加速器车载成像系统图像质量参数及质量保证阈值设置的定量评价

Huang Jinyan, Pei Yuntong, Ma Yangguang, Wang Haiyang, L. Lele, Guo Yuexin
{"title":"加速器车载成像系统图像质量参数及质量保证阈值设置的定量评价","authors":"Huang Jinyan, Pei Yuntong, Ma Yangguang, Wang Haiyang, L. Lele, Guo Yuexin","doi":"10.3760/CMA.J.ISSN.1004-4221.2019.12.009","DOIUrl":null,"url":null,"abstract":"Objective \nTo achieve quantitative analysis of image quality parameters and establish warning and action thresholds for the on-board imaging (OBI) system of linear accelerator. \n \n \nMethods \nThe Catphan604 phantom was repeatedly scanned in the Full-Fan and Half-Fan CBCT scanning modes on a Varian EDGE linear accelerator, and the software based on Python language development in-house was utilized to analyze image quality parameters, such as CT number linearity, geometric consistency, slice thickness, spatial resolution, uniformity and low-contrast resolution. The quantitative analysis results of each image quality parameter obtained from 16 times of scanning within 16 months were normalized to the mean and the standard deviations were recorded. A run chart analysis was created to determine the warnings and action thresholds. \n \n \nResults \nThe software built in-house can quantitatively analyze the image parameters of the two scanning modes of OBI system. The low-contrast resolution of Half-Fan was better than that of Full-Fan, whereas the spatial resolution of Full-Fan was superior to that of Half-Fan. One standard deviation (1σ) was set as the warning threshold and 2 standard deviations (2σ) as the action threshold, respectively. The tolerance level of Half-Fan was smaller than that of Full-Fan. \n \n \nConclusion \nSelf-developed software enables quantitative analysis of accelerator image quality parameters, establishes warning and action tolerance of quality assurance and provides guidance for image quality assurance under image-guided radiotherapy specification. \n \n \nKey words: \nOn-board imaging; Image quality; Quantitative evaluation; Threshold setting","PeriodicalId":10288,"journal":{"name":"中华放射肿瘤学杂志","volume":"28 1","pages":"919-923"},"PeriodicalIF":0.0000,"publicationDate":"2019-12-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A quantitative evaluation on the image-quality parameters and quality assurance thresholds setting of accelerator on-board imaging system\",\"authors\":\"Huang Jinyan, Pei Yuntong, Ma Yangguang, Wang Haiyang, L. Lele, Guo Yuexin\",\"doi\":\"10.3760/CMA.J.ISSN.1004-4221.2019.12.009\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective \\nTo achieve quantitative analysis of image quality parameters and establish warning and action thresholds for the on-board imaging (OBI) system of linear accelerator. \\n \\n \\nMethods \\nThe Catphan604 phantom was repeatedly scanned in the Full-Fan and Half-Fan CBCT scanning modes on a Varian EDGE linear accelerator, and the software based on Python language development in-house was utilized to analyze image quality parameters, such as CT number linearity, geometric consistency, slice thickness, spatial resolution, uniformity and low-contrast resolution. The quantitative analysis results of each image quality parameter obtained from 16 times of scanning within 16 months were normalized to the mean and the standard deviations were recorded. A run chart analysis was created to determine the warnings and action thresholds. \\n \\n \\nResults \\nThe software built in-house can quantitatively analyze the image parameters of the two scanning modes of OBI system. The low-contrast resolution of Half-Fan was better than that of Full-Fan, whereas the spatial resolution of Full-Fan was superior to that of Half-Fan. One standard deviation (1σ) was set as the warning threshold and 2 standard deviations (2σ) as the action threshold, respectively. The tolerance level of Half-Fan was smaller than that of Full-Fan. \\n \\n \\nConclusion \\nSelf-developed software enables quantitative analysis of accelerator image quality parameters, establishes warning and action tolerance of quality assurance and provides guidance for image quality assurance under image-guided radiotherapy specification. \\n \\n \\nKey words: \\nOn-board imaging; Image quality; Quantitative evaluation; Threshold setting\",\"PeriodicalId\":10288,\"journal\":{\"name\":\"中华放射肿瘤学杂志\",\"volume\":\"28 1\",\"pages\":\"919-923\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2019-12-15\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华放射肿瘤学杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/CMA.J.ISSN.1004-4221.2019.12.009\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华放射肿瘤学杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/CMA.J.ISSN.1004-4221.2019.12.009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的对直线加速器车载成像系统的图像质量参数进行定量分析,建立预警和动作阈值。