前段OCT成像PAUL®青光眼植入片:一种有效的随访和风险管理方法。

Pascal Schipper, Constance Weber, Ke Lu, Siqi Fan, Verena Prokosch, Frank G Holz, Karl Mercieca
{"title":"前段OCT成像PAUL®青光眼植入片:一种有效的随访和风险管理方法。","authors":"Pascal Schipper, Constance Weber, Ke Lu, Siqi Fan, Verena Prokosch, Frank G Holz, Karl Mercieca","doi":"10.1007/s00417-024-06708-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>To evaluate a useful, non-contact method for the follow-up of pericardium patch graft changes in patients undergoing PAUL<sup>®</sup> Glaucoma Implant (PGI) surgery using high-resolution anterior segment optical coherence tomography (OCT) to predict tube erosions.</p><p><strong>Methods: </strong>Prospective analysis over six months of tube pericardium patch graft thickness of PGI surgical cases at the University Eye Hospital Bonn, Germany, from November 2021 to August 2022. In all eyes, Tutopatch<sup>®</sup> (RTI Surgical, United States) pericardium was used to cover the implant intra-operatively. Anterior segment OCT (AS-OCT, Heidelberg ANTERION<sup>®</sup> Swept-Source-OCT) examinations were performed following a standardized protocol to measure quantitative and qualitative aspects of the patch grafts before surgery, and at three and six months after surgery.</p><p><strong>Results: </strong>Twenty-six eyes of 26 patients were included. Thickness of the patch material was 1188 µm (IQR 415 µm) directly after implantation and decreased over time to 1068 µm (IQR 478 µm) at 3 months and 846 µm (IQR 677 µm) at 6 months. No significant differences between groups were shown concerning gender (p = 0.128), ethnicity (p = 1.000), age (p = 0.741), glaucoma type (p = 0.173), other concurrent diseases (p = 0.302), former glaucoma surgeries (p = 1.000) and the quadrant of implantation (p = 0.555). Five eyes developed implant exposure. When comparing eyes with and without tube exposure, no significant differences were shown in average patch thickness above the tube directly after implantation (p = 0.476). However, significant differences in average thickness were observed at 3 months (p = 0.013) and 6 months (p = 0.005).</p><p><strong>Conclusions: </strong>Pericardial patch grafts tend to thin over time which can be assessed by AS-OCT, the latter proving to be a useful method to follow-up patients who undergo patch graft implantation during PGI surgery. This investigation could potentially help identify patients at risk of tube exposure which in turn could lead to modification of patient management. It could also possibly be used in future studies to find more suitable patch materials.</p>","PeriodicalId":12795,"journal":{"name":"Graefe’s Archive for Clinical and Experimental Ophthalmology","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2024-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Anterior segment OCT for imaging PAUL<sup>®</sup> glaucoma implant patch grafts: a useful method for follow-up and risk management.\",\"authors\":\"Pascal Schipper, Constance Weber, Ke Lu, Siqi Fan, Verena Prokosch, Frank G Holz, Karl Mercieca\",\"doi\":\"10.1007/s00417-024-06708-2\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Aim: </strong>To evaluate a useful, non-contact method for the follow-up of pericardium patch graft changes in patients undergoing PAUL<sup>®</sup> Glaucoma Implant (PGI) surgery using high-resolution anterior segment optical coherence tomography (OCT) to predict tube erosions.</p><p><strong>Methods: </strong>Prospective analysis over six months of tube pericardium patch graft thickness of PGI surgical cases at the University Eye Hospital Bonn, Germany, from November 2021 to August 2022. In all eyes, Tutopatch<sup>®</sup> (RTI Surgical, United States) pericardium was used to cover the implant intra-operatively. Anterior segment OCT (AS-OCT, Heidelberg ANTERION<sup>®</sup> Swept-Source-OCT) examinations were performed following a standardized protocol to measure quantitative and qualitative aspects of the patch grafts before surgery, and at three and six months after surgery.</p><p><strong>Results: </strong>Twenty-six eyes of 26 patients were included. Thickness of the patch material was 1188 µm (IQR 415 µm) directly after implantation and decreased over time to 1068 µm (IQR 478 µm) at 3 months and 846 µm (IQR 677 µm) at 6 months. No significant differences between groups were shown concerning gender (p = 0.128), ethnicity (p = 1.000), age (p = 0.741), glaucoma type (p = 0.173), other concurrent diseases (p = 0.302), former glaucoma surgeries (p = 1.000) and the quadrant of implantation (p = 0.555). Five eyes developed implant exposure. When comparing eyes with and without tube exposure, no significant differences were shown in average patch thickness above the tube directly after implantation (p = 0.476). However, significant differences in average thickness were observed at 3 months (p = 0.013) and 6 months (p = 0.005).</p><p><strong>Conclusions: </strong>Pericardial patch grafts tend to thin over time which can be assessed by AS-OCT, the latter proving to be a useful method to follow-up patients who undergo patch graft implantation during PGI surgery. This investigation could potentially help identify patients at risk of tube exposure which in turn could lead to modification of patient management. It could also possibly be used in future studies to find more suitable patch materials.</p>\",\"PeriodicalId\":12795,\"journal\":{\"name\":\"Graefe’s Archive for Clinical and Experimental Ophthalmology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-12-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Graefe’s Archive for Clinical and Experimental Ophthalmology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s00417-024-06708-2\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Graefe’s Archive for Clinical and Experimental Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00417-024-06708-2","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:评价采用高分辨率前段光学相干断层扫描(OCT)预测导管侵蚀的一种有用的、非接触的方法,用于随访PAUL®青光眼(PGI)手术患者心包贴片移植改变。方法:前瞻性分析2021年11月至2022年8月在德国波恩大学眼科医院进行的PGI手术患者6个月的心包贴片厚度。在所有的眼睛中,术中使用Tutopatch®(RTI Surgical,美国)心包覆盖植入物。前段OCT (AS-OCT, Heidelberg ANTERION®scan - source -OCT)检查按照标准化方案进行,以在术前和术后3个月和6个月测量膜片移植的定量和定性方面。结果:纳入26例患者26只眼。植入后膜片材料的厚度为1188µm (IQR 415µm),随着时间的推移,3个月时膜片材料的厚度为1068µm (IQR 478µm), 6个月时膜片材料的厚度为846µm (IQR 677µm)。性别(p = 0.128)、种族(p = 1.000)、年龄(p = 0.741)、青光眼类型(p = 0.173)、其他并发疾病(p = 0.302)、既往青光眼手术(p = 1.000)和植入象限(p = 0.555)组间差异无统计学意义。5只眼睛出现植入物暴露。与未置管的眼相比,置管后直接置管上方的平均眼片厚度无显著差异(p = 0.476)。然而,在3个月(p = 0.013)和6个月(p = 0.005)时,平均厚度有显著差异。结论:随着时间的推移,心包贴片逐渐变薄,可通过AS-OCT进行评估,AS-OCT是PGI手术中贴片植入术患者随访的一种有效方法。这项调查可能有助于识别有导管暴露风险的患者,从而可能导致患者管理的修改。它也可能在未来的研究中用于寻找更合适的贴片材料。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Anterior segment OCT for imaging PAUL® glaucoma implant patch grafts: a useful method for follow-up and risk management.

