喷雾参数对冷冻治疗中皮肤肿瘤消融体积的影响。

Q3 Biochemistry, Genetics and Molecular Biology Australasian Physical & Engineering Sciences in Medicine Pub Date : 2019-06-01 Epub Date: 2019-03-05 DOI:10.1007/s13246-019-00740-x
Chandrika Kumari, Amitesh Kumar, Sunil Kumar Sarangi, Arunachalam Thirugnanam
{"title":"喷雾参数对冷冻治疗中皮肤肿瘤消融体积的影响。","authors":"Chandrika Kumari,&nbsp;Amitesh Kumar,&nbsp;Sunil Kumar Sarangi,&nbsp;Arunachalam Thirugnanam","doi":"10.1007/s13246-019-00740-x","DOIUrl":null,"url":null,"abstract":"<p><p>The main purpose of the study is to establish correlations for the ablation volume and the ice front as a function of the spray parameters. The ablation volume and the ice front depend upon the nozzle diameter, spraying distance and the freeze duration (spray parameters). The estimation of the ablation volume using the spray parameters shall be useful in surgical practice to ablate the different sizes of tumours. Liquid nitrogen spray cooling is carried out with 0.8 mm, 0.6 mm and 0.4 mm nozzle diameters. The spraying distance is maintained at 9 mm, 18 mm and 27 mm. The spray cooling is carried out for a single freeze-thaw cycle where freezing and thawing consist of 120 s and 130 s duration respectively. A two-dimensional heat flow equation with phase change is considered for the numerical study. The numerically calculated transient temperature (2 mm and 5 mm from the gel surface) and ice front values show confirmatory results with the experimentally measured data. Correlations are obtained to determine the ablation volume (- 50 °C and - 25 °C isothermal surfaces) and ice front (axial and lateral) with a goodness of fit [Formula: see text] 95%. The nozzle diameter has a greater impact on the ablation volume as compared to the spraying distance during 120 s of freezing. The nozzle diameter of 0.8 mm, 0.6 mm and 0.4 mm can be effectively used for cryotherapy with spraying distance up to 27 mm, 18 mm and 9 mm respectively.</p>","PeriodicalId":55430,"journal":{"name":"Australasian Physical & Engineering Sciences in Medicine","volume":"42 2","pages":"453-464"},"PeriodicalIF":0.0000,"publicationDate":"2019-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1007/s13246-019-00740-x","citationCount":"8","resultStr":"{\"title\":\"Effects of spray parameters on skin tumour ablation volume during cryotherapy.\",\"authors\":\"Chandrika Kumari,&nbsp;Amitesh Kumar,&nbsp;Sunil Kumar Sarangi,&nbsp;Arunachalam Thirugnanam\",\"doi\":\"10.1007/s13246-019-00740-x\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The main purpose of the study is to establish correlations for the ablation volume and the ice front as a function of the spray parameters. The ablation volume and the ice front depend upon the nozzle diameter, spraying distance and the freeze duration (spray parameters). The estimation of the ablation volume using the spray parameters shall be useful in surgical practice to ablate the different sizes of tumours. Liquid nitrogen spray cooling is carried out with 0.8 mm, 0.6 mm and 0.4 mm nozzle diameters. The spraying distance is maintained at 9 mm, 18 mm and 27 mm. The spray cooling is carried out for a single freeze-thaw cycle where freezing and thawing consist of 120 s and 130 s duration respectively. A two-dimensional heat flow equation with phase change is considered for the numerical study. The numerically calculated transient temperature (2 mm and 5 mm from the gel surface) and ice front values show confirmatory results with the experimentally measured data. Correlations are obtained to determine the ablation volume (- 50 °C and - 25 °C isothermal surfaces) and ice front (axial and lateral) with a goodness of fit [Formula: see text] 95%. The nozzle diameter has a greater impact on the ablation volume as compared to the spraying distance during 120 s of freezing. The nozzle diameter of 0.8 mm, 0.6 mm and 0.4 mm can be effectively used for cryotherapy with spraying distance up to 27 mm, 18 mm and 9 mm respectively.</p>\",\"PeriodicalId\":55430,\"journal\":{\"name\":\"Australasian Physical & Engineering Sciences in Medicine\",\"volume\":\"42 2\",\"pages\":\"453-464\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1007/s13246-019-00740-x\",\"citationCount\":\"8\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Australasian Physical & Engineering Sciences in Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s13246-019-00740-x\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2019/3/5 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"Biochemistry, Genetics and Molecular Biology\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Australasian Physical & Engineering Sciences in Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s13246-019-00740-x","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2019/3/5 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"Biochemistry, Genetics and Molecular Biology","Score":null,"Total":0}
引用次数: 8

