{"title":"听力计校准:解释和应用标准。","authors":"G R Lightfoot","doi":"10.3109/03005364000000143","DOIUrl":null,"url":null,"abstract":"<p><p>A degree of judgement is required when applying the British Standard associated with the calibration and performance of audiometers, BS EN 60645-1 (IEC 645-1). Consequently there is potential for variation of practice between calibration agencies. This note addresses issues and makes suggestions relating to the measurement of distortion, on-off effect, cross talk, linearity, rise/fall time and to the positioning of the earphone on an artificial ear. The degree of precision appropriate to each measurement is suggested. A warning is given of the audiological consequences of excessive distortion produced by normally functioning audiometers.</p>","PeriodicalId":75616,"journal":{"name":"British journal of audiology","volume":"34 5","pages":"311-6"},"PeriodicalIF":0.0000,"publicationDate":"2000-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/03005364000000143","citationCount":"10","resultStr":"{\"title\":\"Audiometer calibration: interpreting and applying the standards.\",\"authors\":\"G R Lightfoot\",\"doi\":\"10.3109/03005364000000143\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A degree of judgement is required when applying the British Standard associated with the calibration and performance of audiometers, BS EN 60645-1 (IEC 645-1). Consequently there is potential for variation of practice between calibration agencies. This note addresses issues and makes suggestions relating to the measurement of distortion, on-off effect, cross talk, linearity, rise/fall time and to the positioning of the earphone on an artificial ear. The degree of precision appropriate to each measurement is suggested. A warning is given of the audiological consequences of excessive distortion produced by normally functioning audiometers.</p>\",\"PeriodicalId\":75616,\"journal\":{\"name\":\"British journal of audiology\",\"volume\":\"34 5\",\"pages\":\"311-6\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2000-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.3109/03005364000000143\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"British journal of audiology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.3109/03005364000000143\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"British journal of audiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/03005364000000143","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10
摘要
当应用与听力计的校准和性能相关的英国标准BS EN 60645-1 (IEC 645-1)时,需要一定程度的判断。因此,校准机构之间的做法可能存在差异。本说明针对失真、开关效应、串音、线性度、上升/下降时间的测量以及耳机在人造耳上的定位提出问题和建议。提出了适合于每次测量的精度。对正常工作的听力学计产生的过度失真的听力学后果提出警告。
Audiometer calibration: interpreting and applying the standards.
A degree of judgement is required when applying the British Standard associated with the calibration and performance of audiometers, BS EN 60645-1 (IEC 645-1). Consequently there is potential for variation of practice between calibration agencies. This note addresses issues and makes suggestions relating to the measurement of distortion, on-off effect, cross talk, linearity, rise/fall time and to the positioning of the earphone on an artificial ear. The degree of precision appropriate to each measurement is suggested. A warning is given of the audiological consequences of excessive distortion produced by normally functioning audiometers.