Aim: To evaluate a useful, non-contact method for the follow-up of pericardium patch graft changes in patients undergoing PAUL® Glaucoma Implant (PGI) surgery using high-resolution anterior segment optical coherence tomography (OCT) to predict tube erosions.

Methods: Prospective analysis over six months of tube pericardium patch graft thickness of PGI surgical cases at the University Eye Hospital Bonn, Germany, from November 2021 to August 2022. In all eyes, Tutopatch® (RTI Surgical, United States) pericardium was used to cover the implant intra-operatively. Anterior segment OCT (AS-OCT, Heidelberg ANTERION® Swept-Source-OCT) examinations were performed following a standardized protocol to measure quantitative and qualitative aspects of the patch grafts before surgery, and at three and six months after surgery.

Results: Twenty-six eyes of 26 patients were included. Thickness of the patch material was 1188 µm (IQR 415 µm) directly after implantation and decreased over time to 1068 µm (IQR 478 µm) at 3 months and 846 µm (IQR 677 µm) at 6 months. No significant differences between groups were shown concerning gender (p = 0.128), ethnicity (p = 1.000), age (p = 0.741), glaucoma type (p = 0.173), other concurrent diseases (p = 0.302), former glaucoma surgeries (p = 1.000) and the quadrant of implantation (p = 0.555). Five eyes developed implant exposure. When comparing eyes with and without tube exposure, no significant differences were shown in average patch thickness above the tube directly after implantation (p = 0.476). However, significant differences in average thickness were observed at 3 months (p = 0.013) and 6 months (p = 0.005).

Conclusions: Pericardial patch grafts tend to thin over time which can be assessed by AS-OCT, the latter proving to be a useful method to follow-up patients who undergo patch graft implantation during PGI surgery. This investigation could potentially help identify patients at risk of tube exposure which in turn could lead to modification of patient management. It could also possibly be used in future studies to find more suitable patch materials.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.40
自引率
7.40%
发文量
398
审稿时长
3 months
期刊介绍: Graefe''s Archive for Clinical and Experimental Ophthalmology is a distinguished international journal that presents original clinical reports and clini-cally relevant experimental studies. Founded in 1854 by Albrecht von Graefe to serve as a source of useful clinical information and a stimulus for discussion, the journal has published articles by leading ophthalmologists and vision research scientists for more than a century. With peer review by an international Editorial Board and prompt English-language publication, Graefe''s Archive provides rapid dissemination of clinical and clinically related experimental information.
期刊最新文献
Reexamine the link between retinal layer thickness and cognitive function after correction of axial length: the Beijing Eye Study 2011. Comparison of the clinical outcomes after manual capsulorhexis on one eye and precision pulse capsulotomy on the other eye. Reply on comments on the paper "Descemet's membrane transplantation for the treatment of recurrent high myopic macular hole associated with retinal detachment". The rAAV2-ND1 gene therapy for Leber hereditary optic neuropathy. Adverse Events Associated with Devices for Incisional Glaucoma Surgery Performed with Implants as Reported to the FDA MAUDE Database.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1