{"title":"Roche Cobas t711分析仪凝血测定的评估:性能和临床意义。","authors":"María Ordóñez-Robles, Oscar D Pons-Belda, María José Moína, Ángel Bernardo-Gutiérrez, Belén Prieto-García","doi":"10.1097/MBC.0000000000001261","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>We performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation.</p><p><strong>Methods: </strong>Imprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression.</p><p><strong>Results: </strong>Imprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711.</p><p><strong>Conclusion: </strong>The strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.</p>","PeriodicalId":8992,"journal":{"name":"Blood Coagulation & Fibrinolysis","volume":null,"pages":null},"PeriodicalIF":1.2000,"publicationDate":"2023-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Assessment of coagulation assays on Roche Cobas t711 analyzer: performance and clinical implications.\",\"authors\":\"María Ordóñez-Robles, Oscar D Pons-Belda, María José Moína, Ángel Bernardo-Gutiérrez, Belén Prieto-García\",\"doi\":\"10.1097/MBC.0000000000001261\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>We performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation.</p><p><strong>Methods: </strong>Imprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression.</p><p><strong>Results: </strong>Imprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711.</p><p><strong>Conclusion: </strong>The strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.</p>\",\"PeriodicalId\":8992,\"journal\":{\"name\":\"Blood Coagulation & Fibrinolysis\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2023-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Blood Coagulation & Fibrinolysis\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/MBC.0000000000001261\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/10/11 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"HEMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Blood Coagulation & Fibrinolysis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/MBC.0000000000001261","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/10/11 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"HEMATOLOGY","Score":null,"Total":0}
引用次数: 0
摘要
目的:我们在Roche Cobas t711分析仪上对五项凝血测试[即凝血酶原时间(PT)、活化部分凝血活酶时间(aPTT)、纤维蛋白原、凝血酶时间(TT)和D-二聚体]进行了分析评估,并与我们实验室使用的方法(即Werfen ACL Top 750分析仪)进行了比较研究,将分析扩展到Cobas t711实施的临床意义。方法:根据临床和实验室标准研究所(CLSI)H57A:2008指南进行不精确性研究。根据CLSI EP06-A:2003的建议分析D-二聚体和纤维蛋白原测试的线性。为了进行方法比较,使用Bland-Altman图和Passing-Balok回归分析结果。结果:不精确性符合制造商对PT、aPTT和TT的要求。D-二聚体和纤维蛋白原测试显示,在某些浓度水平下,与制造商的要求相比,变异系数(CV)为%。无法验证线性范围。比较研究表明,任何测试的结果都不能互换,aPTT测试的相关性较低,罗氏Cobas t711的D-二聚体结果较低。结论:与我们实验室使用的方法相比,本研究的优势在于分析报告Cobas t 711结果的临床意义。对因子缺乏、抗凝治疗和干扰的不同敏感性可能解释了aPTT测试获得的较低相关性。用于D-二聚体测定的不同单克隆抗体可能解释了Cobas t711分析仪获得的较低结果。鉴于D-二聚体检测与排除血栓事件的相关性,这方面需要进一步研究,并加强了止血实验室的协调需求。
Assessment of coagulation assays on Roche Cobas t711 analyzer: performance and clinical implications.
Objectives: We performed an analytical assessment of five coagulation tests [i.e. prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen, thrombin time (TT) and D-dimer] on the Roche Cobas t711 analyzer and a comparison study with the methodology in use at our laboratory (i.e. Werfen ACL Top 750 analyzer), expanding the analysis to the clinical implications of Cobas t711 implementation.
Methods: Imprecision studies were performed following the Clinical and Laboratory Standards Institute (CLSI) H57 A:2008 guideline. Linearity of D-dimer and fibrinogen tests was analysed according to the CLSI EP06-A: 2003 recommendations. For method comparison, the results were analyzed using the Bland-Altman plot and Passing-Bablok regression.
Results: Imprecision met manufacturer claims for PT, aPTT and TT. D-dimer and fibrinogen tests showed a coefficient of variation (CV)% over manufacturer claims at certain concentration levels. Linearity ranges could not be verified. Comparison study revealed that results are not interchangeable for any test, a lower correlation for aPTT test and lower D-dimer results from Roche Cobas t711.
Conclusion: The strength of this study relies on the analysis of the clinical implications of reporting Cobas t711 results compared to those obtained with the methodology in use at our laboratory. Different sensibility to factor deficiency, anticoagulant therapy and interferences might explain lower correlation rates obtained for the aPTT test. Different monoclonal antibodies used for D-dimer determination might explain the lower results obtained with the Cobas t711 analyzer. This aspect needs further studies given the relevance of D-dimer test to exclude thrombotic events and reinforces the need of harmonization in the haemostasis laboratory.
期刊介绍:
Blood Coagulation & Fibrinolysis is an international fully refereed journal that features review and original research articles on all clinical, laboratory and experimental aspects of haemostasis and thrombosis. The journal is devoted to publishing significant developments worldwide in the field of blood coagulation, fibrinolysis, thrombosis, platelets and the kininogen-kinin system, as well as dealing with those aspects of blood rheology relevant to haemostasis and the effects of drugs on haemostatic components