickness (ST) between the sliding microtome (SM) and rotary microtome (RM). We used the ice pack (IP) condition, in which the paraffin block was not cooled during slicing, and a continuous cooling device (CCD) for continuous cooling during slicing. The ST was greater for the SM than the RM in the IP condition, but it was identical between the devices under the CCD condition. Thus, we used the CCD condition for subsequent studies. In formalin-fixed, paraffin-embedded (FFPE) fish sausage blocks, the ST of the tissue surface (T-surface) was significantly concaved compared to that of the paraffin surface (P-surface) for both microtomes. On the contrary, in FFPE human kidney blocks, ST did not differ between the T-surface and P-surface. Furthermore, the eosin-positive area of PAM-stained specimens was affected by ST, and the color tone of the thickest sample differed from that of the median or thinnest sample. Our data indicated that we should use CCD conditions to ensure that ST is uniform regardless of the type of microtome. In addition, for quantitative analysis, we should utilize ST measure equipment and specimens with a constant ST.</div></div>","PeriodicalId":6961,"journal":{"name":"Acta histochemica","volume":"126 8","pages":"Article 152208"},"PeriodicalIF":2.3000,"publicationDate":"2024-10-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Section thickness is identical for the sliding microtome and rotary microtome under the continuous cooling device condition\",\"authors\":\"Mayuna Fukuzawa, Raia Kushibiki, Yuki Kanehira, Akifumi Ishizawa, Moe Kameda, Sayaka Kobayashi, Yoshimi Nishijima, Masanao Saio\",\"doi\":\"10.1016/j.acthis.2024.152208\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>To date, no report has compared section thickness (ST) between the sliding microtome (SM) and rotary microtome (RM). We used the ice pack (IP) condition, in which the paraffin block was not cooled during slicing, and a continuous cooling device (CCD) for continuous cooling during slicing. The ST was greater for the SM than the RM in the IP condition, but it was identical between the devices under the CCD condition. Thus, we used the CCD condition for subsequent studies. In formalin-fixed, paraffin-embedded (FFPE) fish sausage blocks, the ST of the tissue surface (T-surface) was significantly concaved compared to that of the paraffin surface (P-surface) for both microtomes. On the contrary, in FFPE human kidney blocks, ST did not differ between the T-surface and P-surface. Furthermore, the eosin-positive area of PAM-stained specimens was affected by ST, and the color tone of the thickest sample differed from that of the median or thinnest sample. Our data indicated that we should use CCD conditions to ensure that ST is uniform regardless of the type of microtome. In addition, for quantitative analysis, we should utilize ST measure equipment and specimens with a constant ST.</div></div>\",\"PeriodicalId\":6961,\"journal\":{\"name\":\"Acta histochemica\",\"volume\":\"126 8\",\"pages\":\"Article 152208\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2024-10-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta histochemica\",\"FirstCategoryId\":\"99\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S006512812400076X\",\"RegionNum\":4,\"RegionCategory\":\"生物学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"CELL BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta histochemica","FirstCategoryId":"99","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S006512812400076X","RegionNum":4,"RegionCategory":"生物学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CELL BIOLOGY","Score":null,"Total":0}
引用次数: 0
摘要
迄今为止,还没有报告比较过滑动切片机(SM)和旋转切片机(RM)的切片厚度(ST)。我们使用了冰袋(IP)条件,即在切片过程中不冷却石蜡块,并使用持续冷却装置(CCD)在切片过程中持续冷却。在 IP 条件下,SM 的 ST 值大于 RM,但在 CCD 条件下,两种装置的 ST 值相同。因此,我们在后续研究中使用了 CCD 条件。在福尔马林固定、石蜡包埋(FFPE)的鱼肠块中,两种显微切片的组织表面(T 表面)的 ST 都比石蜡表面(P 表面)的 ST 明显凹陷。相反,在 FFPE 人肾组织块中,T 表面和 P 表面的 ST 没有差异。此外,PAM 染色样本的伊红阳性面积也受 ST 的影响,最厚样本的色调与中位样本或最薄样本的色调不同。我们的数据表明,无论使用哪种显微切片机,我们都应使用 CCD 条件来确保 ST 的一致性。此外,为了进行定量分析,我们应该使用 ST 测量设备和具有恒定 ST 的样本。
Section thickness is identical for the sliding microtome and rotary microtome under the continuous cooling device condition
To date, no report has compared section thickness (ST) between the sliding microtome (SM) and rotary microtome (RM). We used the ice pack (IP) condition, in which the paraffin block was not cooled during slicing, and a continuous cooling device (CCD) for continuous cooling during slicing. The ST was greater for the SM than the RM in the IP condition, but it was identical between the devices under the CCD condition. Thus, we used the CCD condition for subsequent studies. In formalin-fixed, paraffin-embedded (FFPE) fish sausage blocks, the ST of the tissue surface (T-surface) was significantly concaved compared to that of the paraffin surface (P-surface) for both microtomes. On the contrary, in FFPE human kidney blocks, ST did not differ between the T-surface and P-surface. Furthermore, the eosin-positive area of PAM-stained specimens was affected by ST, and the color tone of the thickest sample differed from that of the median or thinnest sample. Our data indicated that we should use CCD conditions to ensure that ST is uniform regardless of the type of microtome. In addition, for quantitative analysis, we should utilize ST measure equipment and specimens with a constant ST.
期刊介绍:
Acta histochemica, a journal of structural biochemistry of cells and tissues, publishes original research articles, short communications, reviews, letters to the editor, meeting reports and abstracts of meetings. The aim of the journal is to provide a forum for the cytochemical and histochemical research community in the life sciences, including cell biology, biotechnology, neurobiology, immunobiology, pathology, pharmacology, botany, zoology and environmental and toxicological research. The journal focuses on new developments in cytochemistry and histochemistry and their applications. Manuscripts reporting on studies of living cells and tissues are particularly welcome. Understanding the complexity of cells and tissues, i.e. their biocomplexity and biodiversity, is a major goal of the journal and reports on this topic are especially encouraged. Original research articles, short communications and reviews that report on new developments in cytochemistry and histochemistry are welcomed, especially when molecular biology is combined with the use of advanced microscopical techniques including image analysis and cytometry. Letters to the editor should comment or interpret previously published articles in the journal to trigger scientific discussions. Meeting reports are considered to be very important publications in the journal because they are excellent opportunities to present state-of-the-art overviews of fields in research where the developments are fast and hard to follow. Authors of meeting reports should consult the editors before writing a report. The editorial policy of the editors and the editorial board is rapid publication. Once a manuscript is received by one of the editors, an editorial decision about acceptance, revision or rejection will be taken within a month. It is the aim of the publishers to have a manuscript published within three months after the manuscript has been accepted