机器人样品处理机的准确度和精密度。

D A Armbruster, L C Hawes, C T Winter
{"title":"机器人样品处理机的准确度和精密度。","authors":"D A Armbruster,&nbsp;L C Hawes,&nbsp;C T Winter","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The accuracy and precision of the Packard Probe 1000, a computer-driven robotic sample processor, were determined using NCCLS Guideline 18-P, employing a rigorous gravimetric procedure. For volumes of 25, 100, and 200 muL, inaccuracy ranged from -0.5 to +0.6% (saline) and -0.8 to +0.7% (human serum), and imprecision from 0.1 to 0.6% (saline) and 0.02 to 0.8% (human serum), in the dip and sip mode of pipetting. For the same volumes of saline in the dispense-through pipetting mode, inaccuracy ranged from -0.1 to -0.9% and imprecision from 0.1 to 0.5%. To obtain this level of performance, the instrument's operating parameters were changed from the factory default settings. This automated liquid-handling system is capable of impressive accuracy and precision, but it must be calibrated in the same manner as any other volumetric device. Performance varies with the volume, type of liquid, and mode of pipetting. Inaccuracy of less than or equal to 1.0% and imprecision of less than =1.0% can be attained along with rapid, walkaway pipetting of calibrators, controls, and large numbers of patient samples.</p>","PeriodicalId":80043,"journal":{"name":"Journal of the International Federation of Clinical Chemistry","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1992-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy and precision of a robotic sample processor.\",\"authors\":\"D A Armbruster,&nbsp;L C Hawes,&nbsp;C T Winter\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The accuracy and precision of the Packard Probe 1000, a computer-driven robotic sample processor, were determined using NCCLS Guideline 18-P, employing a rigorous gravimetric procedure. For volumes of 25, 100, and 200 muL, inaccuracy ranged from -0.5 to +0.6% (saline) and -0.8 to +0.7% (human serum), and imprecision from 0.1 to 0.6% (saline) and 0.02 to 0.8% (human serum), in the dip and sip mode of pipetting. For the same volumes of saline in the dispense-through pipetting mode, inaccuracy ranged from -0.1 to -0.9% and imprecision from 0.1 to 0.5%. To obtain this level of performance, the instrument's operating parameters were changed from the factory default settings. This automated liquid-handling system is capable of impressive accuracy and precision, but it must be calibrated in the same manner as any other volumetric device. Performance varies with the volume, type of liquid, and mode of pipetting. Inaccuracy of less than or equal to 1.0% and imprecision of less than =1.0% can be attained along with rapid, walkaway pipetting of calibrators, controls, and large numbers of patient samples.</p>\",\"PeriodicalId\":80043,\"journal\":{\"name\":\"Journal of the International Federation of Clinical Chemistry\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1992-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of the International Federation of Clinical Chemistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the International Federation of Clinical Chemistry","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

Packard Probe 1000(计算机驱动的机器人样品处理机)的准确度和精密度采用NCCLS指南18-P,采用严格的重量测定程序进行测定。对于25,100和200mul的体积,在dip和sip移液模式下,不精确性范围为-0.5至+0.6%(生理盐水)和-0.8至+0.7%(人血清),不精确性范围为0.1至0.6%(生理盐水)和0.02至0.8%(人血清)。对于相同体积的生理盐水,在点液移液模式下,不精确范围为-0.1至-0.9%,不精确范围为0.1至0.5%。为了获得这种性能水平,仪器的操作参数从出厂默认设置更改。这种自动化液体处理系统具有令人印象深刻的准确度和精度,但它必须以与任何其他体积测量设备相同的方式进行校准。性能随体积、液体类型和移液方式的不同而变化。不准确性小于或等于1.0%,不精密度小于=1.0%,可实现快速、走开移液校准器、对照品和大量患者样品。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Accuracy and precision of a robotic sample processor.

The accuracy and precision of the Packard Probe 1000, a computer-driven robotic sample processor, were determined using NCCLS Guideline 18-P, employing a rigorous gravimetric procedure. For volumes of 25, 100, and 200 muL, inaccuracy ranged from -0.5 to +0.6% (saline) and -0.8 to +0.7% (human serum), and imprecision from 0.1 to 0.6% (saline) and 0.02 to 0.8% (human serum), in the dip and sip mode of pipetting. For the same volumes of saline in the dispense-through pipetting mode, inaccuracy ranged from -0.1 to -0.9% and imprecision from 0.1 to 0.5%. To obtain this level of performance, the instrument's operating parameters were changed from the factory default settings. This automated liquid-handling system is capable of impressive accuracy and precision, but it must be calibrated in the same manner as any other volumetric device. Performance varies with the volume, type of liquid, and mode of pipetting. Inaccuracy of less than or equal to 1.0% and imprecision of less than =1.0% can be attained along with rapid, walkaway pipetting of calibrators, controls, and large numbers of patient samples.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
What is the most accurate, cost-effective way to diagnose acute viral hepatitis? Use of antioxidants in the prevention and treatment of disease. Aging and oxidative stress. Magnetic resonance spectroscopy in clinical chemistry: the present and the future. Methods for detection of point mutations: performance and quality assessment. The IFCC Scientific Division, Committee on Molecular Biology Techniques.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1