摘要

研究的主要目的是建立消融体积和冰锋作为喷雾参数函数的相关性。消融体积和冰锋取决于喷嘴直径、喷射距离和冻结时间(喷射参数)。利用喷雾器参数估算消融体积,在手术实践中可用于消融不同大小的肿瘤。液氮喷雾冷却喷嘴直径分别为0.8 mm、0.6 mm和0.4 mm。喷涂距离保持在9mm、18mm和27mm。喷雾冷却为单次冻融循环,冻结时间为120秒,解冻时间为130秒。采用二维相变热流方程进行数值研究。数值计算的瞬态温度(距凝胶表面2 mm和5 mm)和冰锋值与实验测量数据吻合。获得了确定消融体积(- 50°C和- 25°C等温表面)和冰锋(轴向和侧向)的相关性,拟合优度为95%[公式:见文本]。在冻结120 s期间,喷嘴直径对烧蚀体积的影响大于喷射距离。喷嘴直径为0.8 mm、0.6 mm和0.4 mm,可有效用于冷冻治疗,喷雾距离分别可达27mm、18mm和9mm。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Effects of spray parameters on skin tumour ablation volume during cryotherapy.

The main purpose of the study is to establish correlations for the ablation volume and the ice front as a function of the spray parameters. The ablation volume and the ice front depend upon the nozzle diameter, spraying distance and the freeze duration (spray parameters). The estimation of the ablation volume using the spray parameters shall be useful in surgical practice to ablate the different sizes of tumours. Liquid nitrogen spray cooling is carried out with 0.8 mm, 0.6 mm and 0.4 mm nozzle diameters. The spraying distance is maintained at 9 mm, 18 mm and 27 mm. The spray cooling is carried out for a single freeze-thaw cycle where freezing and thawing consist of 120 s and 130 s duration respectively. A two-dimensional heat flow equation with phase change is considered for the numerical study. The numerically calculated transient temperature (2 mm and 5 mm from the gel surface) and ice front values show confirmatory results with the experimentally measured data. Correlations are obtained to determine the ablation volume (- 50 °C and - 25 °C isothermal surfaces) and ice front (axial and lateral) with a goodness of fit [Formula: see text] 95%. The nozzle diameter has a greater impact on the ablation volume as compared to the spraying distance during 120 s of freezing. The nozzle diameter of 0.8 mm, 0.6 mm and 0.4 mm can be effectively used for cryotherapy with spraying distance up to 27 mm, 18 mm and 9 mm respectively.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.00
自引率
0.00%
发文量
0
审稿时长
6-12 weeks
期刊介绍: Australasian Physical & Engineering Sciences in Medicine (APESM) is a multidisciplinary forum for information and research on the application of physics and engineering to medicine and human physiology. APESM covers a broad range of topics that include but is not limited to: - Medical physics in radiotherapy - Medical physics in diagnostic radiology - Medical physics in nuclear medicine - Mathematical modelling applied to medicine and human biology - Clinical biomedical engineering - Feature extraction, classification of EEG, ECG, EMG, EOG, and other biomedical signals; - Medical imaging - contributions to new and improved methods; - Modelling of physiological systems - Image processing to extract information from images, e.g. fMRI, CT, etc.; - Biomechanics, especially with applications to orthopaedics. - Nanotechnology in medicine APESM offers original reviews, scientific papers, scientific notes, technical papers, educational notes, book reviews and letters to the editor. APESM is the journal of the Australasian College of Physical Scientists and Engineers in Medicine, and also the official journal of the College of Biomedical Engineers, Engineers Australia and the Asia-Oceania Federation of Organizations for Medical Physics.
期刊最新文献
Acknowledgment of Reviewers for Volume 35 Acknowledgment of Reviewers for Volume 34 A comparison between EPSON V700 and EPSON V800 scanners for film dosimetry. Nanodosimetric understanding to the dependence of the relationship between dose-averaged lineal energy on nanoscale and LET on ion species. EPSM 2019, Engineering and Physical Sciences in Medicine : 28-30 October 2019, Perth, Australia.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1