{"title":"体容积描记术测量肺容量变异性和特定气道阻力的方法学因素。","authors":"R Peslin,&nbsp;C Gallina,&nbsp;M Rotger","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Thoracic gas volume (TGV) and specific airway resistance (sRaw) are commonly measured using pressure type and flow type body plethysmographs. Within-subject variability of the data, defined as the coefficient of variation of eight to ten measurements during the same session, was assessed with the two kinds of instruments and compared in fifteen normal subjects. The reproducibility of data obtained several days apart was also compared. All measurements were made in a 480-l body chamber, which could be used in both the pressure and the flow mode. The signals were processed digitally using three different algorithms: 1) simple linear regression (LR); 2) linear regression with drift correction achieved by adding to, or subtracting from the plethysmographic signal a term proportional to time (LRC); 3) Fourier analysis (FFT). Within-subject variability of TGV was much larger with flow than with pressure plethysmography when the signals were processed by LR (14.5 +/- 7.5 vs 6.3 +/- 3.0%; p less than 0.001), but almost the same using LRC (6.7 +/- 3.2 vs 5.4 +/- 2.7%) and FFT (6.1 +/- 2.4 vs 5.0 +/- 2.4%). For sRaw, variabilities were larger and less influenced by methodological factors. Adequate digital processing may therefore largely remedy the inherently greater variability of TGV measurements with flow plethysmographs.</p>","PeriodicalId":75642,"journal":{"name":"Bulletin europeen de physiopathologie respiratoire","volume":"23 4","pages":"323-7"},"PeriodicalIF":0.0000,"publicationDate":"1987-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Methodological factors in the variability of lung volume and specific airway resistance measured by body plethysmography.\",\"authors\":\"R Peslin,&nbsp;C Gallina,&nbsp;M Rotger\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Thoracic gas volume (TGV) and specific airway resistance (sRaw) are commonly measured using pressure type and flow type body plethysmographs. Within-subject variability of the data, defined as the coefficient of variation of eight to ten measurements during the same session, was assessed with the two kinds of instruments and compared in fifteen normal subjects. The reproducibility of data obtained several days apart was also compared. All measurements were made in a 480-l body chamber, which could be used in both the pressure and the flow mode. The signals were processed digitally using three different algorithms: 1) simple linear regression (LR); 2) linear regression with drift correction achieved by adding to, or subtracting from the plethysmographic signal a term proportional to time (LRC); 3) Fourier analysis (FFT). Within-subject variability of TGV was much larger with flow than with pressure plethysmography when the signals were processed by LR (14.5 +/- 7.5 vs 6.3 +/- 3.0%; p less than 0.001), but almost the same using LRC (6.7 +/- 3.2 vs 5.4 +/- 2.7%) and FFT (6.1 +/- 2.4 vs 5.0 +/- 2.4%). For sRaw, variabilities were larger and less influenced by methodological factors. Adequate digital processing may therefore largely remedy the inherently greater variability of TGV measurements with flow plethysmographs.</p>\",\"PeriodicalId\":75642,\"journal\":{\"name\":\"Bulletin europeen de physiopathologie respiratoire\",\"volume\":\"23 4\",\"pages\":\"323-7\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1987-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bulletin europeen de physiopathologie respiratoire\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bulletin europeen de physiopathologie respiratoire","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

胸气量(TGV)和特定气道阻力(sRaw)通常使用压力型和流量型体体积描记仪测量。数据的受试者内可变性,定义为在同一时段内8到10次测量的变异系数,用两种仪器进行评估,并在15个正常受试者中进行比较。还比较了相隔几天获得的数据的再现性。所有的测量都是在480-l的体室中进行的,可以在压力和流动模式下使用。使用三种不同的算法对信号进行数字处理:1)简单线性回归(LR);2)线性回归与漂移校正,通过在容积脉搏信号中添加或减去与时间成比例的项(LRC);傅里叶分析(FFT)。当信号被LR处理时,流量的TGV在受试者内的变异性要比压力的大得多(14.5 +/- 7.5 vs 6.3 +/- 3.0%;p < 0.001),但使用LRC (6.7 +/- 3.2 vs 5.4 +/- 2.7%)和FFT (6.1 +/- 2.4 vs 5.0 +/- 2.4%)几乎相同。sRaw的变异较大,受方法学因素的影响较小。因此,充分的数字处理可以在很大程度上弥补流量容积仪测量TGV固有的更大变异性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Methodological factors in the variability of lung volume and specific airway resistance measured by body plethysmography.

Thoracic gas volume (TGV) and specific airway resistance (sRaw) are commonly measured using pressure type and flow type body plethysmographs. Within-subject variability of the data, defined as the coefficient of variation of eight to ten measurements during the same session, was assessed with the two kinds of instruments and compared in fifteen normal subjects. The reproducibility of data obtained several days apart was also compared. All measurements were made in a 480-l body chamber, which could be used in both the pressure and the flow mode. The signals were processed digitally using three different algorithms: 1) simple linear regression (LR); 2) linear regression with drift correction achieved by adding to, or subtracting from the plethysmographic signal a term proportional to time (LRC); 3) Fourier analysis (FFT). Within-subject variability of TGV was much larger with flow than with pressure plethysmography when the signals were processed by LR (14.5 +/- 7.5 vs 6.3 +/- 3.0%; p less than 0.001), but almost the same using LRC (6.7 +/- 3.2 vs 5.4 +/- 2.7%) and FFT (6.1 +/- 2.4 vs 5.0 +/- 2.4%). For sRaw, variabilities were larger and less influenced by methodological factors. Adequate digital processing may therefore largely remedy the inherently greater variability of TGV measurements with flow plethysmographs.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Status Asthmaticus The development of an animal model for TDI asthma. The role of neuropeptides as neurotransmitters of non-adrenergic, non-cholinergic nerves in bronchial asthma. Epidemiological measurement of bronchial responsiveness in polyurethane workers. Late asthmatic reactions and changes in histamine responsiveness provoked by occupational agents.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1