{"title":"不可逆电穿孔治疗转移性直肠腺癌后输尿管支架断裂。","authors":"Harjivan Kohli,&nbsp;Alexander Tapper,&nbsp;James Relle","doi":"10.1089/cren.2017.0111","DOIUrl":null,"url":null,"abstract":"<p><p><b><i>Background:</i></b> Irreversible electroporation (IRE) is a soft tissue ablation technique using electrical pulses without thermal energy to create pores in the cell membrane, resulting in death from apoptosis rather than necrosis. Advantages include protection of blood vessels, nerves, and surrounding structures. Documented complications include periprocedure nausea/vomiting, infection, and severe pain. Ureteral stents are frequently used in management of hydronephrosis caused by malignant obstruction. We describe what is to our knowledge the first documentation of stent fragmentation secondary to IRE and subsequent management. <b><i>Case Presentation:</i></b> This is a 61-year-old male with history of metastatic rectal adenocarcinoma treated initially with chemotherapy and surgery. Follow-up imaging revealed hydronephrosis and enlarged right iliac lymph node. Ureteral stent was placed for management of the hydronephrosis and the patient was referred to undergo IRE for management of metastatic disease. After treatment, the patient had imaging performed that showed fractured right ureteral stent with proximal portion in the ureter and distal portion floating freely in the bladder. This complication was managed with staged endoscopic procedure involving adjacent ureteral stent placement and subsequent ureteroscopy and stent removal using delta grasper. <b><i>Conclusion:</i></b> We describe to our knowledge the first incidence as well as subsequent management of ureteral stent fracture from an increasingly common treatment modality for metastatic disease. Given the frequency of malignant ureteral obstruction managed with ureteral stents, knowledge of potential complications pertaining to the urologist is imperative.</p>","PeriodicalId":36779,"journal":{"name":"Journal of Endourology Case Reports","volume":"6 4","pages":"548-550"},"PeriodicalIF":0.0000,"publicationDate":"2020-12-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7803275/pdf/cren.2017.0111.pdf","citationCount":"0","resultStr":"{\"title\":\"Ureteral Stent Fracture After Irreversible Electroporation for Treatment of Metastatic Rectal Adenocarcinoma.\",\"authors\":\"Harjivan Kohli,&nbsp;Alexander Tapper,&nbsp;James Relle\",\"doi\":\"10.1089/cren.2017.0111\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b><i>Background:</i></b> Irreversible electroporation (IRE) is a soft tissue ablation technique using electrical pulses without thermal energy to create pores in the cell membrane, resulting in death from apoptosis rather than necrosis. Advantages include protection of blood vessels, nerves, and surrounding structures. Documented complications include periprocedure nausea/vomiting, infection, and severe pain. Ureteral stents are frequently used in management of hydronephrosis caused by malignant obstruction. We describe what is to our knowledge the first documentation of stent fragmentation secondary to IRE and subsequent management. <b><i>Case Presentation:</i></b> This is a 61-year-old male with history of metastatic rectal adenocarcinoma treated initially with chemotherapy and surgery. Follow-up imaging revealed hydronephrosis and enlarged right iliac lymph node. Ureteral stent was placed for management of the hydronephrosis and the patient was referred to undergo IRE for management of metastatic disease. After treatment, the patient had imaging performed that showed fractured right ureteral stent with proximal portion in the ureter and distal portion floating freely in the bladder. This complication was managed with staged endoscopic procedure involving adjacent ureteral stent placement and subsequent ureteroscopy and stent removal using delta grasper. <b><i>Conclusion:</i></b> We describe to our knowledge the first incidence as well as subsequent management of ureteral stent fracture from an increasingly common treatment modality for metastatic disease. Given the frequency of malignant ureteral obstruction managed with ureteral stents, knowledge of potential complications pertaining to the urologist is imperative.</p>\",\"PeriodicalId\":36779,\"journal\":{\"name\":\"Journal of Endourology Case Reports\",\"volume\":\"6 4\",\"pages\":\"548-550\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2020-12-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7803275/pdf/cren.2017.0111.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Endourology Case Reports\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1089/cren.2017.0111\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2020/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Endourology Case Reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1089/cren.2017.0111","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2020/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

背景:不可逆电穿孔(IRE)是一种使用无热能的电脉冲在细胞膜上形成孔的软组织消融技术,导致细胞凋亡而不是坏死。优点包括保护血管、神经和周围结构。记录的并发症包括围手术期恶心/呕吐、感染和剧烈疼痛。输尿管支架常用于治疗恶性梗阻引起的肾积水。我们描述了据我们所知的第一个关于IRE继发支架碎裂和后续处理的文献。病例介绍:这是一个61岁的男性,有转移性直肠腺癌的病史,最初接受化疗和手术治疗。随访影像显示肾积水及右侧髂淋巴结肿大。输尿管支架放置治疗肾积水,患者接受IRE治疗以治疗转移性疾病。治疗后,患者影像学显示右侧输尿管支架断裂,近端部分在输尿管内,远端部分在膀胱内自由漂浮。该并发症通过分阶段的内窥镜手术处理,包括邻近输尿管支架放置和随后的输尿管镜检查,并使用delta钳取出支架。结论:据我们所知,输尿管支架骨折是一种越来越常见的转移性疾病的治疗方式,我们描述了输尿管支架骨折的首次发病率以及随后的治疗。考虑到输尿管支架治疗恶性输尿管梗阻的频率,泌尿科医生对潜在并发症的了解是必要的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Ureteral Stent Fracture After Irreversible Electroporation for Treatment of Metastatic Rectal Adenocarcinoma.

Background: Irreversible electroporation (IRE) is a soft tissue ablation technique using electrical pulses without thermal energy to create pores in the cell membrane, resulting in death from apoptosis rather than necrosis. Advantages include protection of blood vessels, nerves, and surrounding structures. Documented complications include periprocedure nausea/vomiting, infection, and severe pain. Ureteral stents are frequently used in management of hydronephrosis caused by malignant obstruction. We describe what is to our knowledge the first documentation of stent fragmentation secondary to IRE and subsequent management. Case Presentation: This is a 61-year-old male with history of metastatic rectal adenocarcinoma treated initially with chemotherapy and surgery. Follow-up imaging revealed hydronephrosis and enlarged right iliac lymph node. Ureteral stent was placed for management of the hydronephrosis and the patient was referred to undergo IRE for management of metastatic disease. After treatment, the patient had imaging performed that showed fractured right ureteral stent with proximal portion in the ureter and distal portion floating freely in the bladder. This complication was managed with staged endoscopic procedure involving adjacent ureteral stent placement and subsequent ureteroscopy and stent removal using delta grasper. Conclusion: We describe to our knowledge the first incidence as well as subsequent management of ureteral stent fracture from an increasingly common treatment modality for metastatic disease. Given the frequency of malignant ureteral obstruction managed with ureteral stents, knowledge of potential complications pertaining to the urologist is imperative.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
A Rare Complication of Amplatz Sheath: Amplatz Sheath Rupture During Percutaneous Nephrolithotomy. Patient's Fear of Being Infected, Another Complication of COVID-19 Outbreak: A Lesson Learned from a Case of Life-Threatening Urolithiasis. Inflammatory Pseudotumor of the Renal Pelvis. Endoscopic Management of Chyluria Caused by Pyelolymphatic Fistula After Robot-Assisted Laparoscopic Pyeloplasty. Ureteroiliac Fistula: Bleeding of Unknown Origin-Case Report and Review of the Literature.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1