方法在Varian EDGE线性加速器上以全扇形和半扇形CBCT扫描模式对Catphan604体模进行重复扫描,并利用内部基于Python语言开发的软件分析图像质量参数,如CT数线性、几何一致性、切片厚度、空间分辨率、均匀性和低对比度分辨率。将16个月内16次扫描获得的每个图像质量参数的定量分析结果归一化为平均值,并记录标准偏差。已创建运行图分析以确定警告和操作阈值。结果自制的软件可以定量分析OBI系统两种扫描模式的图像参数。半扇形的低对比度分辨率优于全扇形,而全扇形的空间分辨率优于半扇形。将一个标准偏差(1σ)设置为警告阈值,将两个标准差(2σ)分别设置为行动阈值。半风扇的公差水平小于全风扇。结论自行开发的软件能够定量分析加速器图像质量参数,建立质量保证的预警和行动容忍度,为图像引导放疗规范下的图像质量保证提供指导。关键词:车载成像;图像质量;定量评价;阈值设置
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A quantitative evaluation on the image-quality parameters and quality assurance thresholds setting of accelerator on-board imaging system
Objective To achieve quantitative analysis of image quality parameters and establish warning and action thresholds for the on-board imaging (OBI) system of linear accelerator. Methods The Catphan604 phantom was repeatedly scanned in the Full-Fan and Half-Fan CBCT scanning modes on a Varian EDGE linear accelerator, and the software based on Python language development in-house was utilized to analyze image quality parameters, such as CT number linearity, geometric consistency, slice thickness, spatial resolution, uniformity and low-contrast resolution. The quantitative analysis results of each image quality parameter obtained from 16 times of scanning within 16 months were normalized to the mean and the standard deviations were recorded. A run chart analysis was created to determine the warnings and action thresholds. Results The software built in-house can quantitatively analyze the image parameters of the two scanning modes of OBI system. The low-contrast resolution of Half-Fan was better than that of Full-Fan, whereas the spatial resolution of Full-Fan was superior to that of Half-Fan. One standard deviation (1σ) was set as the warning threshold and 2 standard deviations (2σ) as the action threshold, respectively. The tolerance level of Half-Fan was smaller than that of Full-Fan. Conclusion Self-developed software enables quantitative analysis of accelerator image quality parameters, establishes warning and action tolerance of quality assurance and provides guidance for image quality assurance under image-guided radiotherapy specification. Key words: On-board imaging; Image quality; Quantitative evaluation; Threshold setting
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
6375
期刊介绍: The Chinese Journal of Radiation Oncology is a national academic journal sponsored by the Chinese Medical Association. It was founded in 1992 and the title was written by Chen Minzhang, the former Minister of Health. Its predecessor was the Chinese Journal of Radiation Oncology, which was founded in 1987. The journal is an authoritative journal in the field of radiation oncology in my country. It focuses on clinical tumor radiotherapy, tumor radiation physics, tumor radiation biology, and thermal therapy. Its main readers are middle and senior clinical doctors and scientific researchers. It is now a monthly journal with a large 16-page format and 80 pages of text. For many years, it has adhered to the principle of combining theory with practice and combining improvement with popularization. It now has columns such as monographs, head and neck tumors (monographs), chest tumors (monographs), abdominal tumors (monographs), physics, technology, biology (monographs), reviews, and investigations and research.
期刊最新文献
Assessment of Legibility of Handwritten Prescriptions and Adherence to W.H.O. Prescription Writing Guidelines in Ahmadu Bello University Teaching Hospital Zaria - Kaduna State, Nigeria. Current situation and research progress on postoperative adjuvant radiotherapy for thymoma Mechanism of radiation combined with recombinant human endostatin in inducing myocardial fibrosis The diagnostic value of HPV detection in squamous cell carcinoma of cervical lymph node metastasis from an unknown primary site (with report of 6 cases) Application of ArcCHECK system in the verification of tomotherapy plan for esophageal cancer
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1