{"title":"保留在避免排出宫内节育器中的作用——用于基础研究的测量装置。","authors":"K H Kurz","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>320 medicated IUDs were removed from 292 women by means of the dynamometer. This instrument indicates the highest retention force (HRF) in newtons (N). Ages of the women ranged from 17-50 years. The IUDs were fitted in nulligravidae (56%), nulliparous with 1 or more abortions (13%), primiparous (18%), and multiparous (13%) women. The IUDs used were the ML Cu 250-standard, 7 Cu 200, T Cu 200-adapted, with a shortened transversal arm of 15-30 mm and T Cu 200 standard. With the low expulsion rate, the ML Cu 250-standard IUD needed 5.8-5.4N and the 7 Cu 200 3.8-2.6N. The adapted T Cu 200 IUD needed 2.6-1.6N, i.e., about 1N more thant the T Cu 200-standard with an HRF of 1.7-1.0N. As the dynamomoeter only indicated the highest retention force, a retentiograph was devised to demonstrate the different retention forces and phases of the various IUDs by characteristic curves. The possible connections between retention and expulsion of IUDs are discussed. The retention forces of the different IUDs compared are congruent with the published expulsion rates. The retentiograph showed new patterns to test IUDs before they demonstrate side effects in vitro.</p>","PeriodicalId":84493,"journal":{"name":"Contraceptive delivery systems","volume":"3 2","pages":"107-16"},"PeriodicalIF":0.0000,"publicationDate":"1982-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Role of retention in avoiding expulsion of IUDs--measuring devices for basic research.\",\"authors\":\"K H Kurz\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>320 medicated IUDs were removed from 292 women by means of the dynamometer. This instrument indicates the highest retention force (HRF) in newtons (N). Ages of the women ranged from 17-50 years. The IUDs were fitted in nulligravidae (56%), nulliparous with 1 or more abortions (13%), primiparous (18%), and multiparous (13%) women. The IUDs used were the ML Cu 250-standard, 7 Cu 200, T Cu 200-adapted, with a shortened transversal arm of 15-30 mm and T Cu 200 standard. With the low expulsion rate, the ML Cu 250-standard IUD needed 5.8-5.4N and the 7 Cu 200 3.8-2.6N. The adapted T Cu 200 IUD needed 2.6-1.6N, i.e., about 1N more thant the T Cu 200-standard with an HRF of 1.7-1.0N. As the dynamomoeter only indicated the highest retention force, a retentiograph was devised to demonstrate the different retention forces and phases of the various IUDs by characteristic curves. The possible connections between retention and expulsion of IUDs are discussed. The retention forces of the different IUDs compared are congruent with the published expulsion rates. The retentiograph showed new patterns to test IUDs before they demonstrate side effects in vitro.</p>\",\"PeriodicalId\":84493,\"journal\":{\"name\":\"Contraceptive delivery systems\",\"volume\":\"3 2\",\"pages\":\"107-16\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1982-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Contraceptive delivery systems\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Contraceptive delivery systems","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
通过测功仪从292名妇女中取出了320个药物宫内节育器。该仪器显示以牛顿(N)为单位的最高保留力(HRF)。女性年龄从17-50岁不等。宫内节育器适用于无孕(56%)、有1次或1次以上流产的无孕(13%)、初产(18%)和多产(13%)妇女。宫内节育器采用ML - cu250标准、7 - cu200、T - cu200改进型,横臂缩短15-30 mm,采用T - cu200标准。ML Cu 250标准节育器排液率较低,排液率为5.8 ~ 5.4 n, 7 Cu 200标准节育器排液率为3.8 ~ 2.6 n。改造后的T Cu 200 IUD需要2.6-1.6N,比T Cu 200标准增加约1N, HRF为1.7-1.0N。由于动力计只能显示最高的保留力,因此设计了保留图,通过特征曲线来显示各种宫内节育器的不同保留力和相。讨论了宫内节育器保留和排出之间可能存在的联系。不同宫内节育器的保留力与公布的排出率一致。保留图显示了在宫内节育器显示出体外副作用之前测试的新模式。
Role of retention in avoiding expulsion of IUDs--measuring devices for basic research.
320 medicated IUDs were removed from 292 women by means of the dynamometer. This instrument indicates the highest retention force (HRF) in newtons (N). Ages of the women ranged from 17-50 years. The IUDs were fitted in nulligravidae (56%), nulliparous with 1 or more abortions (13%), primiparous (18%), and multiparous (13%) women. The IUDs used were the ML Cu 250-standard, 7 Cu 200, T Cu 200-adapted, with a shortened transversal arm of 15-30 mm and T Cu 200 standard. With the low expulsion rate, the ML Cu 250-standard IUD needed 5.8-5.4N and the 7 Cu 200 3.8-2.6N. The adapted T Cu 200 IUD needed 2.6-1.6N, i.e., about 1N more thant the T Cu 200-standard with an HRF of 1.7-1.0N. As the dynamomoeter only indicated the highest retention force, a retentiograph was devised to demonstrate the different retention forces and phases of the various IUDs by characteristic curves. The possible connections between retention and expulsion of IUDs are discussed. The retention forces of the different IUDs compared are congruent with the published expulsion rates. The retentiograph showed new patterns to test IUDs before they demonstrate side effects in vitro.